Shi Su1,2, Zheyuan Yi1,2, Yujiao Zhao1,2, Linfang Xiao1,2, Ziming Huang1,2, Jiahao Hu1,2, Junhao Zhang1,2, Vick Lau1,2, Christopher Man1,2, Alex T.L. Leong1,2, and Ed X. Wu1,2
1Laboratory of Biomedical Imaging and Signal Processing, The University of Hong Kong, Hong Kong, China, 2Department of Electrical and Electronic Engineering, The University of Hong Kong, Hong Kong, China
Synopsis
Wave
encoding offers high acceleration in parallel imaging by exploring the coil
sensitivity variations in the readout dimension. However, typically preset wave
gradients (i.e., sinusoidal trajectory) have never been optimized for specific receiver
array coil, thus intrinsically limiting the maximum acceleration factor. We
propose to optimize the gradient trajectory in a coil sensitivity specific manner
by minimizing the squared L2-norm of off-diagonal elements in encoding
correlation matrix. To guarantee an allowed maximum gradient slew rate, a
bandlimited constraint is also introduced in optimization. This procedure leads
to significant improvements in g-factor map and artifact reduction, especially
at very high acceleration.
Introduction
The emerging wave encoding methods exploit the
coil sensitivity variation in the readout dimension for highly accelerated parallel
imaging[1-3]. They are typically achieved by applying wave gradients
(i.e., sinusoidal trajectory) in phase or/and partition dimensions to spread
the aliasing along the readout direction. However, gradient trajectory is often
preset in these existing wave encoding methods and non-specific to receiver array
coil. We develop a numerical procedure to optimize the wave gradient trajectory
for a given coil sensitivity so to significantly increase acceleration and
decrease maximum gradient slew rate. It minimizes the squared L2-norm of off-diagonal
elements in the encoding correlation matrix. This matrix reflects the g-factor
noise penalty and is determined by the wave gradients, coil sensitivity, and
undersampling pattern[4,5]. Further, this optimization includes a
bandlimited constraint in the frequency domain, enabling a smooth trajectory
and moderate slew rate in the time domain.Theory and Method
Wave gradient trajectory optimization
The squared L2-norm of off-diagonal elements in
the correlation matrix is calculated as[5]:
$$S_{w}=\sum_{j=1}^{N_{y}/R}\sum_{m=1}^R\sum_{n=1}^R\sum_{i=1}^L\left\|C_{i,(j+m)}^HF_x^HPSF_{j+m}^HPSF_{j+n}F_{x}C_{i,(j+n)}\right\|_2^2 (1)$$
where Ny is the number of phase encoding lines, L
the number of channels, and R the acceleration factor. $$$C_{i,(j+m)}$$$ or $$$C_{i,(j+n)}$$$ is a diagonal matrix with the elements being the ith
channel and (j+m)th or (j+n)th row of the coil sensitivity maps (i = 1, ..., L; j = 1, ..., Ny; n, m = 1, ..., R and n ≠ m). $$$PSF_{j+m}$$$ or $$$PSF_{j+n}$$$ is a diagonal matrix with the elements being the (j+m)th or (j+n)th
row of Point Spread Function (PSF), which is calculated as $$$PSF_{j+m}=e^{-i\gamma(j+m)\sum{g_{w}(t)}}$$$ and $$$PSF_{j+n}=e^{-i\gamma(j+n)\sum{g_{w}(t)}}$$$, where $$$g_{w}(t)$$$ is the gradient trajectory. $$$F_{x}$$$ represents the Fourier transform along x
(readout) dimension. Generally, small off-diagonal elements of the correlation
matrix result in decreased g-factors[5]. Therefore, the optimization
is to find the wave gradients that have the minimum value of Sw.
The main optimization steps are illustrated in Fig. 1.
First, we randomly generate an initial bandlimited complex data ($$$G_{w}(\omega)$$$) in the frequency
domain, and conduct inverse Fourier transform to obtain the wave gradient trajectory ($$$g_{w}(t)$$$) in the time domain. Afterward, with the gradient
trajectory, sensitivity maps, and undersampling pattern, the squared L2-norm of
the off-diagonal elements in the correlation matrix (Sw) is calculated according to Eq. (1) and then
optimized. Finally, the optimized $$$g_w^*(t)$$$ owning the minimum Sw* is obtained.
Simulation experiments
Simulation
experiments were conducted with various acceleration factors and slice
locations. Fully sampled 2D brain T2-weighted data were acquired on a Philips
3T MR system with an 8-channel head coil. The 2D FSE sequence parameters were: TE/TR
= 113ms/3000ms, bandwidth = 246Hz/pixel, FOV = 240×240mm2, in-plane
resolution = 1×1mm2, slice thickness = 4mm, ETL = 20, readout
oversampling factor = 2.
For comparison, a conventional sinusoidal wave
encoding gradient was used (i.e., incorporated into phase encoding direction) with
typical parameters: wave cycle = 4, maximum wave gradient = 11.55 mT/m (2×
readout gradient amplitude)[7,8]. For the proposed optimization, the
maximum gradient was kept the same. The theoretical PSFs were calculated using
the above-generated gradients trajectories. Wave encoded data were simulated based
on the fully sampled k-space data and PSF[8]. In all simulation
experiments, acceleration was implemented through uniform undersampling in phase
encoding steps retrospectively[5]. Also, the central 24 lines
were utilized to estimate the sensitivity maps. The ESPIRiT algorithm was used
for image reconstruction[7]. The g-factor maps were calculated using
the pseudo-multiple replica method.Results
The proposed optimization achieves consistent improvements
in reducing noise amplification, i.e., g-factors, for various accelerations (Fig.
3) and slice locations (Fig. 4). The improvement is particularly significant at very high acceleration
factors (R=6; see Figs. 3 and 4).
The error maps and g-factor maps exhibit very different patterns for sinusoidal
and optimized trajectories (see Fig. 3 as indicated by red arrows). This
illustrates the coil sensitivity specific control of aliasing by the optimized
trajectory, which leads to the reduction in residual artifacts and g-factors.
In addition, given the same maximum gradient amplitude, the optimized trajectory
requires a lower maximum gradient slew rate than the sinusoidal trajectory as
shown in Fig. 2. These results demonstrate that the optimized gradient
trajectory significantly outperforms the traditional sinusoidal trajectory.Discussion and Conclusions
The proposed method offers a new way to further
accelerate wave encoding parallel imaging by coil sensitivity specific
optimization of wave encoding gradient trajectory (i.e., optimization of 2D PSF
for minimizing noise propagation when unfolding
aliasing during parallel imaging reconstruction). Significant improvement is demonstrated
in g-factor reduction at very high acceleration. Note that residual artifacts
can be further reduced in practice by increasing the gradient amplitude or using
virtual conjugate coils[5]. The bandlimited constraint in the
proposed optimization procedure imposes not only low slew rate but also high
optimization efficiency. It corresponds to fewer optimization variables in the
frequency domain than in the time domain, thus resulting in more efficient
optimization. Moreover, this work can be extended to the 3D wave encoding
parallel imaging. For example, with two gradient trajectories and acceleration
in two dimensions, Eq. (1) can be modified by replacing the 2D PSF matrix with a
3D PSF matrix, and incorporating a correlation matrix of aliasing lines in both
dimensions. Acknowledgements
The authors thank Porf. Dong Liang, Prof.
Haifeng Wang and Dr. Zhilang Qiu for earlier and insightful encouragement and discussions. The authors also
thank Prof. Berkin Bilgic for sharing the codes online at http://martinos.org/∼berkin/software.html. The authors also thank Porf. Dong Liang, Prof.
Haifeng Wang and Dr. Zhilang Qiu for insightful discussions. This work was
supported in part by Hong Kong Research Grant Council (R7003-19F, HKU17112120
and HKU17127121 to E.X.W., and HKU17103819, HKU17104020 and HKU17127021 to
A.T.L.L.), Lam Woo Foundation, and Guangdong Key Technologies for Treatment of
Brain Disorders (2018B030332001) to E.X.W..References
[1] Bilgic
B, Gagoski BA, Cauley SF, et al. Wave-CAIPI for highly accelerated 3D imaging.
Magn Reson Med. 2015;73:2152-2162.
[2] Moriguchi
H, Duerk JL. Bunched phase encoding (BPE): a new fast data acquisition method
in MRI. Magn Reson Med. 2006;55:633-648.
[3] Breuer
FA, Moriguchi H, Seiberlich N, et al. Zigzag sampling for improved parallel
imaging. Magn Reson Med. 2008;60:474-478.
[4] Kettinger
AO, Kannengiesser SAR, Breuer FA, et al. Controlling the object phase for
g-factor reduction in phase-constrained parallel MRI using spatially selective
RF pulses. Magn Reson Med. 2018;79:2113-2125.
[5] Qiu Z, Jia
S, Su S, et al. Highly accelerated parallel MRI using wave encoding and virtual
conjugate coils [J]. Magn Reson Med. 2021;86:1345-1359.
[6] Uecker
M, Lai P, Murphy MJ, et al. ESPIRiT—an eigenvalue approach to autocalibrating
parallel MRI: where SENSE meets GRAPPA. Magn Reson Med. 2014;71:990-1001.
[7] Polak D,
Cauley S, Huang SY, et al. Highly-accelerated volumetric brain examination
using optimized Wave-CAIPI encoding. J
Magn Reson Imaging. 2019;50:961-974.
[8] Wang H, Qiu Z, Su
S, et al. Parameter optimization framework on wave gradients of Wave-CAIPI imaging.
Magn Reson Med. 2020;83:1659-1672.