Saeed Jerban1, Alecio Lombardi1, Yajun Ma1, Amir M Afsahi1, Dina Moazamian1, Jiyo Athertya1, Hyungseok Jang1, Christine B Chung1, Jiang Du1, and Eric Y Chang1,2
1Department of Radiology, University of California, San Diego, San Diego, CA, United States, 2Radiology Service, VA San Diego Healthcare System, San Diego, CA, United States
Synopsis
Entheses are transition zones connecting
flexible tissues such as tendons and ligaments to bone. Conventional MRI
sequences detect little or no signal from entheses as a result of their short
T2* values. Alternatively, ultrashort echo time MRI (UTE-MRI) with TE<50 μs
can be used to image tissues with short T2 and T2* for quantitative assessment.
The feasibility
of using quantitative UTE-MRI techniques for the
evaluation of entheses regions of Achilles tendons in cadaveric ankle specimens
was investigated. Using 20 cadaveric ankle specimens, high quality images were
obtained with significant entheses signal which enabled excellent UTE-T1,
UTE-Adiab-T1r, and
UTE-MT model fittings.
INTRODUCTION
Entheses are transition zones connecting flexible tissues
such as tendons and ligaments to bone, with potential mechanical stress
concentration as a result of the tissues’ widely different mechanical
properties. Entheses may be involved in many diseases including the overuse
syndromes and the early stages of osteoarthritis. Conventional MRI sequences
detect little or no signal from entheses as a result of their short T2* values.
Particularly, normal entheses have not been identifiable with clinical MR
imaging.
Fortunately, ultrashort echo time MRI (UTE-MRI) with
TE<50 μs can be used to image tissues with short T2 for quantitative
assessment (1).
UTE-MRI combined with magnetization transfer (MT) modeling (UTE-MT) and
adiabatic spin-locking T1r
(UTE-Adib-T1r)
have recently been introduced as orientation insensitive techniques for compositional
evaluation of different musculoskeletal tissues(2,3).
Both UTE-Adiab-T1ρ and
UTE-MT require T1 compensation which can be performed more accurately using actual
variable flip angle (AFI-VFA) techniques of short T2 tissues (4).
This study was aimed to investigate the
feasibility of using UTE-T1, UTE-Adiab-T1ρ, magnetization transfer ratio
(MTR) imaging, and UTE-MT modeling for quantitative evaluation of enthesis
region of Achilles tendon in cadaveric ankle specimens.METHODS
Twenty human cadaveric ankle specimens (65±17
years old) without any known musculoskeletal diseases were imaged with
UTE-MRI sequences on a 3T clinical scanner (MR750, GE). Imaging was performed in
the sagittal plane centered at the enthesis region connecting the Achilles
tendon to the calcaneus bone. To measure T1 as a prerequisite for
the two-pool UTE-MT modeling and UTE-Adiab-T1ρ fitting, AFI-VFA
sequence (AFI: TE=0.032ms, TRs=20,100ms, FA=45˚; VFA: TE=0.032ms, TR=20ms, FAs=5,
12, 24˚) was performed (5).
A 3D-UTE-Cones-MT sequence (pulse power=400° and 800°; frequency offset=2, 5,
10, 20, and 50kHz; FA=7˚; 11 spokes per MT preparation) was performed for the
two-pool MT modeling (3,6).
Additionally, a set of 3D UTE-Adiab-T1r sequences
with five different spin-locking times (TSLs) (TR=500ms, TSL= 0, 12, 24, 36, and
48 ms) were performed (2). Field of view, matrix
dimension, pixel size, slice thickness, and total scan time were 11cm, 256×256,
2mm, and 60 mins, respectively.
UTE-MRI analyses were performed within a region of interest (ROI)
covering the entheses. ROI selection was performed by an experienced image
analyst and confirmed by a musculoskeletal radiologist. For each specimen,
average macromolecular proton fraction (MMF) and T2 of the macromolecules
(T2mm) were estimated using UTE-MRI-MT modeling performed on two sagittal
slices 4 mm apart. UTE-T1 and UTE-Adiab-T1ρ values
were calculated for each specimen using a single exponential fitting model.Results
The UTE-MRI sequence acquired high signal from the entheses
of all scanned specimens. Figure 1A shows schematically a representative ROI
selected on a UTE-Adiab-T1ρ image performed on the sagittal plane of an ankle specimen from a 62-year-old
female donor. Representative UTE-T1, UTE-Adiab-T1ρ, and two-pool MT model fittings
are demonstrated in Figure 1B-C, respectively for the selected ROI. Figure 2
illustrates the representative T1, T1ρ, and MMF pixel maps generated over Achilles tendon including the
enthesis area. Average, standard deviation (SD), maximum and minimum values for
all the investigated UTE-MRI parameters are presented in Figure.3 (Table 1).DISCUSSION
The feasibility
of using UTE-T1, UTE-Adiab-T1ρ, magnetization transfer ratio (MTR) imaging, and
UTE-MT modeling for the evaluation of enthesis region of Achilles tendon in
cadaveric ankle specimens were investigated. Acquired high quality images with
significant signal in enthesis enabled excellent UTE-T1, UTE-Adiab-T1ρ, and UTE-MT model fittings. The range of the
calculated MMF, UTE-T1, UTE-Adiab-T1ρ values in enthesis were roughly close to the reported
values for cadaveric articular cartilage and tendons in the literature. Future
investigations are required to understand the potential correlations between UTE-MRI
parameters of enthesis and its disease stage, mechanical properties,
histological characteristics.CONCLUSION
This study highlighted the UTE-MRI techniques as useful quantitative methods to assess enthesis
regions of tendons and ligaments which possess short T2* values and are not
identifiable using conventional MRI techniques.Acknowledgements
The authors acknowledge grant support from the NIH (R01AR068987,
R01AR075825, R01AR062581), Veterans Affairs (I01RX002604 and I01CX001388), and GE Healthcare. References
1. Chang EY, Du J, Chung
CB. UTE imaging in the musculoskeletal system. J. Magn. Reson. Imaging 2015;41:870–883
doi: 10.1002/jmri.24713.
2. Ma Y, Carl M, Searleman A, Lu X, Chang EY, Du J. 3D adiabatic
T1ρprepared ultrashort echo time cones sequence for whole knee imaging. Magn.
Reson. Med. 2018;80:1429–1439 doi: 10.1002/mrm.27131.
3. Ma Y, Chang EY, Carl M, Du J. Quantitative magnetization transfer
ultrashort echo time imaging using a time-efficient 3D multispoke Cones
sequence. Magn. Reson. Med. 2017;79:692–700 doi: 10.1002/mrm.26716.
4. Ma Y, Lu X, Carl M, et al. Accurate T 1 mapping of short T 2 tissues
using a three-dimensional ultrashort echo time cones actual flip angle
imaging-variable repetition time (3D UTE-Cones AFI-VTR) method. Magn. Reson.
Med. 2018;80:598–608 doi: 10.1002/mrm.27066.
5. Ma YJ, Zhao W, Wan L, et al. Whole knee joint T1 values measured in
vivo at 3T by combined 3D ultrashort echo time cones actual flip angle and
variable flip angle methods. Magn. Reson. Med. 2019;81:1634–1644 doi:
10.1002/mrm.27510.
6. Ma Y, Tadros A, Du J, Chang EY. Quantitative two-dimensional ultrashort
echo time magnetization transfer (2D UTE-MT) imaging of cortical bone. Magn.
Reson. Med. 2018;79:1941–1949 doi: 10.1002/mrm.26846.