Yue Pan1,2, Juliet Varghese1, Ning Jin3, Carmel Hayes4, Peter Speier4, Rizwan Ahmad2,5, and Orlando Simonetti1,2,6
1Davis Heart and Lung Research Institute, The Ohio State University Wexner Medical Center, Columbus, OH, United States, 2Department of Biomedical Engineering, The Ohio State University, Columbus, OH, United States, 3Siemens Medical Solutions USA, Malvern, PA, United States, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Department of Electrical and Computer Engineering, The Ohio State University, Columbus, OH, United States, 6Department of Internal Medicine, The Ohio State University Wexner Medical Center, Columbus, OH, United States
Synopsis
In
this study, the performance of Beat Sensor Cardiac (BSC) triggering was qualitatively
and quantitatively compared with electrocardiogram (ECG) triggering. MR images typically
acquired in a comprehensive exam including localizer, morphology, cine, 2D
flow, parametric mapping and late gadolinium enhancement (LGE) were acquired
using both BSC and ECG triggering in volunteers and patients. The overall image
quality for BSC was equivalent to ECG. Quantitative measurements of function,
flow, and parametric maps in healthy volunteers showed no significant
differences except for peak aortic velocity.
Introduction
ECG
is the most commonly used triggering method for cardiac MRI. However, the
patient preparation required to apply and adjust ECG electrodes can be time
consuming. At higher field strengths, the ECG signal is prone to distortion due
to gradient interference and the magnetohydrodynamic effect. BSC triggering is
a Pilot Tone based method that does not require extra patient preparation. It
utilizes an external radiofrequency (RF) signal that is modulated by the patient’s
physiological motion1-3. In this study, we evaluated the performance of BSC vs.
ECG triggering in a comprehensive cardiac MRI exam by comparing the quality of
images and quantitative measurements.Methods
Both volunteer
and patient data were collected with prototype sequences on a clinical 3T
system (MAGETOM Vida, Siemens Healthcare, Erlangen, Germany). A calibration
procedure was performed once at the beginning of the exam to train and
calibrate BSC triggering signals and correct for signal interference caused by
RF pulses. A comprehensive, non-contrast cardiac exam including localizer,
morphology, cine, 2D flow, and parametric mapping was acquired in 14 volunteers
(38 ± 14.5 years, 6 females) with both BSC and ECG triggering. Biventricular cardiac
function and flow in aorta and main pulmonary artery (MPA, collected in 9
volunteers) were evaluated in suiteHEART (NeoSoft, Pewaukee, WI, USA). Myocardial
T1, T2, and T2* values were measured in the interventricular septum. Pairwise
student t-tests were performed to compare quantitative measurements from images
acquired with BSC vs. ECG triggering.
To evaluate the
effectiveness of BSC triggering post contrast agent administration, BSC
triggered pre- and post-contrast T1 maps and LGE were added to a standard ECG
triggered clinical exam in 6 patients (27 ± 10.4 years, 4 females). Calibration
was performed prior to contrast injection to evaluate the robustness of BSC
triggering post-contrast. Scan parameters of BSC triggered sequences were set
to match the ECG triggered clinical scans performed in the same patient. Inversion
time (TI) in BSC triggered LGE scans were increased slightly relative to the
ECG scans to account for the longer time after contrast injection. Pre- and
post-contrast T1 maps were used to calculate extracellular volume (ECV). LGE
image quality and sharpness were evaluated pairwise by a CMR expert (30 years
of experience) blinded to the triggering method. Sequence types and
corresponding imaging parameters for both volunteer and patient scans are shown
in Table 1.Results
BSC
triggered exams were successfully performed in all healthy volunteers and
patients. All data from one volunteer were discarded due to gross patient
motion observed between scans, resulting in poor correspondence of slice
positions between BSC and ECG triggered images. Example images from healthy
volunteers using both triggering methods are shown in Figure 1. All LGE images acquired
in patients were reviewed and no significant differences in image quality or sharpness
were detected. Figure 2 shows the LGE and T1 maps from one patient, showing mild
enhancement in the basal to mid anterior and inferior wall in the long axis
view, and at the RV attachment site in the mid short axis view. A summary of
quantitative results measured in both volunteers and patients is shown in Table
2. Due to the limited size of the patient cohort, statistical analysis was
performed in only healthy volunteers.Discussion
The
feasibility of BSC triggering was evaluated by comparing with ECG triggering. The
BSC utilizes an external RF signal to detect physiological motion. Depending on
Pilot Tone transducer placement relative to the patient’s anatomy, the received
signal characteristics can vary. Thus, a patient-specific training scan is
required to calibrate the BSC trigger. Overall, nearly all comparisons
demonstrated equivalent image quality and quantitative results. However, we did
observe a systematic reduction in peak aortic velocity with BSC triggering.
This could have been due to either a misalignment of image frames between
ECG and BSC, or slight variations in BSC triggering between heartbeats. The
triggering constancy relative to ECG needs further investigation. In Figure 2,
signal to noise ratio in the blood pool appears slightly lower in the BSC
triggered LGE and the post contrast T1 map. This could be due to the contrast
agent washout from the blood pool because BSC triggered images were acquired
later after contrast injection.Conclusion
Beat
Sensor Cardiac triggered MR exam was successfully evaluated and validated in
healthy volunteers and patients. Its performance was compared with standard ECG
triggering across multiple imaging applications. The overall image quality was equivalent
for BSC and ECG. Quantitative measurements in function, flow, and parametric
maps in healthy volunteers showed no significant difference in all measurements
except peak aortic velocity. Acknowledgements
Debbie Scandling, The Ohio State University, Columbus, OH, United States, for helping with volunteer recruitmentReferences
1. Schroeder,
L., et al. A novel method for
contact-free cardiac synchronization using the pilot tone navigator. in Proceedings of the 24th Annual Meeting of
ISMRM, Singapore. 2016.
2. Bacher, M., Cardiac triggering based on locally generated pilot-tones in a
commercial MRI scanner: a feasibility study. 2017.
3. Bacher M, Speier P, Bollenbeck J, Fenchel M, Stuber M: Pilot tone navigation enables contactless prospective cardiac triggering: initial volunteer results for prospective cine. In: Intl. Soc. Mag. Reson. Med, vol.26, p. 4798 (2018)