Nelson Homero Gil1, Wei-Ching Lo2, Bryan Clifford2, Min Lang1, Komal Awan1, Daniel Nicolas Splitthoff3, Daniel Polak3, Stephen Cauley1, and Susie Huang1
1Radiology, Massachusetts General Hospital, Boston, MA, United States, 2Siemens Medical Solutions USA, Boston, MA, United States, 3Siemens Healthcare GmbH, Erlangen, Germany
Synopsis
Keywords: Motion Correction, Neurodegeneration
Volumetric
brain MRI using the T1-weighted-MPRAGE sequence is an important component of
the clinical evaluation of dementia. However, MPRAGE has a long acquisition
time and is especially prone to patient motion artifact. A recent development
to address this issue is a technique called Scout Accelerated Motion Estimation
and Reduction (SAMER). In this work, we used a set of 90 MPRAGE scans derived
from 10 healthy volunteers to demonstrate that SAMER is effective at correcting
various degrees of motion, including severe motion with non-diagnostic image
quality, and greatly increases the accuracy of volumetric brain measurements.
Introduction
Magnetic
Resonance Imaging (MRI) is limited in practice by patient motion artifacts,
which can result in costly non-diagnostic scans [1-4]. One of the most motion-prone
exams is volumetric brain MRI, which quantifies characteristics such as
cortical volume and thickness and can help diagnose neurodegenerative diseases [5]. Specifically, the 3D-T1-weighted-MPRAGE
sequence is a standard part of MRI protocols to evaluate memory loss and serves
as the basis for volumetric measurements; nevertheless, it is especially prone
to long acquisition times and motion artifact [6].
One approach to shorten
acquisition time of the MPRAGE sequence uses a novel technique called Wave
Controlled Aliasing in Parallel Imaging (Wave-CAIPI) [7]. Wave-CAIPI-accelerated
MPRAGE has been validated clinically against standard MPRAGE and shown to equivalently
assess cortical volume and thickness [8]. Subsequently, a
complementary approach, Scout-Accelerated Motion Estimation and Correction
(SAMER) was developed to reduce motion artifact [9, 10], and is currently
undergoing clinical evaluation.
Preliminary work on one
volunteer suggested that SAMER-MPRAGE successfully corrected mild motion
artifact and resulted in cortical volume and thickness estimates comparable to
those obtained with standard MPRAGE [11]. In this study, we
used 90 MRI scans derived from ten volunteers to show that SAMER can correct
motion-induced errors even in the most motion-prone regions of the brain and
facilitate accurate volumetric analysis. Methods
Ten
volunteers underwent five different in vivo MPRAGE scans at R=4-fold
acceleration. The first was a Wave-CAIPI-MPRAGE sequence which was considered the
reference standard, while the remaining four corresponded to acquisitions
performed with SAMER-MPRAGE with varying degrees of motion corruption: “no”,
“mild”, “moderate”, and “severe” motion [1]. Motion-corrupted scans
underwent motion correction with SAMER. In total, each volunteer generated nine
scans: Wave-CAIPI-MPRAGE and SAMER-MPRAGE, pre- and post-motion correction for
each motion state, resulting in 90 effective total number of MRI scans
analyzed.
All scans were acquired
on a 3T system (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany) using a
20-channel head coil. The sequence parameters used were: FOV 256x256x192mm;
matrix 256x256x192; TR/TE/TI 2300/3.2 (3.3 for SAMER)/1000ms; flip angles 8°;
bandwidth 200 Hz/Px; acquisition times 2:39 min. Each volunteer was instructed
to perform varying degrees of deep breathing during the motion scans, something
known to introduce motion artifact. Motion estimation and image reconstruction
were performed with an online reconstruction directly on the scanner and
generated additional motion-corrected
images.
For the nine sets of
images from each volunteer, Freesurfer [12] calculated cortical
volume for 11 brain regions (frontal, parietal, temporal, occipital lobes,
cingulate gyrus, insula, hippocampus, basal ganglia, brain stem, cerebellum,
cerebral white matter (WM)) and thickness for 6 brain regions (frontal,
parietal, temporal, occipital lobes, cingulate gyrus, insula). Cortical volumes
and thicknesses for each region were averaged across all volunteers. Percent-error
for the averaged cortical volume and thickness was calculated, treating Wave-CAIPI-MPRAGE
results as the reference standard:
$$100*\frac{\left|AveragedVolumeOrThickness - Reference\right|}{Reference}$$
The
statistical significance of each SAMER motion correction was assessed using
paired Wilcoxon rank sum tests comparing the cortical thickness and volume
across each anatomical region for uncorrected versus corrected scans.Results and Discussion
Freesurfer
completed cortical volume and thickness calculations on 84/90 (~93%) scans; the
6 calculations that failed were mostly due to severe motion artifact before
motion correction, and 5/6 of these were successfully completed in the
corresponding SAMER-corrected scans. Representative cases illustrate the effect
of various degrees of motion and subsequent correction with SAMER (Figure 1),
as well as the corresponding Freesurfer segmentation (Figure 2).
Errors introduced by
motion artifact resulted in Freesurfer generally underestimating the reference
volume (Figures 2/3A) in a manner directly proportional with the degree of
motion, consistent with prior reported associations [13]. The most affected
anatomical areas appear to be the temporal lobe (Figure 2) and the cerebral
white matter, with percent error values on severe motion scans reaching 38% and
32%, respectively (Figure 3B). At the same time, SAMER considerably reduced the
volume calculation error across all anatomical areas; for example, the percent
error values corresponding to the temporal lobe and cerebral white matter were
19% and 10%, respectively.
Estimates of the
cortical thickness behaved similarly to the ones for volume in that the
temporal lobe is the most affected (Figure 4A). However, the magnitude of the
percent error is generally less across the examined anatomical areas (Figure 4B),
with the highest value for the severe motion scans being 15% for the temporal
lobe. SAMER’s contributions to the accuracy of thickness estimates are more
modest. For example, the temporal lobe percent error was reduced to 11%. There were
also areas that showed no significant correction, such as the insula.
Interestingly, the largest relative reduction in error with SAMER occurred in
the occipital lobe (11% to 4%), where severe motion resulted in overestimation
of the cortical thickness, in contrast to the remainder of the anatomical
areas. Overall, SAMER results in
statistically significant improvements in calculated cortical volume and
thickness between uncorrected and corrected scans across most anatomical
regions (Figure 5).Conclusion
SAMER
corrects for severe motion artifact and can turn non-diagnostic-quality scans
into ones that allow for accurate computation of cortical volume and thickness.
Our results suggest that SAMER has strong potential for evaluation in practical
clinical settings and can contribute to diagnosing dementia and
neurodegenerative disease.Acknowledgements
This work was supported in part by NIH research grants P41EB030006-01, 5U01EB025121-03 and a research grant from Siemens Healthineers.
References
[1] Andre JB,
Bresnahan BW, Mossa-Basha M, Hoff MN, Smith CP, Anzai Y, et al. Toward
Quantifying the Prevalence, Severity, and Cost Associated With Patient Motion
During Clinical MR Examinations. J Am Coll Radiol. 2015;12:689-95.
[2] Munn Z, Pearson A,
Jordan Z, Murphy F, Pilkington D, Anderson A. Patient Anxiety and Satisfaction
in a Magnetic Resonance Imaging Department: Initial Results from an Action
Research Study. J Med Imaging Radiat Sci. 2015;46:23-9.
[3] Havsteen I, Ohlhues
A, Madsen KH, Nybing JD, Christensen H, Christensen A. Are Movement Artifacts
in Magnetic Resonance Imaging a Real Problem?-A Narrative Review. Front Neurol.
2017;8:232.
[4] Slipsager JM,
Glimberg SL, Sogaard J, Paulsen RR, Johannesen HH, Martens PC, et al.
Quantifying the Financial Savings of Motion Correction in Brain MRI: A
Model-Based Estimate of the Costs Arising From Patient Head Motion and
Potential Savings From Implementation of Motion Correction. J Magn Reson
Imaging. 2020;52:731-8.
[5] Lehmann M, Douiri A,
Kim LG, Modat M, Chan D, Ourselin S, et al. Atrophy patterns in Alzheimer's
disease and semantic dementia: a comparison of FreeSurfer and manual volumetric
measurements. Neuroimage. 2010;49:2264-74.
[6] Mugler JP, 3rd,
Brookeman JR. Rapid three-dimensional T1-weighted MR imaging with the MP-RAGE
sequence. J Magn Reson Imaging. 1991;1:561-7.
[7] Polak D, Setsompop
K, Cauley SF, Gagoski BA, Bhat H, Maier F, et al. Wave-CAIPI for highly
accelerated MP-RAGE imaging. Magn Reson Med. 2018;79:401-6.
[8] Longo MGF, Conklin
J, Cauley SF, Setsompop K, Tian Q, Polak D, et al. Evaluation of Ultrafast
Wave-CAIPI MPRAGE for Visual Grading and Automated Measurement of Brain Tissue
Volume. AJNR Am J Neuroradiol. 2020;41:1388-96.
[9] Polak D, Splitthoff
DN, Clifford B, Lo WC, Huang SY, Conklin J, et al. Scout accelerated motion
estimation and reduction (SAMER). Magn Reson Med. 2022;87:163-78.
[10] Polak D, Hossbach
J, Splitthoff DN, Clifford B, Lo WC, Tabari A, et al. Motion guidance lines for
robust data-consistency based retrospective motion correction in 2D and 3D MRI.
Magn Reson Med. 2023. doi:10.1002/mrm.29534
[11] Lo WC, Conklin J,
Clifford B, Tian Q, Polak D, Splitthoff D, et al. Evaluation of Scout
Accelerated Motion Estimation and Reduction (SAMER) for measuring brain volume
and cortical thickness. International
Society for Magnetic Resonance in Medicine. London, England, United
Kingdom 2022.
[12] Fischl B.
FreeSurfer. Neuroimage. 2012;62:774-81.
[13] Reuter M, Tisdall MD, Qureshi A, Buckner RL,
van der Kouwe AJW, Fischl B. Head motion during MRI acquisition reduces gray
matter volume and thickness estimates. Neuroimage. 2015;107:107-15.