Zhehao Hu1,2, Nan Wang3, Anthony Christodoulou2,3, Mark Shiroishi1, Gabriel Zada4, Debiao Li2,3, and Zhaoyang Fan1,5,6
1Department of Radiology, Keck School of Medicine, University of Southern California, Los Angeles, CA, United States, 2Department of Bioengineering, University of California, Los Angeles, Los Angeles, CA, United States, 3Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, CA, United States, 4Department of Neurosurgery, Keck School of Medicine, University of Southern California, Los Angeles, CA, United States, 5Department of Radiation Oncology, Keck School of Medicine, University of Southern California, Los Angeles, CA, United States, 6Department of Biomedical Engineering, University of Southern California, Los Angeles, CA, United States
Synopsis
MR Multitasking-based Dynamic Imaging for Cerebrovascular Evaluation
(MT-DICE) technique is recently developed and has the potential to provide DCE- and leakage-corrected DSC-MRI parameters
simultaneously with one 7.6-minute scan and a single-dose contrast injection. However, it remains a practical challenge to validate the technique against their
respective references with repeated contrast injections in separate days. In
this work, we perform a numerical simulation study to validate the accuracy of
MT-DICE in the estimation of permeability and leakage-corrected perfusion
metrics.
Introduction
DSC-MRI and DCE-MRI provide perfusion- and permeability-related
parameters, respectively, and are evolving as increasingly common modalities
for evaluating a variety of brain cancers1,2. Their different but
complementary information may form a more complete basis for evaluation of the
complex and heterogeneous tumor microenvironment. However, acquiring both in
one exam requires two separate scans as well as two contrast injections. Recently,
we developed an MR Multitasking-based Dynamic Imaging for Cerebrovascular
Evaluation (MT-DICE) technique that provides DCE- and leakage-corrected DSC-MRI
parameters simultaneously with one 7.6-minute scan and a single-dose contrast
injection3,4. It remains a practical challenge to validate the
technique against their respective references with repeated contrast injections
in separate days. In this work, we perform a numerical simulation study to validate
the accuracy of MT-DICE in the estimation of permeability and leakage-corrected
perfusion metrics.Methods
To better simulate the highly heterogeneous environment within brain
tumors, a 3D anthropomorphic digital reference brain phantom incorporating a
tumor model from a deidentified glioblastoma patient was created5-7.
The $$${T_{1}}$$$-/$$${T_{2}^{*}}$$$-based arterial input functions (AIFs) were
generated according to Jaspers et al.8 and Simpson et al.9,
respectively, with realistic parameters at a 0.1-s temporal resolution. The
dynamic $$${T_{1}}$$$/$$${T_{2}^{*}}$$$ curves
were generated for gray matter (GM), white matter (WM) and tumor with the
parameters listed in Table 1, and the residual function $$$R(t)$$$ was
modeled as:$$R(t)=exp(-CBF\cdot t/CBV)$$where $$$CBF$$$ represents cerebral blood flow and $$$CBV$$$ represents
cerebral blood volume. Subsequently, the simulated dynamic signal intensities
were calculated based on the signal equation, which depends on our sequence
design, using the downsampled dynamic $$${T_{1}}$$$/$$${T_{2}^{*}}$$$ curves at a temporal resolution of 1 s, which is consistent with the
temporal resolution used in our in vivo study4. The simulated images were Fourier transformed with complex white
Gaussian noise added to achieve a signal-to-noise ratio of 30. The generated
k-space data were first undersampled and then reconstructed using the MT-DICE
technique (Figure 1). Specifically, MT-DICE models the 6-dimensional image $$$A(x,y,z,\tau,{T_{E}},t)$$$ as a low-rank tensor $$$\mathfrak{A}=\phi \times_{1} \mathbf{U_{r}}$$$. The temporal factor tensor $$$\phi$$$ is first determined from the training data and the spatial coefficients $$$\mathbf{U_{r}}$$$ are reconstructed by fitting $$$\phi$$$ to the imaging data $$$\mathbf{d}$$$, with undersampling pattern $$$\Omega$$$, spatial encoding model $$$\mathbf{E}$$$ and regularization parameter $$$\lambda$$$ for spatial total variation penalty $$$TV(\cdot)$$$. $$\hat{\textbf{U}}_\textbf{r}=\arg \min \left \| \textbf{d}-\Omega (\phi \times_{1} \textbf{E}\textbf{U}_{\textbf{r}}) \right \|_{2}^{2}+\lambda TV(\textbf{U}_{\textbf{r}})$$Dynamic $$${T_{1}}$$$/$$${T_{2}^{*}}$$$ fitting
and kinetic modeling were performed on all slices involving the tumor. The
derived permeability metrics were adopted to perform leakage correction for the
estimations of DSC-MRI metrics based on a combined biophysical
and pharmacokinetic approach10. In
addition to the leakage-corrected perfusion parameters, the
non-leakage-corrected perfusion metrics were derived in the conventional way11.Results
The simulated dynamic image series of one representative slice from the
digital brain phantom are displayed in Figure 2A. The ground-truth and MT-DICE
derived maps of DCE-MRI and DSC-MRI parameters as well as their differences are
shown in Figure 2B. MT-DICE was capable of correcting contrast leakage effects,
which is more evident in $$$CBV$$$ quantification, thus leading to smaller
percentage errors compared with the non-leakage-corrected counterpart (13.99%
vs. 18.87%). Table 2 summarizes the quantitative results measured from the
entire tumor model except the necrotic core.Discussion and Conclusion
In this work, a simulation study was performed to validate the accuracy
of MT-DICE in the estimation of permeability and leakage-corrected perfusion
metrics. In the field of DCE-MRI and DSC-MRI, the common method of assessing
bias and variance resulting from an algorithm is through the development and
application of a grid-based digital reference objects (DRO). To better simulate
the highly heterogeneous environment within brain tumors, a 3D anthropomorphic
digital reference brain phantom incorporating a tumor model was adopted, which
is more sophisticated and accurate. The simulation results validated the
accuracy of the proposed MT-DICE in kinetic parameter estimation.Acknowledgements
No acknowledgement found.References
1. Essig M, Nguyen TB,
Shiroishi MS, Saake M, Provenzale JM, Enterline DS, Anzalone N, Dörfler A,
Rovira À, Wintermark M. Perfusion MRI: the five most frequently asked clinical
questions. American Journal of Roentgenology 2013;201(3):W495-W510.
2. Bergamino M, Bonzano
L, Levrero F, Mancardi G, Roccatagliata L. A review of technical aspects of
T1-weighted dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) in
human brain tumors. Physica Medica 2014;30(6):635-643.
3. Hu Z, Christodoulou
A, Wang N, Xie Y, Sun B, Bi X, Han F, Song S, Maya M, Li D, Fan Z. MR
multitasking-based dynamic imaging for cerebrovascular evaluation (MT-DICE):
development and feasibility study. Proceeding International Society of Magnetic
Resonance in Medicine 2020; 1827.
4. Hu Z, Christodoulou
A, Wang N, Xie Y, Cao T, Maya M, Yang W, Li D, Fan Z. MR multitasking-based
dynamic imaging for cerebrovascular evaluation (MT-DICE): further development
and feasibility study on brain cancer. Proceeding International Society of
Magnetic Resonance in Medicine 2021; 0048.
5. Bosca RJ, Jackson
EF. Creating an anthropomorphic digital MR phantom—an extensible tool for
comparing and evaluating quantitative imaging algorithms. Physics in Medicine
& Biology 2016;61(2):974.
6. Cocosco CA,
Kollokian V, Kwan RK-S, Pike GB, Evans AC. Brainweb: Online interface to a 3D
MRI simulated brain database. 1997. Citeseer.
7. Collins DL,
Zijdenbos AP, Kollokian V, Sled JG, Kabani NJ, Holmes CJ, Evans AC. Design and
construction of a realistic digital brain phantom. IEEE transactions on medical
imaging 1998;17(3):463-468.
8. Jaspers K, Aerts HJ,
Leiner T, Oostendorp M, van Riel NA, Post MJ, Backes WH. Reliability of
pharmacokinetic parameters: Small vs. medium‐sized
contrast agents. Magnetic Resonance in Medicine: An Official Journal of the
International Society for Magnetic Resonance in Medicine 2009;62(3):779-787.
9. Simpson NE, He Z,
Evelhoch JL. Deuterium NMR tissue perfusion measurements using the tracer
uptake approach: I. Optimization of methods. Magnetic Resonance in Medicine: An
Official Journal of the International Society for Magnetic Resonance in
Medicine 1999;42(1):42-52.
10. Stokes AM, Semmineh
N, Quarles CC. Validation of a T1 and leakage correction method based on
multiecho dynamic susceptibility contrast MRI using MION as a reference
standard. Magnetic resonance in medicine 2016;76(2):613-625.
11.
Straka M, Albers GW, Bammer R. Real‐time
diffusion‐perfusion
mismatch analysis in acute stroke. Journal of Magnetic Resonance Imaging 2010;32(5):1024-1037.