Amir Seginer1 and Rita Schmidt2
1Siemens Healthcare Ltd, Rosh Ha'ayin, Israel, 2Department of Brain Sciences, Weizmann Institute of Science, Rehovot, Israel
Synopsis
Keywords: Data Acquisition, Artifacts, ultra-high field
Rapid 3D steady-state sequences such as SWI are sensitive to semi-periodic physiological fluctuations (e.g., cardiac pulsation, breathing, and eye movement) resulting in repeating artifacts in the images. Randomization of the phase-encoding order reduces the above artifacts but results in apparent noise from slow global changes (like motion or eddy currents changes). We propose a new semi-randomized acquisition order that allows to set a cutoff frequency for artifact suppression; above which artifacts are suppressed, whereas artifacts from slower changes are unaffected. Simulations and SWI human brain scanning at 7T validate the method.
Introduction
Rapid steady-state sequences offering 3D acquisitions and multiple contrasts1 are widely used. These sequences are, however, sensitive to semi-periodic physiological fluctuations of the signal and/or its phase, e.g., from cardiac pulsation in the blood vessels (0.3-1 Hz), breathing (0.7-2 Hz), and eye/eyelids movement (0.25-0.33 Hz). The semi-periodicity results in incorrect interpretation of the signal’s source location, manifested as repeating artifacts in the image whose intensity depends on the scan parameters (like TR and flip angle). Susceptibility Weighted Imaging (SWI) that utilizes the benefits of a 3D steady-state acquisition with short TR is widely used, but is also very sensitive to the above artifacts.
Here we examine how by simply reordering the phase encodes (PEs) of the acquisition one can reduce the artifacts from local fluctuations, as long as the source of the fluctuating signal is small. We pick up on previous works2-8 considering random and semi-random acquisition ordering to suppress the above artifacts. A randomized order can suppress any repeating artifact but it may also increase the apparent noise, e.g., in cases of slow movement of the subject or slow changes due to eddy currents. Here we propose a new semi-randomized space-filing curve that allows to set a cutoff frequency for artifact suppression. Thus, fluctuations above the cutoff are suppressed, whereas changes from slow movement are not affected. We examine and characterize the above methods using simulations and SWI human scanning at 7T including parallel imaging (using a Cartesian acquisition).Methods
Five ordering schemes were
tested in simulation and in vivo:
-
'Ordered' - The standard
order, sequentially going through a 2D phase encoding grid (column by column,
or row by row).
- 'Full
Scrambling' - Random shuffling of all the possible PEs.
-
'Gilbert'
- A generalized version of the Hilbert curve, supporting rectangular domains9.
- 'Segmented
Scrambling' - Targeted at modulations faster than a cutoff frequency $$$f_\mathrm{cutoff}$$$. The
ordered set is treated as a 1D array, split into segments of length $$$1/(\mathrm{TR}\cdot{f_\mathrm{cutoff}})$$$ and
each, independently, randomly permuted. [In practice, segments where not of
identical length, as integer lengths are involved.]
- 'Local
Scrambling' - A random shuffling aiming to mimic the time shift distribution of
Segmented Scrambling but without using explicit segments. (For the target
cutoff, an effective cutoff $$$f_\mathrm{cutoff}/1.2$$$ was used.)
Simulations
were limited to 2D, discarding the readout dimension. Images of a 2D object
(point or disk) with a periodic phase $$$\exp(i\cdot2\pi{ft})$$$ were
generated for different PE orderings. As the only present object was
fluctuating, the image was considered as an artifact only image. The maximum
image signal, at each frequency, was taken as the worst-case artifact, or
“error”, for that frequency. A simulation comparing modulating disks of different
diameters was also performed to assess the qualitative effect of less localized
artifacts.
In vivo
scans were performed on a 7 T system (MAGNETOM Terra, Siemens Healthcare,
Erlangen) using a modified GRE sequence that allows arbitrary ordering of the PE
lines and an option to interleave two orderings. Interleaving two orderings
allows a simultaneous acquisition of both and thus a comparison at the same
conditions. See figures for scan parameters.
Results
Fig. 1
shows the examined orderings and compares the intensity of the worst “artifact”
as a function of modulation frequency. The effect of the TR, of the number of PE
samples ($$$N_\mathrm{tot}$$$), and
of the cutoff $$$f_\mathrm{cutoff}$$$ on the
worst artifacts is shown in Fig. 2. Fig. 3 shows the effect of the size of the modulating
object on the resulting “noise”. In vivo results comparing the different ordering
schemes are shown in Fig. 4, each acquired interleaved with the Ordered scheme.
Fig. 5 shows an in vivo example demonstrating movement and breathing artifacts.
This case emphasizes that Full Scrambling may increase the overall noise in the
image while Local Scrambling still provides a good quality image.Discussion
As
expected, Full Scrambling minimizes the artifact from a periodic signal (except
at extreme modulation frequencies of $$$\sim1/(\mathrm{scan\ time})$$$ and $$$\sim1/\mathrm{TR}$$$). However, Full Scrambling involves large steps in k-space which have led to their own artifacts (affected also by the readout bandwidth, but not shown here). A Hilbert-like curve minimizes the PE k-space steps but suppresses the artifact less effectively (Figs. 1, 4) and in addition is more difficult to implement for non-uniform sampling, e.g., GRAPPA with its center of k-space fully sampled.
Local Scrambling, however, benefits from both worlds. K-space steps are limited in size and the suppression of the artifacts practically matches that of Full Scrambling for frequencies above $$$f_\mathrm{cutoff}$$$. In
this approach, the lower frequency modulations (typically associated with
global motion, e.g., head motion) are hardly affected, thus no extra noise
appears (Fig. 5). The scrambling approach benefits from 3D acquisitions and
increasing resolution since the artifact magnitude depends on the total number
of PE samples, behaving like $$$\sim1/\sqrt{N_\mathrm{tot.}}$$$.Conclusions
Local
Scrambling of steady-state acquisitions, especially 3D with high resolution,
achieves a significant reduction of artifacts arising from semi-periodic localized
fluctuations. This is especially useful in SWI which is sensitive to
fluctuations in blood vessels and the eyes. The method was demonstrated with
high resolution SWI brain imaging at 7T MRI. Global changes (motion or phase)
still require complementary corrections.Acknowledgements
No acknowledgement found.References
- Hargreaves BA. Rapid gradient-echo imaging. Journal of Magnetic Resonance Imaging 2012;36:1300–1313.
- Bailes DR, Gilderdale DJ, Bydder GM, Collins AG, Firmin DN. Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artefacts in MR imaging. Journal of Computer Assisted Tomography 1985;9:835–838.
- Haacke ME, Patrick JL. Reducing motion artifacts in two-dimensional Fourier transform imaging. Magnetic Resonance Imaging 1986;4:359–376.
- Korin HW, Riederer SJ, Bampton AEH, Ehman RL. Altered phase-encoding order for reduced sensitivity to motion in three-dimensional MR imaging. Journal of Magnetic Resonance Imaging 1992;2:687–693.
- Wilman AH, Riederer SJ. Improved centric phase encoding orders for three-dimensional magnetization-prepared MR angiography. Magnetic Resonance in Medicine 1996;36:384–392.
- Jhooti P, Wiesmann F, Taylor AM, Gatehouse PD, Yang GZ, Keegan J, Pennell DJ, Firmin DN. Hybrid ordered phase encoding (HOPE): An improved approach for respiratory artifact reduction. Journal of Magnetic Resonance Imaging 1998;8:968–980.
- Parker DL, Goodrich KC, Roberts JA, Chapman BE, Jeong EK, Kim SE, Tsuruda JS, Katzman GL. The need for phase-encoding flow compensation in high-resolution intracranial magnetic resonance angiography. Journal of Magnetic Resonance Imaging 2003;18:121–127.
- Wermer MJH, van Walderveen MAA, Garpebring A, van Osch MJP, Versluis MJ. 7 Tesla MRA for the differentiation between intracranial aneurysms and infundibula. Magnetic Resonance Imaging 2017;37:16–20.
- Červený J, Generalized Hilbert ("gilbert") space-filling curve for rectangular domains of arbitrary (non-power of two) sizes. https://github.com/jakubcerveny/gilbert/blob/master/gilbert2d.py, accessed:2022.07.10.