Ehsan Kazemivalipour1,2, Lawrence L. Wald1,2,3, and Bastien Guerin1,2
1A.A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Harvard-MIT Division of Health Sciences Technology, Cambridge, MA, United States
Synopsis
We
investigate the performance of parallel transmission arrays with 8, 16, 24, and
32 channels and varying loop sizes (18 coils total) for brain imaging at 7T. We
compare RF-shimming performance of the arrays using the L-curve method showing
optimal tradeoffs between excitation uniformity (slice), local-SAR, global-SAR
and peak-power. For the numbers of channels simulated here, the coils with
larger loops achieved a better local-SAR vs. excitation uniformity than those
with smaller loops (this was true for axial/sagittal slice orientations). At
constant local-SAR constraint, the 24-channel arrays showed the best
performance in improving the excitation uniformity in the coronal/axial slices.
Introduction
Parallel transmission (pTx) allows control of
the magnetization and the E-field/specific absorption rate (SAR)1,2. This is
especially impactful at 7T, where pTx can be used to create uniform excitation
profiles and reduce local-SAR. The added transmit degrees of freedom (DoF)
provided by pTx pulses and coils depends in a complex manner on the number of
the transmit channels, the size of the channels, and their position with
respect to the head. Ideally, optimal operating points need to be compared for
different coil designs, which requires solving a pulse-design problem with explicit
constraints for the variables being assessed. Since we are interested in the
excitation uniformity and local-SAR, this means that local-SAR needs to be
constrained explicitly in the pulse-design problem. Here, we take a
practical approach to pTx coil optimization by electromagnetic simulation and
RF-shimming pulse-design of pTx coils with 8, 16, 24, and 32 channels with
varying loop diameters.Methods
We simulated the 18
pTx arrays for head imaging at 7T shown in Fig.1. For each coil, the loops
are placed on an elliptic former with a
hemi-ellipsoid top-end that approaches the shapes used in closed-fitting
helmets. The loop locations were optimized by
minimization of the maximum Euclidean distance between loop centers (feasible
since the former surface is an analytical formula). At constant number of
channels, we evaluated the impact of varying the loop diameter (all loops have
the same diameter in this study). The smallest loop diameter considered was 50mm
and the maximum was the one resulting in contact between nearest-neighbors (for
the 8-channel Dmax=70mm, 16-channel Dmax=46mm,
24-channel Dmax=40mm, and 32-channel Dmax=34mm). Each array was loaded with a detailed human
head model (with shoulders, 22 tissues3,4). Simulations and circuit co-simulations were performed with ANSYS Electronics
by replacing all sources and
capacitors with lumped ports. The multiport S-matrix output of HFSS was imported
into MATLAB, where we used custom code to find the tuning/matching capacitor
values for each coil design. Generation of the coil models, tuning, matching
and export of the field results was done in batch mode using MATLAB (HFSS
can be controlled externally by MATLAB), thus allowing to efficiently simulate
a large number of coil designs.
We extracted B1+-maps and E-fields
on a 191×121×171, 2mm isotropic image grid (we did not attempt decoupling of
the nearest-neighbors) and computed 10g-averaged SAR-matrices (Q-matrices)
for each voxel5,6, which we then compressed using the virtual observation points (VOPs)7. We designed
slice-selective RF-shimming pulses (target flip-angle of 10°) in central axial/coronal/sagittal brain slices. Each pulse was a 5-lobe sinc profile of 2ms
duration, duty-cycle of 100%, and slice thickness of 5mm. The pulses were
designed using least-squares optimization for a zero-target phase2,8
subject to a fixed peak-power constraint (56.25W/channel). L-curves were
obtained by varying local-SAR constraints (global-SAR was not constrained).Results
Fig.2A shows the S-matrices for the
tuned and matched coils (natural coupling). The matching level was better than
-35dB for all arrays. Increasing the number of array elements or the loop diameter
led to an increase in the maximum coupling. Greater coupling is associated with
greater eigenvalues of the SHS-matrix (power-matrix)9, indicating
greater reflected power of the eigenmode excitations (Fig.2B).
Figs.3-5 show the shim performance of the
array designs. For the 8-channel array (and, to some extent the 16-channel),
increasing the loops diameter improved the excitation uniformity vs. local-SAR
tradeoff for the transverse/sagittal designs. However, for the coronal
slice, increasing the loop diameter did not necessarily improve the imaging
performance, and, for the 8-channel coils, even created lower-quality
excitations at the lower local-SAR range of values studied.
Fig.5 shows that increasing the channel
count from 8 to 16 and then to 24 improved the pulse performance for the axial/coronal slices, while increasing the number of channels from 24 to 32 did
not always improve transmit performance further. In some cases, it even
worsened it, indicating that for some slice orientations, the 24-channel coil
has more effective DoFs than the 32-channel coil. This can be due
to the specifics of the coil DoFs projected on the slice of interest and to
increased coupling in the 32-channel coil that reduces the effective transmit
DoFs.Conclusion and Discussion
This simulation study shows that pTx coil
comparison depends on the pulse-design strategy and is highly non-linear, with
some L-curves crossing even when using a convex pulse-design strategy
(least-squares)8,10. As
expected, increasing the number of channels generally led to better transmit
performance (local-SAR vs. excitation uniformity) in the moderate-high local-SAR
regime, although this simple conclusion was not always true at lower SAR values
(16-channel sometimes outperform 24- and 32-channel). Increasing the
loop diameter (at constant number of channels) led to better transmit
performance for the 8-channel coil only, and only for the transverse/sagittal slice orientations (for the coronal slice orientation, the performance
initially increased with D, then decreased). In next steps, we will vary
individual loop diameters. Although more complex, this may actually lead to a
simpler ranking and conclusions since we would be in essence comparing optimal
coil configurations (or close to optimal) for varying number of channels. The MATLAB-based
simulation/tuning/matching/pulse design strategy used here will make it easier
to run through the hundreds of simulations required for that goal.Acknowledgements
No acknowledgement found.References
-
Katscher
U, Bornert P, Leussler C, et al. Transmit SENSE. Magn Reson Med.
2003;49(1):144-150.
- Zhu Y. Parallel excitation with an
array of transmit coils. Magn Reson Med. 2004;51(4):775-784.
- Massire A, Cloos MA, Luong M, et al.
Thermal simulations in the human head for high field MRI using parallel transmission.
J Magn Reson Imaging. 2012;35(6):1312-1321.
- Makris N, Angelone L, Tulloch S, et
al. MRI-based anatomical model of the human head for specific absorption rate
mapping. Med Biol Eng Comput. 2008;46(12):1239-1251.
- Sbrizzi A, Hoogduin H, Lagendijk JJ,
et al. Fast design of local N-gram-specific absorption rate-optimized
radiofrequency pulses for parallel transmit systems. Magn Reson Med.
2012;67(3):824-834.
- Das SK, Clegg ST, Samulski TV.
Computational techniques for fast hyperthermia temperature optimization. Med
Phys. 1999;26(2):319-328.
- Eichfelder G, Gebhardt M. Local
specific absorption rate control for parallel transmission by virtual
observation points. Magn Reson Med. 2011;66(5):1468-1476.
- Guerin B, Gebhardt M, Cauley S, et
al. Local specific absorption rate (SAR), global SAR, transmitter power, and
excitation accuracy trade-offs in low flip-angle parallel transmit pulse
design. Magn Reson Med. 2014;71(4):1446-1457.
- Kazemivalipour E, Sadeghi-Tarakameh
A, Atalar E. Eigenmode analysis of the scattering matrix for the design of MRI
transmit array coils. Magn Reson Med. 2021;85(3):1727-1741.
- Wu
X, Zhang X, Tian J, et al. Comparison of RF body coils for MRI at 3 T: a simulation
study using parallel transmission on various anatomical targets. NMR Biomed.
2015;28(10):1332-1344.