Siyuan Liu1, Jiahao Lin1,2, Marvin Bergsneider3, Rock Hadley4, Giyarpuram N Prashant3, Sophie Peeters3, Kyunghyun Sung2, and Robert N Candler1,5
1Electrical and Computer Engineering, University of California Los Angeles, Los Angeles, CA, United States, 2Department of Radiological Sciences, University of California Los Angeles, Los Angeles, CA, United States, 3Department of Neurosurgery, University of California Los Angeles, Los Angeles, CA, United States, 4Department of Radiology, University of Utah, Salt Lake City, UT, United States, 5California NanoSystems Institute, Los Angeles, CA, United States
Synopsis
We simulate and experimentally verify
SNR improvement from a miniature flexible coil with the ultimate goal to
identify microadenomas in the pituitary gland that are currently undetectable. Through
scans of a phantom at multiple angles, we show a maximum of 12 to 20 times, and a
minimum of 2 to 4 times of SNR improvements compared to the commercial head coil.
We also perform high-resolution PD-TSE scans to confirm the visual improvement
of our flexible coil.
INTRODUCTION
There is a strong clinical need
to improve the resolution of pituitary MRI1,2. Cushing’s disease is a potentially
fatal disorder that is typically caused by small pituitary tumors,
microadenomas. The median size of microadenomas is 5mm3, and a
significant percentage of them are less than 3mm in size4. A novel
miniature flexible coil was recently developed to increase the SNR of pituitary
MRI by placing it close to the pituitary gland intra-operatively5.
In this study, we measure the potential SNR improvement of the flexible coil
compared to a commercial head coil, validated by both experiments and a
numerical simulation model.METHODS
Clinical proton density turbo
spin echo (PD-TSE) scans were performed for both the miniature coil
and a Siemens commercial head coil using 3T Siemens Prisma scanner (Siemens
Healthineers, Erlangen, Germany). An agar phantom and an acrylic resolution plate were designed for the
scanning, and a 3D printed igloo cavity was placed in the phantom to mimic the
sphenoid sinus (Figure 1). The coil was placed inside the cavity during the scanning. The SNR improvement
was measured with various rotation angles relative to the B0 field6.
Coil
Performance Simulation:
A 3D coil model was developed in
COMSOL Multiphysics® (COMSOLAB, Stockholm, Sweden) to study the
magnetic field distributions of the coil. The amplitude of the effective
transverse field at a rotation angle θ at
the resonance frequency was simulated and then compared with the MRI scan
results:
$$B_{1xy~effective}= \sqrt{ B_{1x}{B_{1x}}^{*} +(B_{1y}\cdot cos \theta- B_{1z}\cdot sin \theta){(B_{1y}\cdot cos \theta- B_{1z}\cdot sin \theta)}^{*} }$$
where B1x,
B1y, and B1z are the magnetic field components for our RF
receiving coil in x, y, z directions at θ = 0°.
SNR Measurements:
For the flexible coil, SNR
measurements were calculated for each angle from two repeated multi-slice 2D
clinical PD-TSE scans. The region of interest (ROI) was divided into five
cylindrical slices, each with 1cm diameter and 3mm thickness, with a size comparable to a typical pituitary gland7. SNR was
calculated with methods described by Constantinides et al. for magnitude images
of a single-coil array8.
For the 20-channel commercial
head coil, the SNR was calculated based on Kellman’s method for
root-sum-of-squares magnitude combining images9. Two repeated noise
only scans and two repeated clinical PD-TSE scans were used for the
calculation.
High-resolution MRI Scans:
To assess the visual improvement
of the flexible coil, a high-resolution 2D PD-TSE sequence of voxel size 0.2 × 0.2 × 0.7 mm3 was scanned for both the miniature coil and the commercial head coil. Images were reconstructed directly
via inverse Fast Fourier Transform (iFFT). The head coil images were
sum-of-square combined after coil reduction.RESULTS
The mean SNR of the ROI in the
phantom scan using the miniature coil were compared with the mean SNR using the head coil at various rotation angles and distances below the coil.
The factor of the SNR improvement is shown in Figure 2. The mean B1xy
effective field within
the ROI from the simulation was also plotted in Figure 2, normalized based on
the maximum field at 4.5 mm below the coil. The mean of
the simulated effective field agrees with the mean of scan experiment SNR with an error of 1.1% ± 0.8%.
An example of the hypothetical
miniature coil surgical placement with the simulated SNR improvement in the
pituitary region is shown in Figure 3. The gradient line represent the
SNR improvement factor of the coil compared to the head coil. In
this case, the pituitary gland enjoys a 12 to 20 times of SNR improvement at
the region close to the coil, and at least 2 to 4 times of SNR improvement at
the region further away.
To show the improved quality of
phantom images, high-resolution PD-TSE images from the head coil and the
miniature coil are shown in Figure 4. Though the SNR decreases with increasing
rotation angle, the image SNR is high enough to clearly show the 1 mm hole on
the resolution plate.
DISCUSSION
Comparison of SNR between the
miniature coil and the head coil demonstrates the potential of the miniature
coil to be used for high-resolution pituitary MRI. A maximum of 20-fold SNR
improvement is achieved.
Within θ = 60° and a depth of 16.5 mm, the
miniature coil maintains a reasonably high SNR for pituitary MRI that is at
least 2 times higher than the commercial head coil.
Validated with the scan experiment, this coil
simulation model is helpful for optimizing the coil designs for future studies
and predicting the SNR results for specific patients, which is much more
convenient and more cost and time-efficient than conducting MRI scans. Simulations could also be used to select from a
menu of coils to find the optimal
coil for each patient, as the anatomy of the sphenoid sinus can vary between
patients.CONCLUSION
A flexible miniature coil has
been developed for high-SNR pituitary gland MRI scanning. This work
demonstrated an improved SNR (between a factor of 12 and 20) and improved
quality of phantom images of the coil. The retrospective surgical placement example
showed the feasibility of using the simulation model
to streamline and accelerate the future coil designs.
Acknowledgements
This study was supported by the DGSOM Seed Grant
Program Award (The Spitzer Grant Research Program) and American Association of
Neurological Surgeons.References
1.Chittiboina P,
Talagala SL, Merkle H, et al. Endosphenoidal coil for intraoperative magnetic
resonance imaging of the pituitary gland during transsphenoidal surgery. J
Neurosurg. 2016;125(6):1451-1459. doi:10.3171/2015.11.JNS151465
2.Law M, Wang R,
Liu CSJ, et al. Value of pituitary gland MRI at 7 T in Cushing’s disease and
relationship to inferior petrosal sinus sampling: Case report. J Neurosurg.
2018;1-5. doi:10.3171/2017.9.JNS171969
3.Jagannathan
J, Smith R, Devroom HL, et al. Outcome of using the histological pseudocapsule
as a surgical capsule in Cushing disease. J
Neurosurg. 2009;111(3):531-539. doi:10.3171/2008.8.JNS08339
4.Hamid
O, Fiky L El, Hassan O, Kotb A, Fiky S El. Anatomic Variations of the Sphenoid
Sinus and Their Impact on Trans-sphenoid Pituitary Surgery. Skull Base. 2008;18(1):9-15.
doi:10.1055/S-2007-992764
5.Lin
J, Hadley R, Li L, Merrill R, Bergsneider M, Candler R, Sung K. Development of a Local
Pituitary Coil for Assessment of Pituitary Microadenomas. Proc ISMRM. 2019.
6.Lin
J, Liu S, Hadley R, Bergsneider M, Prashant G, Peeters S, Candler R, Sung K.
Exploration of the Surgical Placement of the Local Pituitary Coil for
Microadenomas. Proc ISMRM. 2021.
7. Amar
AP, Weiss MH. Pituitary anatomy and physiology. Neurosurg Clin N Am. 2003;14(1):11-v.
doi:10.1016/S1042-3680(02)00017-7
8.Constantinides
CD, Atalar E, McVeigh ER. Signal-to-noise measurements in magnitude images from
NMR phased arrays. Magn Reson Med.
1997;38(5):852-857. doi:10.1002/mrm.1910380524
9. Kellman
P, McVeigh ER. Image reconstruction in SNR units: a general method for SNR
measurement. Magn Reson Med.
2005;54(6):1439-1447. doi:10.1002/mrm.20713