Ziyi Pan1, Jianxiong Hu2, Hai Luo2, Simin Liu1, Sisi Li1, Ziyue Wu2, and Hua Guo1
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China, 2Marvel Stone Healthcare Co., Ltd., Wuxi, China
Synopsis
Low field MR-guided thermotherapy can provide some key
advantages over the high-field alternative, including reduced cost, decreased
susceptibility artifacts, and improved safety of interventional devices. However, both the accuracy
and the speed of PRF temperature measurement suffer at the low field due to the reduced
SNR, limited receive channels, and
declined temperature-induced phase changes, making it unreliable for clinical
MRgLITT treatments. In this study, we demonstrate that the multi-echo
thermometry together with the view-sharing acceleration can be utilized to achieve high-quality PRF thermometry
at 0.5T with satisfactory temperature measurement precision and temporal resolution.
Introduction
MR-guided laser interstitial thermal therapy (MRgLITT)
has been applied as a minimally invasive treatment in neurosurgery. Recent
research1 has
shown the great potential of low field MR systems (such as 0.55T) on MR-guided
interventions and thermotherapies, benefiting from reduced cost, declined interventional
device heating and decreased susceptibility artifacts. However, the precision
of the proton resonance frequency (PRF) shift-based thermometry2 can be deteriorated at low-field due to low SNR3,4, hindering its
clinical reliability for LITT temperature monitoring. Additionally, sampling acceleration is needed in MRgLITT to ensure enough temporal resolution, yet
maintaining SNR in the meantime has to be considered.
This study aims to
improve the temperature precision at low-field using multi-echo GRE and to accelerate
the acquisition using a view-sharing approach5-9
without sacrificing SNR. The performance of this method was validated on a 0.5T
scanner.Theory and Methods
The
influence of B0 Decrease on PRF Thermometry
PRF thermometry
is challenging at 0.5T due to the decrease of the magnetic field B0.
- The accuracy of temperature measurement drops dramatically due to the B0
decrease. The uncertainty of temperature measurement at 0.5T is 20 times higher than that
at 1.5T according to (1)3:
$$\delta T=\sqrt{2}/(2\pi\cdot\gamma\cdot B_0\cdot\alpha\cdot TE\cdot SNR),\ \ SNR\propto B_0^{7/4}\ \ \ (1)$$
- The sensitivity of temperature measurement is declined at 0.5T because
temperature-induced phase difference is inversely proportional to B0 according to (2)2 :
$$\triangle \phi = \alpha\triangle T\cdot\gamma B_0\cdot TE\ \ \ (2)$$
Multi-echo
Thermometry
Longer TE can provide
better temperature precision and sensitivity according to Eq. (1) and (2) in
GRE-based thermometry. Therefore, multi-echo thermometry10,11 with its ability to use a
wide range of TEs without increasing scan time is well suited at low fields. A
bipolar GRE sequence12 was
applied to achieve higher SNR efficiency.
-
Echo Misregistration Correction: Since
the receiver bandwidth is kept low to
ensure sufficient SNR in this study, field inhomogeneity and eddy current-induced
misregistration between even and odd echoes is corrected13 with an
estimated B0 field map from multi-echo data.
- Echo
Combination: The measured temperature maps are
combined into a single estimate using a tSNR optimal weighted echo combination
approach14,
whereby the weights are the production of image magnitude and TE.
View-sharing
Acceleration
The view-sharing-based approach, with its ability to achieve fast reconstruction, and to preserve accelerated
image SNR without the need for multiple receive channels, is beneficial for the low field. So this study adopted the view-sharing acceleration strategy
and investigated its feasibility.
Experiments
All experiments
were performed in a 0.5T MRI scanner (Marvelstone, Wuxi, China) with an
eight-channel receive coil. Temperature data were acquired by a multi-echo
bipolar GRE with the following parameters: flip angle = 30°, TE = 7.5~40.5ms,
nTE / delta TE = 7 / 5.5ms, TR = 138ms, matrix = 108×110, FOV = 220×220 mm2,
slice thickness = 5mm, 3 slices (no gap), sampling bandwidth (BW) = 27.8 kHz,
view-sharing acceleration = 3 (variable density), temporal resolution = 5
sec/volume.
- Simulation
Experiments: Fully sampled data were first acquired on
pork tissues using the sequence mentioned above but
without acceleration (temporal resolution=12 sec/volume, heat at 4W for 120s). Two different k-t
undersampling patterns (Fig.1A) were tested. A sliding window-based algorithm5, which
recovered undersampled k-space using previous time frames, was implemented for
reconstruction. The reconstructed temperature maps were compared to those
reconstructed using GRAPPA.
- In Vitro Tissue
Heating Experiments:
In vitro MRgLITT experiments (MR-Guided Laser Ablation
System, Sinovation Medical, Beijing, China) were carried out in pork tenderloin samples (N=3, heat at 6W for the 60s,
wait for 60s, then heat for another 60s) and pig brain tissues (N=2, heat at 4W for
50s) separately. An MR-compatible fiber optical sensor was inserted into the
tissue to obtain ground truth. Temperature uncertainty was calculated as the standard
deviation (SD) of the temperature difference between MR and fiber-optic
measurements. Results and Discussion
Multi-echo
Thermometry
As illustrated in Fig. 2, the
image misregistration between odd and even echoes can lead to wrong temperature
measurements (e.g. reduced temperatures at the heating center). Therefore, echo
misregistration correction is necessary for bipolar multi-echo thermometry,
especially when using low readout bandwidth (greater SNR and also
greater echo mismatch).
Fig. 3 demonstrates that the precision
of temperature measurement can be substantially improved after echo combination
when compared with single-echo measurements. Therefore, multi-echo
thermometry is highly recommended for low-field PRF measurement.
View-sharing
Acceleration
Simulation experiments show finer
temperature maps with lower errors in the 3-fold view-sharing acquisition (VarDensity, Fig. 1B).
GRAPPA reconstructed images suffer from SNR decrease which can lead to increased
temperature uncertainty. View-sharing acceleration (Interleaved) fails at rapid
temperature rise (Fig. 1B, first row) probably because the reconstruction uses too
much temperature data from previous time frames.
Temperature
Measurement Accuracy
Multi-echo
thermometry with view-sharing acceleration (3-fold, VarDensity) results in high-quality
temperature maps (Fig. 4) and excellent precision with the average temperature uncertainty of less than 0.8°C (Tab. 1) in in vitro heating experiments.Conclusion
We have
shown experimentally that multi-echo thermometry with view-sharing acceleration
can achieve high-quality PRF thermometry at 0.5T. The measured temperature uncertainty
is within a clinically acceptable threshold (<3°C). Still,
it needs further investigation on whether the view-sharing acceleration is acceptable
from a clinical perspective, as the temperatures may change more complicatedly during treatment. Acknowledgements
No acknowledgement found.References
1. Campbell-Washburn AE, Ramasawmy R,
Restivo MC, et al. Opportunities in Interventional and Diagnostic Imaging by
Using High-Performance Low-Field-Strength MRI. Radiology 2019;293(2):384-393.
2. Ishihara
Y, Calderon A, Watanabe H, et al. A precise and fast temperature mapping using
water proton chemical shift. Magn Reson Med 1995;34(6):814-823.
3. Botnar
RM, Steiner P, Dubno B, Erhart P, von Schulthess GK, Debatin JF. Temperature
quantification using the proton frequency shift technique: In vitro and in vivo
validation in an open 0.5 tesla interventional MR scanner during RF ablation.
Journal of Magnetic Resonance Imaging 2001;13(3):437-444.
4. Waqas
Majeed AJK, Sunil Patil, Henrik Odéen, John Roberts, Florian Maier, Dennis L.
Parker, and Himanshu Bhat. Feasibility of Magnetic Resonance Thermometry at
0.55T. Proc Intl Soc Mag Reson Med 29 (ISMRM & SMRT Virtual Conference
& Exhibition), 2021. p.
5. Riederer
SJ, Tasciyan T, Farzaneh F, Lee JN, Wright RC, Herfkens RJ. MR fluoroscopy:
technical feasibility. Magn Reson Med 1988;8(1):1-15.
6. Jones
RA, Haraldseth O, Muller TB, Rinck PA, Oksendal AN. K-space substitution: a
novel dynamic imaging technique. Magn Reson Med 1993;29(6):830-834.
7. van
Vaals JJ, Brummer ME, Dixon WT, et al. "Keyhole" method for
accelerating imaging of contrast agent uptake. J Magn Reson Imaging
1993;3(4):671-675.
8. Doyle
M, Walsh EG, Blackwell GG, Pohost GM. Block regional interpolation scheme for
k-space (BRISK): a rapid cardiac imaging technique. Magn Reson Med 1995;33(2):163-170.
9. Han
Y, Mun C. Evaluation of the keyhole technique applied to the proton resonance
frequency method for magnetic resonance temperature imaging. J Magn Reson
Imaging 2011;34(5):1231-1239.
10. Marx
M, Pauly KB. Improved MRI Thermometry with Multiple-Echo Spirals. Magn Reson
Med 2016;76(3):747-756.
11. Svedin
BT, Payne A, Bolster BD, Parker DL. Multiecho pseudo-golden angle stack of
stars thermometry with high spatial and temporal resolution using k-space
weighted image contrast. Magn Reson Med 2018;79(3):1407-1419.
12. Wieben
O LJ, Mansson S, Hennig J. Multi-echo balanced SSFP imaging for iterative Dixon
reconstruction. Proceedings of the 13th Annual Meeting of ISMRM Miami Beach,
2005. p.
13. Lu
W, Yu H, Shimakawa A, Alley M, Reeder SB, Hargreaves BA. Water-fat separation
with bipolar multiecho sequences. Magn Reson Med 2008;60(1):198-209.
14. Madore
B, Panych LP, Mei CS, Yuan J, Chu RX. Multipathway Sequences for MR
Thermometry. Magn Reson Med 2011;66(3):658-668.