Keywords: Pulse Sequence Design, Non-Proton, High-Field MRI, Body, 23Na MRI, Field Mapping
23Na MRI is a promising imaging method, but suffers from lower SNR, hence from longer acquisition times and lower image resolution than 1H MRI. Especially in the torso, motion and magnetic field inhomogeneities impede quantitative analysis of 23Na concentrations. To tackle this, a new pulse sequence is presented that yields self-gated respiratory-sorted B1+ and B0 maps from a single measurement. B1+ mapping with the new sequence is less prone to motion artefacts, since k-space projections of two flip angles are acquired in an interleaved manner. This reduces the influence of varying respiration, which can lead to artefacts in conventional mapping.
1. G. Madelin and R.R. Regatte. Biomedical applications of sodium MRI in vivo. Journal of Magnetic
Resonance Imaging, 38(3), 511-529
(2013)
2. M. A. Bernstein, K. F. King and X. Joe Zhou. Handbook of MRI Pulse Sequences (2004)
3. E.K. Insko and L. Bolinger. Mapping of the Radiofrequency Field. Journal of Magnetic Resonance. Series A 103, 82-85 (1993)
4. A. M. Nagel et al. Sodium MRI Using a Density-Adapted 3D Radial Acquisition Technique. Magnetic Resonance in Medicine 62:1565–1573 (2009)
5. R.W. Chan, E.A. Ramsay, C.H. Cunningham and D.B. Plewes. Temporal stability of adaptive 3D radial MRI using multidimensional golden means. Magnetic Resonance in Medicine 61: 354-363. https://doi.org/10.1002/mrm.21837 (2009)
6. T. Platt et al. In vivo self-gated 23Na MRI at 7T using an oval-shaped body resonator. Magnetic Resonance in Medicine 2018; 80:1005–1019 (2018)
7. MAGNETOM 7T, Siemens Healthcare, Erlangen, Germany
8. L. Leiner. Bachelor’s thesis - Planung und Umsetzung eines Phantoms für die MR-geführte Therapie. Heidelberg University (2017)
9. J. A. Fessler and B. P. Sutton. Nonuniform fast Fourier transforms using min-max interpolation. IEEE Transactions on Signal Processing. Vol. 51, no. 2, pp. 560-574. doi: 10.1109/TSP.2002.807005 (2003)
10. G. Yuan, N. A. Drost and A. McIvor. Respiratory rate and breathing pattern. McMaster Univ. Med. J. Vol. 10, no. 1, pp. 23–28 (2013)
Figure 1:
A) Density-adapted radial pulse sequence with alternating excitation by pulses with flip angle α and 2α, a double-echo readout and an ADC shift for all echos.
B) Schematic illustration of the new sequence (here: α =45°, 2α=90°) and three standard sequences used as reference. In the new sequence the same radial projection needed for determining B0 and B1+ maps is acquired within 300ms (thus likely to still be in the same or at least similar breathing state, as on average one ex- or inhalation each takes 2s10), whereas in the reference this time difference is much larger with 5min 45s.
Figure 2:
A) Excerpt from the graph of the phantom’s respiratory signal, which was simulated by motion of the small phantom by hand in the rhythm of the mover’s own respiratory frequency. The time of one breathing cycle (inhaling + exhaling) was on average 5s.
B) Excerpt from the graph of the in vivo respiratory signal. The time of one breathing cycle is in a range of 3s to 4.8s (on average 3.6s). The duration of the breathing cycles is in the range known from literature10.
Figure 3:
A) 23Na images (90°, TE1) of the stationary
respiratory phantom (reconstructed FOV=(360mm)3, zero filling
factor=2). The smaller movable phantom (high signal) was placed in breathing state 1. No motion was present during the acquisitions.
B) B0 and relative B1+ maps of the new sequence and its reference determined from the stationary MR images of breathing state 1 from A).
C) Quantitative comparison of determined maps (B0 distribution between -83Hz to 83Hz). Results for reference and the new sequence are in good agreement in the stationary case.
Figure 5:
Respiratory-sorted and unsorted B0 and relative B1+ maps (coronal plane) measured in a healthy volunteer (female, 32 years) using the new sequence and the reference (reconstructed FOV=(360mm)3, zero filling factor=2). For the new sequence the unsorted and sorted B1+ maps are similar, whereas the sorted B1+ maps of the reference slightly differ from the unsorted maps (white arrows).