Bhumi Bhusal1, Fuchang Jiang1, Pia Panravi Sanpitak1, Boris Keil2, Gurinder Kaur Multani2, Nicolas Kutscha2, Giorgio Bonmassar3, Gregory Webster1, Andrada Popescu4, Daniel Kim1, and Laleh Golestanirad1
1Northwestern University, Chicago, IL, United States, 2TH Mittelhessen University of Applied Sciences, Giessen, Germany, 3Massachusetts General Hospital, Charlestown, MA, United States, 4Lurie Children's Hospital, Chicago, IL, United States
Synopsis
Infants
and children with congenital heart disease may require cardiovascular
implantable electronic devices (CIEDs), which typically
precludes future evaluation by MRI due to risks associated with RF
heating of implants. Here we demonstrate that a novel concept based on
patient-adjustable rotating RF coil technology that was originally developed to
reduce RF heating of brain implants can be adopted in pediatric patients with
CIEDs to substantially reduce RF heating.
Our simulations in a cohort of children with both epicardial and
endocardial devices showed an 87% reduction in average SAR compared to
conventional body coils.
Introduction
Infants
and children with congenital heart disease (CHD) often require cardiovascular implantable
electronic devices (CIED)1. Some infants receive CIED
within hours, or even minutes, of birth2. Unfortunately, once CIEDs
have been implanted, there is a relative contraindication for MRI exam due to
risks of RF heating of the device. Recently,
an RF coil technology based on a patient-adjustable rotating transmit coil was shown
to reduce RF heating of neurological implants in adult patients3-6. Here we show that this novel concept can be
adopted in children with CIEDs. We designed and constructed a linearly
polarized (LP) mechanically rotating birdcage transmitter for MRI of children
(0-3 y/o) in a 1.5 T scanner. The coil is a 16-rung low-pass birdcage and has a
slab-like region of very low electric field which can be steered by
mechanically rotating the coil around patient’s body. Once the coil is
positioned at its optimum angle for an individual patient, i.e., the angle at
which patient’s implanted device is maximally contained within the low E field region of the coil, the local SAR amplification around the implanted lead
tip is virtually eliminated. The
SAR-reduction performance of the coil was assessed in electromagnetic simulations
in 10 patient-derived realistic models of CIEDs with both endocardial and
epicardial leads. The mean reduction in the 0.1g-averaged SAR was 87 % (> 7
fold). A prototype of the rotating coil is built and will be tested on a
commercial CIED (Medtronic Azure XT DR) implanted in an anthropomorphic
pediatric phantom to validate simulation results.Methods
Rotating coil: A linearly polarized 16-rung lowpass birdcage coil was
constructed on a 350 mm diameter acrylic cylinder (Figure 1). The coil’s ladder network was constructed from 12 mm
adhesive copper tape (Chomerics Inc., Woburn,MA,USA) and had a rung length of
290 mm. Two
mechanical annuli were secured to the birdcage ends and rested on four guide
wheels located on a sliding frame, allowing the birdcage to rotate freely
around its axis without touching the load.
Simulations:
A child body model was created from segmented MRI images of a 29-months-old
child7. Segmented images
were manually smoothed and closed 3D surfaces representing average tissue,
skull, brain, and heart were created in a CAD tool (Rhino 7, Robert McNeel and
Associates, Seattle). Ten unique CIED lead trajectories consisting of 5
epicardial and 5 endocardial cases were created based on X-ray images of
pediatric patients with CIEDs and incorporated in the body model (Figure 2).
Electromagnetic simulations were
implemented in ANSYS Electronic Desktop 2021 R1 HFSS (ANSYS Inc., Canonsburg,
PA). Model of a local birdcage transmit coil with dimensions matching the
constructed prototype was created and tuned to 64 MHz. The coil was fed with a
single cable connected to the bottom end ring which generated a linearly
polarized B1 field and
an E field with a slab-like region of low intensity, which was then
steered by rotating the coil around the body at 22.5°
increments (16 positions in total) (Figure 3). The simulations were
repeated with a conventional body coil (50 cm length, 71 cm diameter) in
circular polarization (CP). Coil SAR-reduction performance was quantified by a
SAR Reduction Efficiency (SRE) metric calculated as
$$$%$$$SREn(φ) $$$=$$$ $$$100×$$${Max0.1gSARCP,n $$$-$$$Max0.1gSAR(φ)LP,n}$$$/$$$Max0.1gSARCP,n
where
Max0.1gSARCP,n was the maximum of 0.1g-averaged SAR at the
tip of the CIED lead 'n' produced by a CP body coil whereas Max0.1gSAR(φ)LP,n was the maximum of 0.1g-averaged SAR produced
by the rotating LP coil at rotational angle $$$φ ε [0° - 360°]$$$. Results
SAR reduction performance:
For all device configurations, we were able to find an optimum rotation angle
that reduced the SAR at tips of implanted leads well below SAR levels produced
by a CP body coil, even when the input power of both coils were adjusted to
produce the same B1+ (2μT) at the iso-center (Figure 4 & 5). The SRE ranged from
58% to 98% with an average of 87% which is equivalent to > 7-fold reduction in SAR. The position of
the optimum angle could be reliably determined from B1+ maps as they corresponded to a minimized
RF-induced B1-inhomogeneity around the lead (Figure 3).
Sensitivity to operational error:
The safety margin of the coil was
quantified with a metric of Permissible Rotational Error (PRE) defined as
PREn $$$=$$$|φn,opt$$$-$$$φn,CP|
where
φn,opt is the optimum angle that minimizes the SAR for
lead 'n',
and φn,cp is the the nearest angle from
optimum angle at which the SAR due to rotating coil equals
that of CP body coil. PRE
was found to be 25° ± 7.9°, which is substantially higher than the rotational
resolution of the coil (5°), indicating that risks due to operational error could
be kept at minimum.Discussions and Conclusions
Application
of a new coil technology for full body MRI of infants and children (<3 y/o)
with CIEDs was demonstrated. The constructed local transmit coil fits inside a 50-cm
bore magnet and can be rotated smoothly around the child’s body to an optimum
angle that virtually eliminates implant RF heating. Position of the optimum
angle can be reliably determined by minimizing the B1-inhomogeneity around the lead on B1+ maps using a
low-SAR B1 mapping sequence.Acknowledgements
No acknowledgement found.References
1. Epstein A. American College of
Cardiology/American Heart Association Task Force on Practice Guidelines
(Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for
Implantation of Cardiac Pacemakers and Antiarrhythmia Devices); American
Association for Thoracic Surgery; Society of Thoracic Surgeons. Circulation.
2008;117:e350-e408.
2. Taşar M, Yaman ND, Arıcı B, et al. Minimally invasive permanent pacemaker implantation
immediately after birth: from delivery room to heart surgery: Permanent
Pacemaker Implantation in First Month of Life. 2021:1-3.
3. Golestanirad L, Keil B, Angelone LM, et al. Feasibility of using linearly polarized
rotating birdcage transmitters and close‐fitting receive arrays in MRI to
reduce SAR in the vicinity of deep brain simulation implants. Magnetic
resonance in medicine. 2017;77(4):1701-12.
4. Kazemivalipour E, Keil B, Vali A, et al. Reconfigurable MRI technology for low-SAR
imaging of deep brain stimulation at 3T: Application in bilateral leads,
fully-implanted systems, and surgically modified lead trajectories. NeuroImage.
2019;199:18-29.
5. Bhusal B, Keil B, Rosenow J, et al. Physics in Medicine and Biology. Patient’s body composition can
significantly affect RF power deposition in the tissue around DBS implants:
ramifications for lead management strategies and MRI field-shaping techniques.
2021;66(1):015008.
6. Golestanirad L, Iacono MI, Keil B,
et al. Construction and modeling of a
reconfigurable MRI coil for lowering SAR in patients with deep brain
stimulation implants. Neuroimage. 2017;147:577-88.
7. Jeong
H, Ntolkeras G, Alhilani M, et al. Development,
validation, and pilot MRI safety study of a high-resolution, open source, whole
body pediatric numerical simulation model. 2021;16(1):e0241682.