Naoharu Kobayashi1 and Michael Garwood1
1CMRR, Department of Radiology, University of Minnesota, Minneapolis, MN, United States
Synopsis
Keywords: Data Acquisition, Pulse Sequence Design, Inhomogeneous field
3D Fast spin echo (FSE) with frequency-modulated
(FM) pulses has been introduced for MRI in highly inhomogeneous fields. B
1 dependent phase in FM pulse
excitation and refocus was adjusted in the FM-FSE pulse sequence. Refocus flip angles in FM-FSE were determined
with prospective extended phase graph (EPG). Proposed flip angle and phase in
FM-FSE were validated in EPG simulation and experiments at 3T by introducing a
linear inhomogeneous field. Finally, in
vivo human brain imaging with T
1- and T
2-weighting was
performed using the proposed 3D FM-FSE sequence.
Introduction
Frequency-modulated (FM) pulses
can achieve broadband excitation with long pulse width, while
amplitude-modulated pulses (e.g. hard, sinc and gaussian) require very short
pulse width for broadband excitation. FM
pulses have recently been implemented in 3D MP-SSFP for MRI in highly
inhomogeneous fields to cover the broad frequencies distributed by the field
inhomogeneity1. Spin echo
(SE) is also a well-known sequence for its immunity to the field inhomogeneity. Adiabatic double spin echo, which is the most
common SE based technique with FM pulses, requires a pair of adiabatic/FM
pulses to rewind the quadratic phase2.
Therefore, when FM pulses are used in fast spin echo (FSE), even and odd
echoes show inconsistent phase so that only even or odd echoes can be used for
imaging (or simultaneous 2 k-space acquisitions). In this study, we introduce 3D FM-FSE with
consistent phase in even and odd echoes for MRI in highly inhomogeneous fields.Methods
The commonly used clinical 3D FSE sequence (aka SPACE/CUBE)
uses a short hard pulse for refocusing without slab selection to minimize echo
spacing, τ. While a hard pulse is highly efficient to
refocus the magnetization with a short pulse width, it entails a high RF peak power
especially when requiring broadband refocusing.
For example, a 12-kHz hard pulse (pulse width, pw = 100 µs)
requires RF power (B1max) of 58.7 µT for 90° flip
angle, which exceeds the limit of standard clinical MRI systems. A flattened hyperbolic secant pulse (HS2)3 yields
a better excitation profile and only requires B1max = 18.1 µT, although at the expense of pw = 3.6 ms (Fig.1A). In the proposed FM-FSE, an HS2 pulse was used
for excitation and refocusing to make even and odd echo phase consistent
(Fig.1B). Since FM pulses introduce B1/flip
angle dependent phase to the magnetization, the excitation and refocusing pulse
phase was adjusted with the B1 dependent phase, φ(α)
(Fig.1C).
Flip angles in 3D FM-FSE were
determined with prospective extended phase graph (EPG)4 including diffusion effects5,6. First, optimal
flip angles were calculated with a target of |Mxy/M0|
= 0.08 and a limit of 60° for an agar gel phantom with
known parameters (Fig.2A,B). Refocused echo signal intensities matched the
target until the flip angle reached the 60° limit. EPG simulations were also performed for
experimental validation. For brain
imaging, optimal flip angles were separately calculated with the brain
parameters at 3T7,8 (Fig.2C).
MRI studies were conducted with a Siemens Prisma 3T
MRI using a 32ch head coil. The optimal flip angles and B1 dependent
phases were first validated with the agar gel phantom. Then, in vivo human brain T1/T2-weighted
imaging was performed with healthy volunteers under an IRB approved protocol. Sequence parameters were: τ = 11
ms, TEeff = 11/396 ms, TR = 900/1800 ms, HS2 excitation and refocus (pw = 3.6 ms and BWexc = 12 kHz), BWro
= 40 kHz, echo train length = 40/72, 1.5 mm isotropic resolution and scan time
= 4:30/5:00. A linear inhomogeneous
field ΔB0 of
10, 15 or 20 mT/m was turned on during the entire scan with the z channel of
the gradient coil set, which was a field inhomogeneity of 61, 92, or 123 kHz
over the FOV. The linear inhomogeneous field was partly compensated during
excitation and readout to achieve BWexc
and BWro by modulating the
z gradient amplitude.Results
In EPG simulation, the frequency profiles
of the refocused echoes attained the desired flatness with the B1
dependent phase adjustment (Mopt), whereas the profiles without the
phase adjustment (i.e. φ(α) = 0, Mzero) showed a
mild descending slope towards the low frequency edge (Fig.3A,B). Consistent quadratic phase was observed in
all echoes for either Mopt or Mzero (Fig.3C,D). The magnitude profile for Mopt and
Mzero was identical at echo 1, but the non-flat intensity profile
was clearly seen at echo numbers >18 (Fig.3E). Phase profiles for Mopt matched well
for all echoes, but those for Mzero showed minor fluctuation
(Fig.3F).
In experiments, echo signal
intensities for Mopt showed a good match with the echo signal
intensities from EPG simulation (Fig.4A).
In profile validation, the echo profiles acquired in experiments for Mopt
and Mzero showed similar profile at echo 1, but a consistent intensity
drop towards the low frequency edge was observed in Mzero/Mopt
at echo 18 and after (Fig.4B), which is consistent to the EPG simulation (Fig.3E). In in vivo brain imaging, SAR was 50-70% of
the limit with the vendor provided SAR monitor depending on subjects. 3D FM-FSE
provided good T1/T2-weighted images by using centric-out
and sequential k-space filling (Fig.5).
Discussion
In this study, 1.5-mm isotropic
brain imaging was demonstrated with acceptable scan time of 5 min. The proposed 3D FM-FSE can readily be
accelerated by introducing partial Fourier reconstruction, parallel imaging
and/or compressed sensing reconstruction.
SAR is a serious limitation for
3D FM-FSE in highly inhomogeneous fields. However, most inhomogeneous field MRI
systems (e.g. portable MRIs) target lower field strengths (typically <1.5T)
than 3T herein. Therefore, this SAR limitation is highly mitigated in actual
highly inhomogeneous MRI systems.
Conclusion
3D FSE with frequency-modulated
pulse has been introduced for MRI in highly inhomogeneous fields.
Acknowledgements
Supporting grants: NIH
U01EB025153 and P41EB027061References
1. Kobayashi N, Parkinson B, Idiyatullin
D, Adriany G, Theilenberg S, Juchem C, et al. Development and validation of 3D
MP-SSFP to enable MRI in inhomogeneous magnetic fields. Magnetic resonance in
medicine. 2021;85(2):831-44. Epub 2020/09/07. doi: 10.1002/mrm.28469. PubMed
PMID: 32892400; PubMed Central PMCID: PMC8132587.
2. Conolly S, Glover G, Nishimura D,
Macovski A. A reduced power selective adiabatic spin-echo pulse sequence.
Magnetic resonance in medicine. 1991;18(1):28-38. Epub 1991/03/01. doi:
10.1002/mrm.1910180105. PubMed PMID: 2062239.
3. Garwood M, DelaBarre L. The return of
the frequency sweep: designing adiabatic pulses for contemporary NMR. J Magn
Reson. 2001;153(2):155-77. Epub 2001/12/13. doi: 10.1006/jmre.2001.2340. PubMed
PMID: 11740891.
4. Busse RF, Hariharan H, Vu A, Brittain
JH. Fast spin echo sequences with very long echo trains: design of variable
refocusing flip angle schedules and generation of clinical T2 contrast.
Magnetic resonance in medicine. 2006;55(5):1030-7. Epub 2006/04/07. doi:
10.1002/mrm.20863. PubMed PMID: 16598719.
5. Weigel M. Extended phase graphs:
dephasing, RF pulses, and echoes - pure and simple. Journal of magnetic
resonance imaging : JMRI. 2015;41(2):266-95. Epub 2014/04/17. doi:
10.1002/jmri.24619. PubMed PMID: 24737382.
6. Kiselev VG. Calculation of diffusion
effect for arbitrary pulse sequences. J Magn Reson. 2003;164(2):205-11. Epub
2003/09/27. PubMed PMID: 14511588.
7. Bojorquez JZ, Bricq S, Acquitter C,
Brunotte F, Walker PM, Lalande A. What are normal relaxation times of tissues
at 3 T? Magnetic resonance imaging. 2017;35:69-80. Epub 2016/10/28. doi:
10.1016/j.mri.2016.08.021. PubMed PMID: 27594531.
8. DeLano MC, Cooper TG, Siebert JE,
Potchen MJ, Kuppusamy K. High-b-value diffusion-weighted MR imaging of adult
brain: image contrast and apparent diffusion coefficient map features. AJNR
American journal of neuroradiology. 2000;21(10):1830-6. Epub 2000/12/08. PubMed
PMID: 11110534.