Graeme Harris1, Stephen Jermy1,2, and Ernesta Meintjes1,2
1Department of Human Biology, Division of Biomedical Engineering, University of Cape Town, Cape Town, South Africa, 2Cape Universities Body Imaging Center, University of Cape Town, Cape Town, South Africa
Synopsis
Respiratory motion of the heart is a fundamental
challenge to cardiac MR imaging (CMR), frequently compensated for with
breath-holding and acceptance-window methods. These methods are not always
viable and result in inefficient acquisitions, creating longer scan times1.
Here, an adaptive Kalman-filter with a control system was implemented in a
FLASH sequence to update the position of the slice during the imaging segment
based on repeated navigator measurements acquired during the non-imaging
segment. This predictive tracking results in a free-breathing sequence which is
less strenuous for subjects, more efficient and reduces scan times.
Introduction
Respiration causes severe image degradation in
CMR. A common correction method uses diaphragmatic navigators to accept only
information acquired when the diaphragm is within certain window positions2.
This substantially increases the acquisition time. Here we implement a prospective
motion correction control system based on an adaptive Kalman filter in a FLASH
sequence3. Based on diaphragm positions measured during non-imaging
segments, the Kalman filter predicts the position of the diaphragm during the
imaging segment. The position of the slice is then prospectively adjusted using
a linear scaling factor. Typically, a generalized scaling factor of 0.6 is used
but this does not compensate for the variation amongst subjects nor the 3D
nature of the heart.4Methods
The FLASH sequence was modified to include
multiple navigator readings between imaging segments. Initially a set of 256
diaphragm positions is acquired to pre-train the Kalman filter. Before each
imaging segment navigator positions are gathered and used as the input into the
control system. The predicted outputs of the control system are then used for
slice following during the imaging segment, during which no navigator
information is available. Imaging was tested with phantoms initially and
followed by a set of in vivo scans from 8 healthy volunteers. All
imaging was performed on a 3T Skyra (Siemens AG, Erlangen). Initial phantom testing was performed utilizing a motion rig (Figure 1)
that simulates breathing motion. Five sets of ECG-triggered FLASH acquisitions
were performed in each healthy volunteer: (i) breath holds (BH)s, (ii) free-breathing
with no motion correction, (iii) free-breathing-navigated-FLASH with a 4mm
acceptance window, (iv) free-breathing FLASH adapted to utilize the control
system, and (v) a set of low-resolution cine FLASH images (TR=86ms, 50 images).
The log data from the acquisitions with the control system adapted sequence were
then analysed to measure the accuracy of the control system’s predictions. Images
acquired with the standard BH sequence were compared to the control system
adapted sequence, the acceptance window slice following sequence and the
uncorrected free-breathing sequence. Finally, the set of cine images were segmented
at the lung-liver interface and around the heart. The edge of the lung liver
interface and an edge of the heart were tracked to calculate the proportional
change of the diaphragm’s position to the heart’s position, for each subject.Results
Figure 2 shows the control system’s predictions for
the diaphragm position during the imaging segments (red), given the diaphragm
positions measured by the navigator during non-imaging segments (blue). Figures
3 and 4 show example images from uncorrected breathing, free-breathing windowed,
free-breathing control system corrected acquisitions. Tracking the edge of the
lung-liver interface and the heart yielded variable tracking factors across
subjects (0.65 to 0.77). For the subject shown, the tracking scaling factor was
0.77.Discussion
Although the position tracking of the control
system is accurate, tracking during non-linear sections of the breathing cycle
remains challenging. The resultant images show improved quality using the
control system compared with no correction and similar quality when compared
with the windowed acquisition, although artefacts due to the expansion and
contraction of the chest wall remain.Conclusion
The control system adapted acquisition provides
similar image quality to the windows acquisition, whilst maintaining 100%
imaging efficiency through the respiratory cycle. There is clear evidence for
variation in the tracking scaling factor amongst subjects and therefore a need
for subject specific adaptation when considering navigated sequences.Acknowledgements
No acknowledgement found.References
1. Holland, A. E., Goldfarb, J. W., &
Edelman, R. R. (1998). Diaphragmatic and cardiac motion during suspended
breathing: preliminary experience and implications for breath-hold MR imaging. Radiology,
209(2), 483–489.
2. Sachs, T. S., Meyer, C. H., Hu, B.
S., et al. (1994). Real-time motion detection in spiral MRI using navigators. Magnetic
Resonance in Medicine, 32(5), 639–45.
3. Taylor, A. M., Keegan, J., Jhooti,
P., Firmin, D. N., & Pennell, D. J. (1999). Calculation of a subject-specific
adaptive motion-correction factor for improved real-time navigator echo-gated
magnetic resonance coronary angiography. J Cardiovasc Magn
Reson, 1 (2), 131{138. doi: 10.3109/10976649909080841
4. Burger, I. H., Keegan, J., Meintjes, E., et al. Prospective diaphragm
position prediction for Cardiac MR using multiple navigators. Proc ISMRM-ESMRMB
18, ISSN# 1545-4428, Stockholm, Sweden, 1-7 May 2010. (#5013)