Jeehun Kim1,2,3, Qi Peng4, Can Wu5, and Xiaojuan Li1,2,6
1Program of Advanced Musculoskeletal Imaging (PAMI), Cleveland Clinic, Cleveland, OH, United States, 2Department of Biomedical Engineering, Cleveland Clinic, Cleveland, OH, United States, 3Department of Electrical Engineering, Case Western Reserve University, Cleveland, OH, United States, 4Department of Radiology, Albert Einstein College of Medicine and Montefiore Medical Center, Bronx, NY, United States, 5Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 6Department of Diagnostic Radiology, Cleveland Clinic, Cleveland, OH, United States
Synopsis
Quantitative T1ρ mapping is a promising biomarker for
detecting tissue compositional changes at early stages of diseases. For
reliable and reproducible measurements, T1ρ preparation pulses should be robust
to B0 and B1 inhomogeneity. In this work, a Bloch equation based numerical
simulation tool was developed and validated. The performance of six different T1ρ
preparation schemes were evaluated in terms of their responses to B0 and B1
inhomogeneities using the simulation tool and in phantoms and human subjects. T2ρ
measured in phantoms and human subjects were used during simulation.
Introduction
Quantitative T1ρ mapping is a promising technique in detecting tissue compositional changes
at the early stages of diseases such as Osteoarthritis.1 For this
method to be successfully implemented in clinical routines, the imaging
technique should be stable and robust, including responses to B0 and B1
inhomogeneities. This is especially important in body parts that are
off-center, and in situations such as with metal-induced inhomogeneities due to
surgeries. In the previous study, we compared 6 different preparation schemes
in phantoms and in-vivo acquisitions.2 However, a numerical simulation was needed to
better quantify the response to B0/B1 inhomogeneities in larger ranges. Furthermore,
T2ρ needs to be
considered during simulation. However, T2ρ measures are scarce and inconsistent
(regarding its relationship to T1ρ values) in the literature.3,4 Therefore, the
aim of this study was to investigate 6 T1ρ preparation schemes using Bloch equation-based
numerical simulation, validate it by phantom acquisition, and evaluate the
performance in in-vivo acquisitions. T2ρ values were measured and used in
both phantoms and in-vivo.Methods
Bloch equation-based simulation
A time-series Bloch equation-based simulation was developed
for the 6 preparation schemes (Figure
1).
The simulator follows the RF profile created for each preparation with 0.1us
time steps (dT).2,5-8
When creating RF profile, B1 inhomogeneity was translated as a constant applied
on RF field, and B0 inhomogeneity was translated as an B1 in z-axis (B0
direction). For all non-spinlock pulses, the RF strength was equivalent
to 400us 90 degrees pulse, which matches how the preparations were implemented
in the MR scanner. During a non-spinlock RF pulse, the change in magnetization
(dM) was calculated by Bloch equation. During a spinlock RF pulse, dM was
calculated with a slightly modified Bloch equation, where T1/T2 were changed to
T1ρ/T2ρ, and T1 recovery term removed. 500Hz
spinlock frequency was used in both simulation and acquisition.
Evaluation of T1ρ preparation schemes with simulation
The simulation was used evaluate the performance of
preparations in various B0 and B1 conditions. First, the decay curve with
regards to Spinlock time (TSL=0-70ms) was generated with B0/B1 inhomogeneity of
-100~+100Hz/0.8~1.2. In the curve, T1ρ/T2ρ were
set to 40ms. Second, a heatmap of preparation error was drawn from B0/B1
inhomogeneity of -200~+200Hz/0.8~1.2. In the heatmap, T1ρ was locked in 40ms while changing T2ρ (40-70ms). Area under 5% quantification
error was calculated by deriving the ratio of the area that has smaller than 5%
T1ρ
difference to evaluate
the robustness of preparation regarding B0/B1 inhomogeneity.
Phantom Imaging and Simulation Validation
A 3% agarose phantom was imaged using a 3T Prisma MR scanner
(Siemens
Healthcare AG, Erlangen, Germany) with 1Tx/15Rx knee coil (QED, Mayfield, OH). 3D
magnetization-prepared angle-modulated partitioned k-space spoiled gradient-echo
snapshots (MAPSS) sequence9,10 was used to
quantify T2ρ and
T1ρ with
different preparations. TSLs of 0,10,20,30ms were used. During acquisition, a 300uT/m
gradient was used to apply ±250Hz inhomogeneity in 4cm range. The B0 was
measured by double-echo gradient-echo sequence, and B1 was measured by
Bloch-Siegert shift B1 mapping.11 To validate the simulation,
B0/B1 maps and reference T1ρ/T2ρ values of
phantom acquisition (43/70ms for T1ρ/T2ρ as measured in the phantom without B0/B1 offset)
were used. The calculated T1ρ values from the simulation were compared with phantom acquisition values
at all positions.
Human Subject Imaging
A healthy volunteer was scanned with all T1ρ preparations and T2ρ. TSL of 0,10,30,70ms was used, and an
additional dual-echo steady-state image was collected for cartilage
segmentation using Deep Learning-based automatic segmentation.12 Six compartments were segmented, and average
T1ρ/T2ρ were reported.Results
Figure 2 shows the T1ρ curves generated by simulation. The oscillation in
the curve increases with increased B0/B1 inhomogeneities, and these
oscillations result in banding artifacts in the acquisition. Among the preparations, Prep6 showed the
least amount of oscillation.
Figure 3 shows the heatmap of T1ρ quantification error with different
B0/B1 inhomogeneities. The trend changed with different combinations of T1ρ/T2ρ, and in this setup, Prep6 showed the best result
when T2ρ was smaller
than 65ms.
Figure 4 shows the validation of simulation to the
acquisition. It clearly shows the simulation reflects the actual acquisition,
verifying that the simulation can be used to evaluate different preparations.
Figure 5 shows the T1ρ/T2ρ values acquired from a healthy volunteer. Even with
the change in preparation methods, the quantitative values were close in
general. Also, T2ρ was longer by 10ms than T1ρ in general.Discussion
We have developed and validated a simulation tool for evaluating
responses of T1ρ
preparation schemes to B0/B1 inhomogeneity. Such tools will be helpful for
future continuous optimization of T1ρ preparation design. According to the simulation, Prep6 showed the
least oscillation in signal curve with B0/B1 offset. This characteristic does
not only decrease the banding artifact easily seen in areas with large
inhomogeneity but also decreases the variability of T1ρ quantification caused
by different choices of TSLs. Also, Prep6 was most robust to B0/B1 inhomogeneity
unless T2ρ was 30ms longer than T1ρ. According to the in-vivo measurement, T2ρ
was different by 10ms. Therefore, we expect Prep6 to show the best performance
in in-vivo. The sequence will be evaluated in more human subjects for responses
to B0/B1 inhomogeneity, scan-rescan repeatability, and inter-vendor inter-site
variations in future studies.Acknowledgements
The study was supported by NIH/NIAMS R01 AR077452References
1. MacKay JW, Low SBL, Smith TO, et al. Systematic review and meta-analysis of the
reliability and discriminative validity of cartilage compositional MRI in knee
osteoarthritis. Osteoarthritis and cartilage. 2018;26(9):1140-52
2. Kim J, Peng Q, Wu C, et al. MR T1rho
preparations: B1 and B0 inhomogeneity response on 3T and 7T systems.
International Society for Magnetic Resonance in Medicine. 2021
3. Wheaton AJ, Borthakur A, Kneeland
JB, et al. In vivo quantification of T1rho using a
multislice spin-lock pulse sequence. Magnetic resonance in medicine.
2004;52(6):1453-8
4. Casula V, Nissi MJ, Podlipská J,
et al. Elevated adiabatic T(1ρ) and T(2ρ) in
articular cartilage are associated with cartilage and bone lesions in early
osteoarthritis: A preliminary study. Journal of magnetic resonance imaging :
JMRI. 2017;46(3):678-89
5. Chen W. Errors in quantitative T1rho
imaging and the correction methods. Quantitative imaging in medicine and
surgery. 2015;5(4):583-91
6. Pang Y. A self-compensated
spin-locking scheme for quantitative R1ρ dispersion in articular cartilage.
International Society for Magnetic Resonance in Medicine. 2020
7. Gram M, Seethaler M, Gensler D, et al. Balanced spin-lock preparation for B(1)
-insensitive and B(0) -insensitive quantification of the rotating frame relaxation
time T(1ρ). Magnetic resonance in medicine. 2021;85(5):2771-80
8. Mitrea BG, Krafft AJ, Song R, et al. Paired self-compensated spin-lock preparation for
improved T1ρ quantification. Journal of magnetic resonance (San Diego, Calif :
1997). 2016;268:49-57
9. Li X, Han ET, Busse RF, et al.
In vivo T(1rho) mapping in cartilage using 3D magnetization-prepared
angle-modulated partitioned k-space spoiled gradient echo snapshots (3D MAPSS).
Magnetic resonance in medicine. 2008;59(2):298-307
10. Li X, Pedoia V, Kumar D, et al. Cartilage T1rho and T2 relaxation times:
longitudinal reproducibility and variations using different coils, MR systems
and sites. Osteoarthritis and cartilage. 2015;23(12):2214-23
11. Sacolick LI, Wiesinger F, Hancu I, et al. B1 mapping by Bloch-Siegert shift. Magnetic resonance in medicine.
2010;63(5):1315-22
12. Gaj S, Yang M, Nakamura K, et al.
Automated cartilage and meniscus segmentation of knee MRI with conditional
generative adversarial networks. Magnetic resonance in medicine.
2020;84(1):437-49