Yu Chen1, Runke Wang1, Ruokun Li2, Fuhua Yan2, and Yuan Feng1,2,3
1Shanghai JiaoTong University, School of Biomedical Engineering, Shanghai, China, 2Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China, 3Institute of Medical Robotics, Shanghai Jiao Tong University, Shanghai, China
Synopsis
Keywords: Elastography, Brain
Conventional gradient echo (GRE) based
Magnetic Resonance Elastography (MRE) suffers from long TE and scan time at low
mechanical excitation (<25Hz). Displacement Encoding with a Stimulated Echo
(DENSE) can record displacement with short TR, useful for low-frequency
excitation. Here, 3D excitation was used to compensate for potential signal
loss. A multiphase acquisition scheme with stack-of-star radial sampling was also
used along with interleaved multi-slab acquisition and Hadamard encoding for
MRE acceleration. Phantom experiments were conducted for validation. 32 slices
brain MRE scan at 20Hz can be achieved in 11min30s, 5 folds faster than conventional GRE-based
MRE.
Summary of Main Findings
A 3D
GRE-based fast Multiphase DENSE MRE sequence with stack-of-star radial
sampling, interleaved multi-slab acquisition and Hadamard encoding was proposed
for MRE acceleration. Compared with conventional GRE-based MRE, the proposed
sequence can achieve 5-fold acceleration.Introduction
Magnetic resonance elastography (MRE) is a useful method
to detect the biomechanical properties of soft tissues noninvasively1, 2. Conventional gradient-echo (GRE) based MRE
sequence has to be synchronized with mechanical excitation3, elongating the scanning time when the
excitation frequency is low (<25Hz). Displacement encoding with stimulated
echo (DENSE) can record displacement with short TR. Therefore, DENSE is ideal
for low-frequency encoding than conventional GRE-MRE at low mechanical
excitation scenarios. It has been shown that multiphase DENSE MRE with radial
sampling has the advantage of acceleration and artifact suppression4. In this study, we proposed a 3D GRE-based
fast multiphase DENSE MRE sequence with stack-of-star radial sampling, interleaved
multi-slab acquisition5, 6, and Hadamard encoding7 for acceleration of MRE. Phantom experiments
were conducted for validation. A human brain MRE scan was also performed to
demonstrate its potential for MRE acceleration.Methods
The DENSE method can be divided into two
parts: Encode and Decode. Spatial modulation of magnetization block was used
for encoding the initial place. In the imaging block, a decode gradient was
used to generate a DENSE echo.
$$M_{x y}^{D E N S E}=\frac{1}{2} M_{0} \sin \theta \cdot e^{-\frac{T M}{T_{1}}} \cdot e^{i \mathrm{~K} \Delta \mathrm{r}}$$ where $$$M_0$$$ is the steady state longitudinal magnetization, $$$\theta$$$ is the flip angle, $$$TM$$$ is the mixing time, $$$K$$$ is the area under the motion encoding gradient and $$$\Delta r$$$ represents the displacement.
The proposed sequence timing diagram is shown
in Figure
1. Motion encoding block (MEB)
encoded the initial position and motion decoding blocks (MDB) decoded and
generated the DENSE echo. In
MEB, the encoding gradients were set up in three encoding directions
simultaneously for Hadamard encoding. 1 MEB was followed by N times MDB (N =
slice phase encoding (SPE) number * slice oversample rate / 2). One MEB plus N
MDB is called an imaging section, which filled the SPE-Spokes k-space a half
from the center to the edge. In one MDB, 8 times in plane golden-angle radial
sampling acquisition was performed for imaging 4 motion phases for the 2
imaging slabs respectively. A broad-band minimum-phase RF pulse is used for 3D
excitation. The decoding gradient was also set up ib three encoding directions
simultaneously. The Hadamard DENSE signal was composited of three direction
displacements. With 4 times Hadamard encoding, the origin displacement can be
decoded using the least square method.
$$M_{x y}^{DENSE}=\frac{1}{2} M_{0} \sin \theta \cdot e^{-\frac{T M}{T_{1}}} \cdot e^{i\left(\mathrm{~K}_{\mathrm{x}} \Delta \mathrm{x}+\mathrm{K}_{\mathrm{y}} \Delta \mathrm{y}+\mathrm{K}_{\mathrm{z}} \Delta \mathrm{z}\right)}$$ $$\left[\begin{array}{l}\Phi_{1} \\\Phi_{2} \\\Phi_{3} \\\Phi_{4}\end{array}\right]=\left[\begin{array}{llll}+1 & +1 & +1 & +1 \\-1 & +1 & +1 & +1 \\+1 & -1 & +1 & +1 \\+1 & +1 & -1 & +1\end{array}\right]\left[\begin{array}{c}K_{x} \Delta x \\K_{y} \Delta y \\K_{z} \Delta z \\\Phi_{0}\end{array}\right]=E\left[\begin{array}{c}K_{x} \Delta x \\K_{y} \Delta y \\K_{z} \Delta z \\\Phi_{0}\end{array}\right]$$ $$\left[\begin{array}{c}\Delta x \\\Delta y \\\Delta z \\\sim\end{array}\right]=\operatorname{inv}(E)\left[\begin{array}{l}\Phi_{1} \\\Phi_{2} \\\Phi_{3} \\\Phi_{4}\end{array}\right] \cdot *\left[\begin{array}{c}1 / K_{x} \\1 / K_{y} \\1 / K_{z} \\\sim\end{array}\right]$$ where $$$K_x, K_y, K_z$$$ are the areas of the encoding gradients on three axes respectively and $$$\Delta x, \Delta y, \Delta z$$$ represent the displacements of three directions. $$$\Phi_i$$$ is the $$$i^{th}$$$ time composite phase with Hadamard encoding.
The proposed sequence was implemented and
tested on a 3T scanner (uMR 790, United Imaging Healthcare, Shanghai, China). A custom-built electromagnetic actuator
was used to generate the vibration. The imaging parameters were TR=12.5ms,
TE=4.2ms, matrix size=80*80, slice=32, voxel size=3*3*3 mm, spokes /slab=125, BW=500Hz/Pixel, $$$G_{MEG}$$$=35mT/m, $$$G_{dura}$$$=0.85ms. A phantom with two stiff inclusions,(I) was
stiffer than (II) (Figure 2), was constructed for sequence validation at
60Hz8. A healthy volunteer was recruited for
testing the proposed sequence at 20Hz.Results
Phantom
experiments result showed that the displacement field recorded using the proposed
was similar to that from GRE-MRE sequence. (Figure 2)
Moreover, the two stiff inclusions can also be detected and distinguished from
each other in the inversion shear modulus map (Figure 3). The
human brain experiment showed the proposed sequence can clearly record the
displacement and the potential in GRE-based MRE acceleration with low mechanical excitation.
(Figure 4)Conclusion
We proposed a
3D GRE-based fast multiphase DENSE MRE sequence with stack-of-star radial sampling,
interleaved multi-slab acquisition and Hadamard encoding for MRE acceleration. The
phantom experiment result showed wave images acquired were similar to that of
GRE-MRE. The human brain scan showed its potential for acceleration with low-frequency
mechanical excitation.Acknowledgements
Funding support from the National Natural Science Foundation of China (grant 32271359) and the Natural Science Foundation of Shanghai (grant 22ZR1429600) is acknowledged.References
[1] S. K. Venkatesh, "Clinical Applications of Liver Magnetic Resonance Elastography: Chronic Liver Disease," in Magnetic Resonance Elastography, 2014, ch. Chapter 4, pp. 39-60.
[2] L. V. Hiscox et al., "Magnetic resonance elastography (MRE) of the human brain: technique, findings and clinical applications," Phys Med Biol, vol. 61, no. 24, pp. R401-R437, Dec 21 2016, doi: 10.1088/0031-9155/61/24/R401.
[3] J. Strasser, M. T. Haindl, R. Stollberger, F. Fazekas, and S. Ropele, "Magnetic resonance elastography of the human brain using a multiphase DENSE acquisition," Magn Reson Med, vol. 81, no. 6, pp. 3578-3587, Jun 2019, doi: 10.1002/mrm.27672.
[4] R. K. Wang et al., "Fast magnetic resonance elastography with multiphase radial encoding and harmonic motion sparsity based reconstruction," (in English), Phys. Med. Biol., Article vol. 67, no. 2, p. 13, Jan 2022, Art no. 025007, doi: 10.1088/1361-6560/ac4a42.
[5] P. Garteiser et al., "Rapid acquisition of multifrequency, multislice and multidirectional MR elastography data with a fractionally encoded gradient echo sequence," NMR Biomed, vol. 26, no. 10, pp. 1326-35, Oct 2013, doi: 10.1002/nbm.2958.
[6] C. Guenthner, S. Sethi, M. Troelstra, A. S. Dokumaci, R. Sinkus, and S. Kozerke, "Ristretto MRE: A generalized multi-shot GRE-MRE sequence," NMR Biomed, vol. 32, no. 5, p. e4049, May 2019, doi: 10.1002/nbm.4049.
[7] C. L. Dumoulin, S. P. Souza, R. D. Darrow, N. J. Pelc, W. J. Adams, and S. A. Ash, "Simultaneous acquisition of phase-contrast angiograms and stationary-tissue images with Hadamard encoding of flow-induced phase shifts," (in English), Journal of magnetic resonance imaging : JMRI, vol. 1, no. 4, pp. 399-404, 1991 1991, doi: 10.1002/jmri.1880010403.
[8] Y. Feng et al., "Viscoelastic Characterization of Soft Tissue-Mimicking Gelatin Phantoms using Indentation and Magnetic Resonance Elastography," (in English), J. Vis. Exp., Article no. 183, p. 12, May 2022, Art no. e63770, doi: 10.3791/63770.