Carmel Hayes1, Yan Tu Huang2, Manuela Rick1, Randall Kroeker1, Mario Bacher1, Yue Pan3, Juliet Varghese3, Orlando Simonetti3, Ning Jin4, Rachel Davids4, Kelvin Chow4, Rudy Vanliedekerke 5, Marieke Vangelder5, Johan Dehem5, Peter Gatehouse6, Raj Kumar Soundarajan 6, Ricardo Wage6, Alessia Azzu6, Sonia Nielles-Vallespin6, and Peter Speier1
1Siemens Healthcare GmbH, Erlangen, Germany, 2Siemens Shenzhen Magnetic Resonance, Shenzhen, China, 3The Ohio State University, Columbus, OH, United States, 4Siemens Medical Solutions USA, Malvern, PA, United States, 5Jan Yperman Ziekenhuis, Ieper, Belgium, 6Royal Brompton Hospital, London, United Kingdom
Synopsis
A multi-center evaluation of a Pilot Tone-based
triggering technique using the BioMatrix Beat Sensor has been performed in healthy
volunteers and patients undergoing a cardiac MRI examination. At five different
sites, and two different field strengths, a wide range of MR sequences typically used in standard cardiac MRI examinations
were successfully triggered using the prototype Beat Sensor Cardiac triggering
technology. By including a short signal
training and RF calibration step at the start of the examination, the device proved
robust across a range of subjects, thereby offering an alternative means to
trigger cardiac examinations without the need to attach electrodes.
Introduction
The
Beat Sensor Cardiac triggering technology, which is
the application of Pilot Tone to
cardiac triggering, is a relatively new technique that allows for
electrode-free triggering of cardiac MRI sequences (1 - 3). In this work we evaluated
this technique at five MRI centers and at two field strengths in both patients
and in healthy volunteers. We were interested in assessing the robustness of
the method in different subject types, field strengths, and sequence flavors,
as well as the stability of the technique with respect to workflow and usability.
Methods
A prototype MR sequence and signal processing package was developed that
allows for Beat Sensor Cardiac triggering of a wide range of MR sequence types
typically used in standard cardiac MRI examinations. The pilot tone signal transmission and detection were performed using the commercially
available BioMatrix Beat Sensor body array coil (Siemens Healthcare GmbH,
Erlangen, Germany). The coil was positioned over the thorax, centered on the estimated heart position. The prototype software package 1) trains and calibrates the pilot tone signal, combining signals from all active local coil arrays and extracting the cardiac motion component, as well as correcting for signal corrupted e.g., during an RF pulse and 2) calculates a final filtered signal that is the inverted temporal derivative of the cardiac motion component. The trigger time point corresponds to the time of early systolic contraction and occurs ca. 200 ms after the R-wave. To facilitate the acquisition of static images of the heart in the end-diastolic phase of the cardiac cycle, the timepoint of the R-wave was estimated for each individual case from the cardiac motion component and used to adjust the timing parameters of the sequence protocol.
The prototype was tested in a total of 55 subjects, of which 20 were patients, at five different sites and at two different field strengths (1.5 T, 3 T) running the software version syngo XA31A. Relevant for this evaluation
were the subject sizes (height and weight) as well as the range of heart rates
and heart rate variations, the presence of devices in the scanner
room that could potentially interfere with the pilot tone signal and the number
of different operators. Of the patients examined, indications included sarcoidosis, hypertrophic cardiomyopathy, myocarditis, chest pain, atrial fibrillation, among others. For reference purposes an ECG or pulse triggering
device was attached and additional comparison data were acquired with ECG triggering
in 30 subjects. Patients were positioned supine, head first in the scanner. All scanning facts are summarized in Figure 1.Results
In
all 55 subjects a wide range of standard clinical cardiac MR sequences were
successfully triggered with the novel Beat Sensor Cardiac trigger technique.
The exact type of sequences tested varied from examination to examination and
site to site. However, in all cases at least localization - with single shot
TrueFISP and dark-blood HASTE - and/or cine or late-enhancement sequences (TrueFISP) were performed
and were accompanied by additional sequences depending on the clinical query or
test protocol. In total, the following sequences were acquired successfully:
- 47 cases cine
- 46 cases T1 mapping
- 27 cases T2 mapping
- 45 cases late-enhancement in free-breathing or breath-hold
- 21 cases dynamic sequences, of which 7 with contrast agent and one with adenosine
- 22 cases flow quantification in breath-hold or free-breathing
- 22 cases dark-blood prepared turbo spin echo both with and without STIR
preparation.
At
three sites in-room devices including monitors and contrast agent power injectors were switched on
and the Beat Sensor signal monitored for disturbances: no noteworthy signal
changes were found. On 8 of 55 occasions the pilot tone signal calibration and
learning phase had to be repeated. This was due to incorrect operator timing of the
two-step signal training and calibration process and associated with the prototype nature of these adjustment
steps. In all cases, once the training and
calibration were successfully completed, no further signal training and
calibration was necessary, despite patient table movement for contrast administration. Thus, even for longer examinations (> 60 minutes) a stable signal was maintained. On 5 occasions an inverted trigger signal was documented, which usually could be resolved by repeating the calibration, although in all cases
this did not compromise the trigger quality but had to be considered when optimizing the timing of static image acquisitions. In several individuals with highly variable heart rates, including a patient with atrial fibrillation, Beat Sensor Cardiac triggering functioned well (Figure 2). All examinations were
successfully performed in breath-hold on expiration, combined with
free-breathing scans as appropriate. In a subset of 5 subjects, comparison of quantitative parameters derived from Beat Sensor Cardiac triggered image data e.g., EF, EDV, ESV, T1, T2 and T2* values were comparable to those derived from ECG triggered image data (4). See figure 3 for example images.
Conclusions
We were able to demonstrate in a five-center evaluation of the Beat
Sensor Cardiac triggering device that pilot tone-based triggering of standard
cardiac examinations is feasible in a range of different patients and subject types and at
different field strengths. This technology offers an alternative means to
trigger clinical cardiac examinations without the need to attach electrodes, and, unlike
ECG triggering, is robust to gradient interference. Acknowledgements
No acknowledgement found.References
1. Speier
P, Fenchel M, Rehner R: Pt-nav: a novel respiratory navigation method for
continuous acquisitions based on modulation of a pilot tone in the MR-receiver.
Magn Reson Mater Phys Biol Med 28, 97{98 (2015)
2.
Schroeder L, Wetzl J, Maier A, Lauer L, Bollenbeck J, Fenchel M, Speier P: A
novel method for contact-free cardiac synchronization using the pilot tone
navigator. In: Proceedings of the 24th Annual Meeting of ISMRM, Singapore, p.
410 (2016)
3. Bacher
M, Speier P, Bollenbeck J, Fenchel M, Stuber M: Pilot tone navigation enables
contactless prospective cardiac triggering: initial volunteer results for
prospective cine. In: Intl. Soc. Mag. Reson. Med, vol.26, p. 4798 (2018)
4. Varghese
J, Pan Y, Hayes C, Jin N, Simonetti 0, Speier P: Comparison of Beat Sensor
Cardiac Triggering with ECG Triggering in a Comprehensive Cardiac MR
Examination: Initial Volunteer Experience. submitted to SCMR 25th Annual
Scientific Sessions (2022)