Lingping Ran1, Lu Huang1, Xianghu Yan1, Yi Luo1, Shuheng Zhang2, Shiyu Zhang2, Yuan Zheng3, Jian Xu3, and Liming Xia1
1Department of Radiology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China, 2United Imaging Healthcare, Shanghai, China, 3UIH America, Inc., Houston, TX, United States
Synopsis
High resolution
T2-weighted dark blood (HR-T2W-DB) imaging is not always robust for clinical
use because of low SNR and long scan time. The purpose of this study was to
evaluate a novel deep learning (DL) based reconstruction method in T2W-DB
sequence that achieves higher spatial resolution and same scan duration
compared with traditional reconstruction method. Quantitative and qualitative
image assessment demonstrated that DL based HR-T2W-DB sequence showed better
CNR in region of edema and LV free wall visibility, which might help detecting
myocardial edema.
Introduction
T2-weighted dark
blood (T2W-DB) imaging is an important technique to evaluate myocardial edema.
Constrained by long breath-hold time, 2D dual inversion breath-hold fast spin
echo sequence for cardiac T2W-DB imaging requires high acceleration factors to
achieve high spatial resolution, which usually causes a compromise in image
quality due to low signal-to-noise ratio (SNR).
Deep learning has
been used frequently in MRI reconstruction, especially in de-noising, and
restoration, et al. It has been proven to have the potential to shorten MR scan
time or to improve image quality.
The purpose of
this study was to implement the combination of deep learning based acceleration1 and reconstruction2 method on a high resolution T2W-DB
sequence, and to compare the image quality, diagnostic value with the
conventional T2W-DB with routine reconstruction method and same scan duration.Methods
Deep
learning reconstruction:
DL-based
HR-T2W-DB imaging acquisition was under the scheme of a novel deep learning
based acceleration method (AI-assisted Compressed Sensing, ACS) 1.
Reconstructed multi-channel data was then transferred to a novel deep learning
based image reconstruction neural network 2 as input.
MR
Scan:
This clinical study was approved by the local
Institution Review Board. Five healthy volunteers (age 29±4, five male) and twenty eight patients (age 43±15,
twenty one male) were prospectively recruited to undergo cardiac magnetic
resonance imaging (CMR). Clinical indications included ischemic and
non-ischemic cardiomyopathy, cardiac valve disease, myocarditis, and other
diseases that were not classified. CMR was performed on a 3T scanner (uMR 790,
United Imaging Healthcare, Shanghai, China) with a 24-channel dedicated cardiac
coil. Conventional T2W-DB imaging was achieved with parallel imaging. Both
conventional T2W-DB and DL-based HR-T2W-DB imaging were applied in short-axial
view at the same position (Figure 1). Detailed scan parameters were shown in Table1.
Image
Analysis:
Two
radiographers (X.Y., Y.L.) contoured regions of interest on the ventricular
septal myocardium and the blood pool on mid-ventricular short-axis images
independently. Peak SNR 3 and CNR which was defined as ($$\frac{SI_{myocardium-mean}-SI_{blood-mean}}{SI_{blood-SD}}$$) were calculated as
quantitative measurements.
Images
were blindly reviewed by two radiologists (L.H., L.R.), both with more than 5
years of experience in cardiovascular imaging. Overall image quality, blood
nulling, right ventricular (RV) free wall visibility, left ventricular (LV)
free wall visibility, and septum visibility were scored individually on
mid-ventricular short-axis images with a 5-point Likert scale as:
1-non-diagnostic; 2-poor; 3-fair; 4-good; 5-excellent 4.
In eleven patients with myocardial edema 5,
CNR-edema, which was defined as ($$\frac{SI_{edema-mean}-SI_{muscle-mean}}{SI_{blood-SD}}$$), was also calculated.
Statistical
Analysis:
All
data were described as mean ± standard deviation. Statistical analysis was
performed using SPSS (version 23.0, Chicago, IL). Ordinal Likert scores from
both observers were averaged prior to analysis, and differences in Conventional
T2W-DB and DL-based HR-T2W-DB scores were assessed using paired t-test and
paired Wilcoxon signed-rank test, respectively. Kendall's W was used to assess
inter-observer agreement of scoring data. ICC was used to assess inter-observer
agreement of CNR. P<0.05 was considered as statistically significant.Results
Peak
SNR and CNR-edema of DL-based HR-T2W-DB sequence were both significantly higher
than conventional T2W-DB (Table 2). Image quality scores of LV free wall
visibility in DL-based HR-T2W-DB sequence were also significantly higher than
conventional T2W-DB (Table 3). Overall image quality, blood nulling, RV free
wall visibility and septum visibility shows no statistical difference between
the two sequences. The results of CNR
measured by the two radiographers in DL-based HR-T2W-DB and conventional T2W-DB
were in good agreement (ICC 0.669 and ICC=0.677, respectively), and scoring
data also had good inter-observer agreement (Table 4).Discussion
The
DL-based HR-T2W-DB sequence doubles the resolution without increasing scan
duration or compromising image quality compared with conventional T2W-DB
sequence. In our study, deep learning based HR-T2W-DB sequence could achieve
better peak SNR, better CNR in region of edema and LV free wall visibility,
which might be benefit for detecting myocardial edema, especially in cases of
myocarditis to reveal LV lateral wall edema.Acknowledgements
No acknowledgement found.References
1. Renkuan Zhai, et al. Intelligent Incorporation of AI with Model Constraints for
MRI Acceleration. Proc. Intl. Soc. Mag. Reson. Med. 27 (2021): 1760.
2. Aowen Liu, et al. A lightweight and efficient convolutional neural network for
MR image restoration Proc. Intl. Soc. Mag. Reson. Med. 27 (2021): 1952.
3. Wang Z, et al.
Image quality assessment: from error visibility to structural
similarity[J].IEEE transactions on image processing : a publication of the IEEE
Signal Processing Society,2004,13(4):600-612.
4. Edelman RR,
Botelho M, Pursnani A, Giri S, Koktzoglou I. Improved dark blood imaging of the
heart using radial balanced steady-state free precession. J Cardiovasc
Magn Reson. 2016;18(1):69.
5. Friedrich MG,
Sechtem U, Schulz-Menger J, et al. Cardiovascular magnetic resonance in
myocarditis: A JACC White Paper. J Am Coll Cardiol.
2009;53(17):1475-1487.