Michael Childers1, Dorab Sethna1, Vick Chen1, and Shiloh Sison2
1Abbott, Sylmar, CA, United States, 2Abbott, Sunnyvale, CA, United States
Synopsis
Simulations and benchtop testing were performed to assess potential
gradient induced device heating of leadless pacemakers. Results of the evaluation
showed leadless pacemakers sized objects, made from commonly used conductive
materials, heated less than 1°C when exposed to the gradient induced device heating
safety test conditions per ISO/TS 10974.
In general, a temperature rise less than 2°C is considered to be
clinically insignificant. As such, the risk of leadless pacemaker gradient
induced device heating is negligible and assessments for this hazard per ISO/TS
10974 may not be necessary.
Introduction
For the assessment of MR
conditional safety there are many hazards to consider. Gradient induced device
heating is one hazard identified in ISO/TS 10974. The imaging gradient dB/dt field can induce eddy currents on the conductive surfaces of
an of Active Implantable Medical Device (AIMD) resulting in heating. The instantaneous
power deposited by a dB/dt field into a conductive material is proportional to its
size1, where larger objects of the same material will have higher
heating and smaller objects will have less heating.
As leadless pacemakers are
relatively small AIMDs, we explore in this abstract the potential gradient
induced heating for this type of device by modeling and conducting benchtop
testing on leadless pacemaker sized objects made from commonly used conductive materials.
The results from this modeling and
testing are then evaluated to see if the temperature rises could result in
clinically significant heating. Methods
To assess potential gradient
induced heating of leadless pacemakers, simulations were performed using COMSOL
Multiphysics software (v5.6). These
simulations were conducted on leadless pacemaker sized objects (solid cylinders
with a length of 38mm and diameter of 6.65mm) constructed from materials commonly
used in the large conductive surfaces of AIMDs (Titanium, MP35NLT, Tantalum,
Lithium, CFx, Pt-Ir, Nickel, Aluminum).
The simulations were set up to emulate
exposure to the gradient induced device heating safety test per ISO/TS 10974,
where the test object is placed inside a tissue simulating gel solution (1.2
S/m HEC) and then exposed to a changing magnetic field of 42T/s RMS using a
sine waveform with a frequency of 1750Hz for a total of 15 minutes per
exposure.
An initial set of simulations was
conducted on a Titanium object to determine the orientation with respect to the
changing magnetic field, parallel vs perpendicular (see Figure 1), which
generates the largest heating. Titanium was selected for this simulation as it
is the most commonly used material for AIMD enclosures.
Once the worst-case orientation was
determined, simulations were conducted on all the materials specified above to determine
the worst-case heating for each material.
To validate the simulation results,
MR gradient induced heating testing was performed on three of the modeled materials:
Titanium, Nickel, and Aluminum. Titanium was selected as it is the most
commonly used material for leadless pacemaker enclosures. Nickel and Aluminum
were chosen as these were the materials which resulted in the largest heating
from the simulation results.
The gradient induced heating testing
was performed using the methods specified in the gradient induced device heating
clause of ISO/TS 10974 with the same exposure conditions as the simulations.
The test object was placed inside a
phantom filled with a tissue simulating gel solution (1.2 S/m HEC) perpendicular
to the dB/dt field. Using a research gradient magnetic field generator, the object
was exposed to a changing magnetic field of 42T/s RMS using a sine waveform
with a frequency of 1750Hz for a total of 15 minutes per exposure.
Fiber optic temperature probes were
placed across the surface of the cylindrical test object, including the hotspot
identified in the modeling results. The approximate temperature probe
placements on the object under test are shown in Figure 2. These temperature probes were used to monitor
and record the test objects’ temperature response throughout the test exposure
with temperature measurements being recorded at least once every second. In addition, one probe was also placed near
the edge of the test phantom to monitor background temperature changes.Results
An initial set of simulations were
conducted on a Titanium cylinder in both the parallel and perpendicular
orientations shown in Figure 1. The
results of the simulations, seen in Figure 3, show that the perpendicular
orientation is worst case, generating the largest amount of heating.
Once the worst-case orientation was
determined (perpendicular), simulations were then conducted for the remaining
materials in that orientation. The
simulation results are shown in Figure 4. As can be seen in this figure, the
resultant temperature rise for all of the materials modeled was less than 1°C,
with Aluminum, Nickel and Lithium generating the highest heating.
To validate the simulation results,
benchtop testing was conducted on three representative materials. The results
of the benchtop testing are shown in Figure 5.
As can be seen in this Figure, the validation test results match well
with the modeling results.Discussion/Conclusions
As seen in the results section
above, the modeling and benchtop testing of the leadless pacemaker sized
objects resulted in temperature rises which were all less than 1°C, with the
largest temperature rise seen in the aluminum test object. This object serves as
a conservative surrogate for leadless pacemakers as the entire test object is
constructed from the worst-case heating material, whereas leadless pacemakers
are constructed from a composite of materials which will not all contribute to
heating. In general, a temperature rise less than 2°C is considered to be
clinically insignificant2. As such, the risk of leadless pacemaker
gradient induced device heating is negligible and assessments for this hazard per
ISO/TS 10974 may not be necessary.Acknowledgements
No acknowledgement found.References
1.
ISO/TS
10974, 2018, "Assessment of the safety of magnetic resonance imaging for
patients with an active implantable medical device" ISO, Geneva,
Switzerland, www.astm.org.
2.
Fda.gov,
2021, “Testing and Labeling Medical Devices for Safety in the Magnetic
Resonance (MR) Environment, Guidance for Industry and Food and Drug
Administration Staff” [online] Available at: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/testing-and-labeling-medical-devices-safety-magnetic-resonance-mr-environment