Tao Jin1 and Julius Juhyun Chung1
1University of Pittsburgh, Pittsburgh, PA, United States
Synopsis
Endogenous CEST signal
usually has low specificity due to contamination from the magnetization
transfer effect and from fast exchanging labile protons with close Larmor
frequencies. We propose to improve CEST signal specificity with an ASEF which
measures the difference between CEST signals acquired with similar average saturation
power but largely different duty cycles (DC), e.g., a continuous wave or a high
DC pulse train versus a low DC one. Simulation and creatine phantom studies showed
that ASEF can improve the specificity of slow to
intermediate exchanging CEST signal with a relatively small loss of
sensitivity.
Introduction
CEST MRI signal is
usually contaminated by magnetization transfer (MT) from semi-solids and
overlapping faster, broad exchanges. Numerous solutions have been proposed to
remove confounding effects but these methods often either increase acquisition
time or sacrifice sensitivity1-4. We propose an Average Saturation
Efficiency Filter (ASEF) which uses two pulse trains with similar average saturation power but highly
unequal duty cycles, so that saturation transfer effects from fast chemical
exchange species and semisolid macromolecules are similar between the two pulse
trains, but drastically different for slow exchange species. Thus, their difference
becomes an exchange rate filter suppressing MT and fast exchange signals. In
this work, the signal properties of
ASEF were evaluated by computer simulation and validated by phantom
experiments.
Methods
Simulations: Assuming
a train of block pulses where the RF in each repeating unit has a bipolar pulse
(which minimizes the rotation effect 2) with duration tp (Fig. 1). ASEF takes
the difference between CEST signals measured by a continuous wave (or high duty
cycle) saturation and a low duty cycle (DCl) one. CEST signals were simulated by Bloch-McConnel
Equations which include 3 exchanging pools of free water protons, labile
protons, and bound water protons, assuming a chemical shift between the labile
proton and water of 1.9 ppm, a fraction of labile proton of 0.001, the T1
(T2) of water, labile proton, and bound water protons is 2 s (66.6
ms), 2 s (66.6 ms), and 2 s (10 ms), respectively.
Phantom experiments: MR
experiments were performed at 9.4 T. Seven phantoms were prepared in 10% Bovine
Serum Albumin (BSA). 50 mM Creatine
was added to six of them and they were then titrated to pH = 6.15, 6.55, 7.04,
7.44, 7.85, 8.25, and 7.0 for the BSA only phantom. These phantoms were heated to
95ºC to denature the BSA and measured at room temperature. The saturation preparation
schemes consisted of either a single CW block pulse or a DCl = 15%
and tp = 24 ms bipolar pulse train, with a duration of 6-s. The B1
was 0.8 μT for the CW and a nominal average B1
= 0.82 μT was determined for the DCl pulse
train. This 2.4% fudge factor was obtained by matching the saturated signal at
6 ppm (where the CEST effect is minimal) to minimize the mismatch due to our RF
power linearity/stability and the mismatch in the MT effect expected from two
saturations with highly unequal duty cycles 5. Images were acquired by single slice
spin-echo EPI. Results
Fig. 1B shows simulated exchange rate filtering of ASEF. The
CEST contrast of the DCl pulse train is like that of CW at fast
exchange rates but is much smaller at slower exchange rates. As a result, the
sensitivity of their difference, i.e., the ASEF signal, is only slightly lower
than CW for slow exchange rates but is suppressed at kex > 2000 s-1.
For larger DCl values, the sensitivity of ASEF decreases but the exchange
rate filtering (e.g., peak position and the linewidth) remains the same (Fig. 1C).
With a higher B1, avg = 1.6 µT, the ASEF signal peak shifts to
a higher kex, and CEST contrast is suppressed for kex >
4000 s-1 (Fig. 1D). For creatine in heated-denatured BSA, the
Z-spectra of the pH = 7.04 phantom show larger differences in the 1.9 ppm dips
for CW than DCl pulse train saturation (Fig.2A). The spectra match
well for offsets > 3.5 ppm indicating the MT effect can be effectively minimized
by ASEF. In Fig. 2B, the CEST contrast of CW saturation is calculated by
subtracting the signals of phantoms with creatine from that of BSA only. The
ASEF signal is only slightly lower than that of CW at lower pH values, but the
difference is much larger at higher pH. Note MTRasym showed a negative
baseline of about -2% due to the intrinsic asymmetry of the MT effect, as shown
in the BSA-only phantom (Fig. 2C). In contrast, the ASEF signal is minimal for
the phantom of BSA only and is small for pH = 8.24 (Fig. 2D), indicating that
signals from both the MT contrast and from fast chemical exchange can be
effectively suppressed. Discussions
ASEF can improve the CEST signal specificity for slow to
intermediate exchanges (e.g., kex <2000 s-1) with only
a small reduction in the sensitivity. It can be acquired at as few as only one
frequency, i.e., the Larmor frequency of the labile proton of interest. ASEF has three parameters that can be adjusted:
the B1,
avg determines the range of
exchange rate filtering, the duty cycles of the pulse trains determine ASEF
sensitivity, and tp which is sensitive to direct water saturation. The
MT mismatch5 between the low and high DC saturations may be
minimized by careful selection of these parameters, as well as a fudge factor
of the B1,
avg between the two
saturations.
Conclusion
ASEF is a simple method that can minimize the
MT effect and provide fast exchange rate filtering for CEST MRI with a
relatively small reduction in sensitivity. It can be a highly useful tool for
CEST study in the slow to intermediate exchange regime.Acknowledgements
This work is supported
by NIH grant NS100703.References
1. Zaiss M,
et al. J Magn Reson 2011;211(2):149-55.
2. Zu Z, et
al. Magn Reson Med 2013;69(3):637-647.
3. Xu X, et
al. Magn Reson Med 2016;75(1):88-96.
4. Chen L,
et al. Magn Reson Med 2019;81(1):69-78.
5. Varma G,
et al. J Magn Reson 2018;296:60-71.