Kartiga Selvaganesan1, Yonghyun Ha2, Chenhao Sun2, Anja Samardzija1, Zhehong Zhang1, Heng Sun1, Gigi Galiana2, and Todd Constable2
1Biomedical Engineering, Yale University, New Haven, CT, United States, 2Radiology and Biomedical Imaging, Yale University, New Haven, CT, United States
Synopsis
Keywords: Pulse Sequence Design, Low-Field MRI
The cost
and complexity of MR scanners can be significantly reduced by eliminating
conventional linear gradient coils, and instead using RF coils for spatial
encoding. Here we have developed a novel multi-echo RF phase encoding pulse
sequence that exploits the Bloch-Siegert shift for nonlinear spatial encoding
in a nonuniform, field-cycling, low-field MR system. Phantoms of varying sizes
were successfully imaged using this pulse sequence, demonstrating that this
technique can be used to perform gradient-free imaging in an inhomogeneous
B0-field at low-field.
Introduction
A growing
body of research explores the development of low-field, open MRI scanners for
low cost, highly accessible, point-of-care screening [1,
2]. However, much of the work in this area has focused
on utilizing a conventional closed-bore geometry for both magnet and gradient
coils which presents significant design constraints for MRI. Spatial encoding
using radiofrequency (RF) coils is a promising solution to opening up the
scanner design, but current methods focus on linear B1-fields or cylindrical geometries
[3-5] which introduces complexities when trying to create on
open MR system. In this work, we present a novel pulse sequence that exploits
the Bloch-Siegert shift [6] imparted by an off-resonance pulse
to perform gradient-free, nonlinear spatial encoding using a multichannel RF
planar array, in a low-field MR system.Methods
Figure 1
shows the multi-echo encoding CPMG sequence used in this study. In this
sequence an excitation pulse is followed by a train of refocusing pulses
interleaved with Bloch-Siegert off-resonance pulses. These pulses act much like
a blipped phase encode gradient in conventional turbo spin echo imaging; after
acquisition of the first spin echo (k0), each subsequent echo accumulates
phase allowing for extended acquisition along a k-trajectory, noting that
different parts of the image will have different k-space trajectories. To obtain a
full encoding trajectory using blipped RF, the off-resonance pulses were first
applied to even echoes, acquiring k0 to kn (Figure 1A),
then again on odd echoes, acquiring k0 to k-n (Figure
1B). 320 echoes were acquired with 160 averages, and an ESP=1.3ms.
Experiments
were conducted on a single-sided, low-field electromagnet [7] featuring a nonuniform main
magnetic field. To make it possible to image in the presence of a nonuniform
B0-field, the system is designed to use field cycling, such that spins are
polarized at a high field (0.6T) and imaged at a lower field (24mT), thereby
minimizing field inhomogeneity effects while improving polarization. The
experimental setup and the phantoms imaged are shown in Figure 2A and B
respectively. A large, Tx-only volume coil placed inside the MR scanner was
used to transmit the 90/180 pulses. A switch tuned RF planar array coil was positioned
at the base of the magnet to transmit the off-resonance Bloch-Siegert pulses
for spatial encoding (Figure 3A) and to receive the NMR signal. Figure 3B shows
a diagram of this 3x3 RF planar array. Five channels in the array were used for
nonlinear spatial encoding (blue), and a single channel in the center of the
FOV for receive (green). Results
Encoding
patterns were created by selecting subsets of two and three coils to transmit
the Bloch-Siegert pulses from the five transmit channels. Figure 4A shows an
example of a positive k-trajectory echo train generated by a single encoding
pattern. From this raw data, the full k-trajectory is formed by re-ordering the
midpoints of each echo (Figure 4B). It is important to note that since the
spatial encoding fields are nonlinear, k-space traversal is also nonlinear, so
each echo contains both kx and ky information.
Images
were reconstructed using the conjugate gradient method for algebraic
reconstruction with nonlinear encoding fields [8]. The algorithm used encoding field
maps calculated from the Biot-Savart for each transmit channel. Image reconstruction
was performed with 8 encoding patterns, over a 64x64 matrix for a 20cmx20cm FOV.
The last column in Figure 5 shows the 2D image of each of the bottle phantoms.
For comparison, a Bloch simulation of the experiment was also conducted using
the same encoding field patterns and maps previously described. The first and
second columns in Figure 5 show the digital phantoms used to simulate the MR
signal, as well as the corresponding 2D reconstruction results. Discussion
We have
developed a novel multi-echo encoding pulse sequence that exploits the
Bloch-Siegert shift for nonlinear spatial phase encoding. This sequence has a
short echo time, and utilizes a series of short RF refocusing pulses in
combination with high power off-resonance pulses to allow for extended
acquisition of k-space within a single TR.
Overall,
the results demonstrate the feasibility of using this technique to perform
spatial encoding and gradient-free imaging in a nonuniform field. The 2D images
show that with just 8 encoding patterns, we can still clearly delineate phantom
geometries and boundaries. The experimental results also match the expected
simulated results, with similar bright areas, and rings seen in both images. These
artifacts come from high frequency noise that can be eliminated by receiving
signal on multiple channels instead of just one, and by incorporating electromagnetic
interference (EMI) correction methods. Future work will focus on expanding this
multi-echo encoding technique to include more encoding patterns and receive
channels for imaging over a larger FOV.Conclusion
In this
study, we have shown that gradient-free imaging is possible in a single-sided,
low-field open magnet system with an inhomogeneous B0-field. By eliminating the
uniform B0-field requirement and need for conventional cylindrical gradients,
we can design MR systems in any size or shape, thereby increasing access to
low-cost, application specific MR imaging.
Acknowledgements
No acknowledgement found.References
1. Nasri,
J., et al., Office-based, point-of-care,
low-field MRI system to guide prostate interventions: recent developments.
EMJ Urology, 2021: p. 83-90.
2. McDaniel, P.C., et al., The MR Cap: A single-sided MRI system designed
for potential point-of-care limited field-of-view brain imaging. Magnetic
resonance in medicine, 2019. 82(5):
p. 1946-1960.
3. Katscher, U., J. Lisinski, and P.
Börnert, RF encoding using a multielement
parallel transmit system. Magnetic Resonance in Medicine, 2010. 63(6): p. 1463-1470.
4. Stockmann, J.P., et al., Transmit Array Spatial Encoding (TRASE)
using broadband WURST pulses for RF spatial encoding in inhomogeneous B0
fields. (1096-0856 (Electronic)).
5. Sun, H., S. Yong, and J.C. Sharp, The twisted solenoid RF phase gradient
transmit coil for TRASE imaging. Journal of Magnetic Resonance, 2019. 299: p. 135-150.
6. Bloch, F. and A. Siegert, Magnetic Resonance for Nonrotating Fields.
Physical Review, 1940. 57(6): p.
522-527.
7. Constable, R.Todd., et al., Design of a novel class of open MRI devices
with nonuniform B0, field cycling, and RF spatial encoding. in 27th ISMRM Annual Meeting.
2019. Montreal, QC, Canada.
8. Wan, Y.Q., Maolin, Galiana, Gigi,
Constable R. Todd. Nonlinear RF spatial
encoding with multiple transmit coils based on Bloch-Siegert shift. in 24th ISMRM Annual Meeting 2016.