Masa Bozic-Iven1,2, Stanislas Rapacchi3, Qian Tao2, Lothar R. Schad1, and Sebastian Weingärtner2
1Heidelberg University, Mannheim, Germany, 2Delft University of Technology, Delft, Netherlands, 3University Aix-Marseille, Marseille, France
Synopsis
Keywords: Myocardium, Arterial spin labelling
Myocardial arterial spin labeling (myoASL) holds promise for
needle-free myocardial blood flow (MBF) quantification but requires tedious
averaging over multiple breath-holds. Here, free-breathing myoASL was
implemented with dual-navigator gating, both with bSSFP and spoiled GRE
readout. Images were processed using individual blood T
1 and inversion time
correction as well as a phase-sensitive (PS) image reconstruction. Phantom
results showed PS reconstruction to reduce MBF variations for short RR and T
1
values. Perfusion values were comparable with breath-held myoASL and on par with the
literature. Ultimately, this can enable faster myoASL acquisitions with
improved patient comfort.
Introduction
First-pass perfusion
imaging is the clinical gold standard for detecting myocardial ischemia and
quantifying myocardial perfusion1. However, it requires the use of exogenous
contrast agents limiting its repeated use. Arterial spin labelling (ASL) can
provide a promising alternative based on magnetically labeled blood as
endogenous tracer. Due to low signal-to-noise ratios and high physiological
noise levels2, however, multiple averages are needed to ensure sufficient
accuracy. While myocardial ASL (myoASL)
has mostly been performed during breath-holds2,3, free-breathing,
retrospectively gated myoASL has recently been demonstrated4. However,
retrospective image selection may lead to excessive scan times. In this work,
we propose a respiratory navigated free-breathing myoASL sequence with
increased scan time efficiency and noise performance, using phase-sensitive
myoASL image reconstruction.FAIR-myoASL sequence
Imaging was performed on a 3T
scanner (Vida, Siemens). For all measurements, a double ECG-gated Flow Alternating
Inversion Recovery (FAIR) ASL sequence3 was implemented (Fig.1). For the
free-breathing sequence, as shown in Fig. 1, a pencil-beam navigator placed at
the liver dome was played prior to both inversion and image acquisitions. To
ensure that images are acquired only upon successful inversion, FAIR images
were accepted only when both consecutive navigators were valid. This
dual-heartbeat navigation also led to matching slice-selective inversion and
excitation in control images. Baseline images were conventionally navigated
within a single heartbeat.Phantom and In vivo Measurements
Data was acquired
both with bSSFP and spoiled GRE (spGRE) readout. In phantom, two control-tag
image pairs were acquired with a 6s delay between the two images. Phantom experiments
were performed with varying simulated RR intervals in a NiCl2-doped agarose
phantom. In vivo images of 4 healthy subjects (1 female, 3 male, 35±4.8
years) were obtained in 12-15s long breath-holds per image pair depending on
the heart rate (scan duration ~3min) as well as in free-breathing (scan
duration ~2-3 min). For each of the four combinations of readout and breathing
strategy, eight control-tag pairs were acquired with a 6s delay. In phantom and
in vivo, two baseline images (no inversion) were acquired with bSSFP, while
with spGRE one of the two was directly preceded by a WET saturation pulse
(“saturated baseline”, Sat-BL) to compensate for the effect of the imaging
readout5,6. All images were acquired with 1.6x1.6x8mm3 voxel size,
TE 1.36/1.97ms, TR 3.6/4.3ms, and FA 60°/17° (bSSFP/spGRE), GRAPPA 2 and
Partial Fourier (6/8). MOLLI7 was used for T1 mapping in phantom and in vivo.Data analysis
In vivo image pairs with
inconsistent inversion times (TI) were excluded, before registering the images
groupwise8. Phase-sensitive (PS) reconstruction was performed to restore the
signal polarity and image contrast. To this end, the phase difference was
unwrapped and rounded to 0 or π to extract the signal polarity9. Pixel- and
segment-wise myocardial blood flow (MBF) were reconstructed using individual
blood T1 and a Sat-BL correction as previously described5,6,10.Results
Phantom MBF values are
underestimated and vary with the RR duration, when no signal polarity
correction is used (Fig. 2b). The inflection point occurs at longer RR
intervals/TIs for longer T1 values and, also for spGRE compared to bSSFP. With
PS reconstruction the phantom MBF is largely constant over the range of simulated heart
rates.
The effect of the
polarity correction on in vivo data is shown in Fig. 3. for one representative
subject with bSSFP readout. The bright blood pool signal in the uncorrected
images (Fig. 3a) indicates that the readout occurred before the inflection
point of the blood signal due to a high heart rate (short RR duration and TI).
The polarity corrected (Fig. 3b) images show a dark blood pool and restored image
contrast.
In vivo MBF maps show visually
homogeneous values throughout the myocardium, with comparable physiological
noise between bSSFP and spGRE (Fig. 4).
Mean septal MBF per subject ranged between 1.25/2.53 and 3.48/5.70 ml/g/min
(bSSFP/spGRE, Fig. 5), with a reduced number of outliers using phase-sensitive
reconstruction in some subjects.Discussion
In this work we evaluate a
free-breathing myocardial ASL sequence, using navigator gating and
phase-sensitive reconstruction. Homogeneous MBF map quality is achieved during
free-breathing. Phase-sensitive reconstruction is shown to reduce the number of
outliers in some subjects.
As a result of short RR
intervals, i.e. short TIs, the magnetization can be negative at the time of the
image readout depending on the tissue T1. The distorted image contrast can then
lead to inaccurate MBF quantification. Phase-sensitive reconstructions were
shown to mitigate this effect in phantom. A
residual phantom MBF deviation was observed due to the impact of the image
readout. Phantom as well as in vivo MBF matched previously reported perfusion
values2,3,4. Respiratory navigation generally yielded shorter scan times than
breath-held acquisitions, while MBF maps were comparable for the two breathing
strategies. Our in vivo data further suggest that the use of phase-sensitive
reconstruction may reduce the number of outliers obtained during the MBF
measurement.Conclusions
Phase-sensitive
perfusion reconstruction restores image contrast and may result in more precise
MBF values. Free-breathing myoASL was demonstrated with dual-respiratory
navigation and double ECG-gating
and yields perfusion values comparable to breath-held myoASL. This approach can
enable faster contrast-agent free perfusion mapping with improved patient
comfort.Acknowledgements
M.B.I. acknowledges is funded by a PhD scholarship from the Landesgraduiertenförderung Baden-Württemberg, and a Procope Mobility stipend. S.W. acknowledges funding from the 4TU Precision Medicine program, an NWO Start-up STU.019.024, and ZonMW OffRoad 04510011910073.References
1. Gerber,
B. L., Raman, S. V., Nayak, K., Epstein, F. H., Ferreira, P., Axel, L., &
Kraitchman, D. L. (2008). Myocardial first-pass perfusion cardiovascular
magnetic resonance: history, theory, and current state of the art. Journal of Cardiovascular Magnetic Resonance,
10(1), 1-18.
2. Zun, Z., Wong, E. C., & Nayak, K. S. (2009). Assessment of
myocardial blood flow (MBF) in humans using arterial spin labeling (ASL):
feasibility and noise analysis. Magnetic Resonance in Medicine: An Official Journal of the
International Society for Magnetic Resonance in Medicine, 62(4), 975-983.
3. Do, H. P., Jao, T. R., & Nayak, K. S. (2014). Myocardial
arterial spin labeling perfusion imaging with improved sensitivity. Journal of Cardiovascular Magnetic
Resonance, 16(1), 1-6.
4. Aramendía‐Vidaurreta,
V., Gordaliza, P. M., Vidorreta, M., Echeverría‐Chasco, R., Bastarrika, G.,
Muñoz‐Barrutia, A., & Fernández‐Seara, M. A. (2022). Reduction of motion
effects in myocardial arterial spin labeling. Magnetic
Resonance in Medicine, 87(3),
1261-1275.
5. Božić-Iven,
M., Rapacchi, S., Pierce, I., Thornton, G., Tao, Q., Schad, L. R., Treibel, T.,
& Weingärtner, S. (2022). Towards reproducible Arterial Spin Labelling in
the myocardium: Impact of blood T1 time and imaging readout parameters. Proceedings of the Joint Annual Meeting ISMRM-ESMRMB.
6. Božić-Iven,
M., Rapacchi, S., Pierce, I., Thornton, G., Tao, Q., Schad, L. R., Treibel, T.,
& Weingärtner, S. (2022). Improving reproducibility of cardiac ASL using T1
and flip angle corrected reconstruction. Proceedings
of the Annual Meeting of the SMRA.
7. Messroghli, D. R., Radjenovic, A., Kozerke, S., Higgins, D. M.,
Sivananthan, M. U., & Ridgway, J. P. (2004). Modified Look‐Locker inversion
recovery (MOLLI) for high‐resolution T1 mapping of the heart. Magnetic Resonance in Medicine: An
Official Journal of the International Society for Magnetic Resonance in
Medicine, 52(1), 141-146.
8. Tao, Q., van der Tol, P., Berendsen, F. F., Paiman, E. H., Lamb, H.
J., & Van Der Geest, R. J. (2018). Robust motion correction for myocardial
T1 and extracellular volume mapping by principle component analysis‐based
groupwise image registration. Journal of Magnetic Resonance Imaging, 47(5), 1397-1405.
9. Gowland,
P. A., & Leach, M. O. (1991). A simple method for the restoration of signal
polarity in multi‐image inversion recovery sequences for measuring T1. Magnetic resonance in medicine, 18(1), 224-231.
10. Buxton, R. B., Frank, L. R., Wong, E. C., Siewert, B., Warach, S.,
& Edelman, R. R. (1998). A general kinetic model for quantitative perfusion
imaging with arterial spin labeling. Magnetic resonance in medicine, 40(3), 383-396.