Yingchao Tan1,2,3, Qiaoyan Chen1,2, Qian Zou3, Yanling Chen3, Junyu Chen3, Xin Liu1,2, Hairong Zheng1,2, and Ye Li1,2
1Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China, 2Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province, Shenzhen, China, 3Shanghai United Imaging Healthcare, Shanghai, China
Synopsis
MR
vessel wall imaging of intracranial and carotid arteries is still challenging
because of the small cross-sectional size of the vessel wall and the cord, and
susceptibility effects, especially in the carotid and the spinal cord. In this
study, a 72-channel coil system that consists of a 64-channel head coil
combined with an 8-channel carotid coil was proposed and characterized in its
performance by comparison with a 40-channel coil system. As a result, the proposed
72-channel coil system provides improved performance in SNR, parallel imaging
capability, and image quality.
Introduction
Atherosclerotic plaque is a primary cause of ischemic stroke and is prevalent in the aorta and carotid vessels [1]. Simultaneous intracranial and extracranial arterial wall imaging would greatly enhance the convenience of clinical diagnosis. However, the normal middle cerebral artery and basilar artery wall thickness is 0.2–0.3 mm, which needs a high spatial resolution imaging. Previous studies showed that a high isotropic spatial resolution of 0.6 mm can be achieved by using a 40-channel coil system including a 32-channel head coil and an 8-channel carotid coil [2]. In this work, we have designed and built a 72-channel coil system with a 64-channel head coil and an 8-channel carotid coil for MR vessel wall imaging at a 3 T MRI system (uMR 880, Shanghai United Imaging Healthcare, Shanghai, China). The coil performance was characterized with improved SNR, enhanced parallel imaging capability, and superior image quality in human studies, compared to a 40-channel coil system with a 32-channel head coil and an 8-channel carotid coil.Methods
Figure 1 shows the
photograph and the circuit schematic of the 72-channel coil system. The head coil
elements with an 80 mm diameter were arranged in six rows in H-F direction. The
inner dimensions of the head coil shell were: 230 mm in the Anterior-Posterior
(AP) direction, 205 mm in the Left-Right (LR) direction, and 255 mm in the
Inferior-Superior (IS) direction. Conventional front-end circuits generally
include matching and tuning circuits, detuning circuits and preamplifier. Each
element was tuned to 128.23 MHz and 50 ohm impedance-matched to minimize the
noise of the preamp.
A 2D density-weighted
gradient echo (GRE) sequence was applied for signal acquisitions with a
phantom. The parameters were as followings: TR/TE =400/20 ms, flip angle=30o,
slice thickness=5mm, matrix=256×256, FOV=250mm×250mm. The noise images were
obtained by setting the flip angle to zero. For SNR comparisons, SNR maps were
calculated using the sum-of-squares method [3]. For parallel imaging capability
evaluation, the inverse g-factor maps were analyzed by using sensitivity
encoding (SENSE) reconstructions [4].
Human anatomy images
were acquired using a 3D GRE sequence with with following parameters: TR/TE=8.4/3.4
ms, flip angle=9o, FOV=220mm×256mm, matrix=316×368, reconstructed
resolution=0.7×0.7×0.7mm3, bandwidth=250 Hz/pixel, compression
channel = 36, uCS (united compressed sensing) acceleration factor=5.
Vessel wall imaging
of intracranial and carotid arteries were acquired using a T1-weighted fast
spin echo (FSE) MATRIX (Modulated
flip angle technique in refocused imaging with extended echo train) sequence with following parameters: TR/TE=800/14.82 ms, flip
angle=75o, FOV=180mm×220mm, matrix=380×464, reconstructed
resolution=0.47×0.47×0.47mm3, bandwidth=440 Hz/pixel, compression channel = 36, uCS acceleration factor=5.5.
Results
Figure 2 shows phantom SNR maps
in the transverse, sagittal and coronal planes. Additionally, SNR profiles in x, y and z directions were also depicted in Figure 2.
The mean SNR values in the ROIs depicted as the circles or rectangles and SNR
profiles both demonstrated that the 72-channel coil system had higher SNR than
the 40-channel coil system. Figure 3 depicts the inverse g-factor maps in the
transverse plane with acceleration factors R from 2 to 6. The results indicate
that the parallel imaging capability of the 72-channel coil system was better
than that of the 40-channel coil system, particularly at high acceleration
factors. Figure 4 displays the anatomy images of human brain with a 0.7 mm
resolution and an acceleration factor of 5 using the 72-channel and 40-channel
coil systems. Figure 5 displays the MR vessel wall images using the 72-channel
and the 32-channel coil systems. Both these images using 72-channel coil system
showed a higher quality than those using 40-channel coil system.
Discussions and Conclusions
A 72-channel array coil system
for MR vessel wall imaging of intracranial and carotid arteries was designed,
constructed and evaluated by imaging experiments in phantom and human studies. Compared
to the 40-channel coil system, the 72-channel coil system achieves better
performance in MR SNR, acceleration capacity and quality of MR vessel wall
images.Acknowledgements
This work is
supported by the Strategic Priority Research Program of Chinese Academy of
Sciences, XDB25000000; National Key R&D Program of China, 2021YFE0204400;
Shenzhen city grant, RCYX20200714114735123, ZDKJ20190204003, ZDKJ20190204004.References
[1] Mandell DM, Mossa-Basha M, Qiao Y, Hess CP,
Hui F, Matouk C, Johnson MH, Daemen MJ, Vossough A, Edjlali M, Saloner D,
Ansari SA, Wasserman BA, Mikulis DJ; Vessel Wall Imaging Study Group of the
American Society of Neuroradiology. Intracranial Vessel Wall MRI: Principles
and Expert Consensus Recommendations of the American Society of Neuroradiology.
AJNR Am J Neuroradiol. 2017, 38(2):218-229.
[2] Q. Chen, Z. Wei,
L. Zhang, C. Tie, Q. He, X. Zhang, X. Liu, H. Zheng, and Y. Li. “MR vessel wall imaging
of intracranial and carotid arteries with a 40-channel coil system at 3 T,” ISMRM, 2019, p1488.
[3] Roemer PB, Edelstein WA, Hayes CE, Souza SP,
Mueller OM. The NMR phased array. Magn Reson Med. 1990, 16(2):192-225.
[4] Pruessmann KP, Weiger M, Scheidegger MB,
Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999,
42(5):952-62.