Sri Kirthi Kandala1, Alberto Fuentes2, Gregory H Turner2, and Sung-Min Sohn1
1Arizona State University, Tempe, AZ, United States, 2Barrow-ASU Center for Preclinical Imaging, Phoenix, AZ, United States
Synopsis
Like most modern
electronic devices, a mobile application based Radio Frequency (RF) coil
monitoring system will make it easy to carry out any real-time changes to the
RF coil. This work focuses on developing a mobile application based RF
monitoring system, that allows the user to gather real-time RF coil characteristics
and control the coil. It gives the user a chance to make changes when necessary.
The opportunity to wirelessly implement such changes without physical contact
with the coil, using a mobile application via Bluetooth is a step in the
direction towards building next-generation RF coils for MRI.
Introduction
With MR scanners becoming
increasingly complex, most of the scanners have a built-in and proprietary
monitoring system for Radio Frequency (RF) coils and interface circuits. These
integrated RF monitoring systems have very little options to incorporate an RF
coil built in-house. Here arises a need for a mobile application based, open
source RF monitoring system that does not disturb the transmit (Tx) and receive
(Rx) chain of the MR scanner. Previous studies have delved slightly into
developing a real time RF monitoring system but none of them were close enough
to a standalone wireless module. This study involves design and development of a
stand-alone, wireless RF monitoring system that presents features like wireless
impedance control to tune and match, forward and reflected RF power
measurement, and decoupling among coil elements along with much potential for
future applications. Method
An L-matching network
with single series and shunt capacitors (trimmer) was designed along with MEMS
(ADGM1304, Analog Devices, USA) switch and 8 capacitors (0505 series
Non-magnetic, PassivePlus, USA) with fixed capacitances to create a capacitor
bank as shown in Fig.1(Bottom). This matching network board can generate a
combined of 255 states for tuning and matching. A directional coupler is used
to couple a portion of the reflected
power using a logarithmic power detector (AD8307, Analog Devices, USA) and fed
to a microcontroller (Espressif Systems, CHINA) which also controls the MEMS
switches. On the user end, an Android application using the
MIT App inventor has been developed, to control the RF monitoring system as shown in Fig.2. After
pairing the module to the application via Classic Bluetooth connection, user
has the option to enter the input power in dBm. This allows the user to get a glimpse
of the return loss in real-time as they change the capacitance values. The
mobile application allows the user to control the impedance with the help of
two sliders to change the capacitance value of the tuning and matching
capacitors and an AUTO T/M button to carry out a fully automatic impedance
matching function. While changing the slider bar, reflected power from the
power detector will be displayed on the canvas as a plot ranging from 0-1023.
The lowest number will correspond to tuned and matched state for the RF coil.Results and Discussion
A surface coil made out
of copper wire and a two-port VNA (FieldFox N9923A, Keysight Technologies, USA)
is used to gather the S-parameters and validate the performance of the system
on the test bench. Fig.3(Left) shows the tuned and detuned conditions of the
coil while using the Android application to control the MEMS switches. All the
MR imaging studies were conducted at Barrow Neurological Institute - Arizona
State University (BNI-ASU), Center for Preclinical Imaging, using a 7T
small-animal, 30-cm horizontal-bore magnet and BioSpec Avance III spectrometer
(Bruker, Billerica, MA) with a 116-mm high-power gradient set (600 mT/m).
Fast-Low-Angle-Shot (FLASH) Sequence with a repetition time (TR) of 350 ms,
echo time (TE) of 5.4 ms, and a flip angle (α) of 40 degrees was used for MR
imaging experiments. A field of view (FOV) of 40 x 40 mm and 256 x 256 matrix
leading to an in-plane resolution of 156 x 156 µm, in addition 9 slices were
acquired along the sample with a slice thickness of 1 mm. A saline solution
filled in a 50 mL centrifuge tube is used as a phantom to replicate small
animals. Same slices (8/9) of the MR images for both tuned and detuned
condition were selected to compare the image quality. Conclusion
This work shows the
feasibility for a general purpose RF monitoring module that includes wireless
impedance control and the ability to detune (Rx coils only) . The mobile application
also gives us the opportunity to view real time reflected power plot. This
feature is helpful when MRI consoles do not possess the ability to plot return
loss. This prototype will be further developed to a multichannel, high power capable
module which will assist in building a stand-alone and open source RF
monitoring system for MRI scanners.Acknowledgements
This work was supported by the National Institute Of Biomedical Imaging And Bioengineering of the National Institutes of Health under Award Number R00EB020058. References
1.
Sohn, S. M., DelaBarre, L., Gopinath, A., & Vaughan, J.
T. (2015). Design of an Electrically Automated RF Transceiver Head Coil in MRI.
IEEE transactions on biomedical circuits and systems, 9(5), 725–732.
https://doi.org/10.1109/TBCAS.2014.2360383
2.
El-Sharkawy, A. M. M., Qian, D., Bottomley, P. A., &
Edelstein, W. A. (2012). A multichannel, real-time MRI RF power monitor for
independent SAR determination. Medical Physics, 39(5), 2334–2341.
https://doi.org/10.1118/1.3700169