Hahnsung Kim1,2, Lisa C. Krishnamurthy 3,4, Kimberly B Hoang5, Ranliang Hu2, and Phillip Zhe Sun1,2
1Yerkes Imaging Center, Yerkes National Primate Research Center, Emory University, Atlanta, GA, United States, 2Department of Radiology and Imaging Sciences, Emory University School of Medicine, Atlanta, GA, United States, 3Center for Visual and Neurocognitive Rehabilitation, Atlanta VA, Decatur, GA, United States, 4Department of Physics & Astronomy, Georgia State University, atlanta, GA, United States, 5Department of Neurosurgery, Emory University School of Medicine, Atlanta, GA, United States
Synopsis
The multi-slice CEST signal evolution was
described by the spin-lock relaxation during saturation duration (Ts) and longitudinal relaxation
during the
relaxation delay time (Td) and post-label
delay (PLD), from which the QUASS
CEST was generalized to fast multi-slice acquisition. In addition, normal human subjects and tumor patients scans were performed to compare the conventional
apparent and QUASS CEST measurements with different Ts, Td,
and PLD. Bland-Altman analysis bias of the proposed QUASS
CEST effects was much smaller than the PLD-corrected apparent
CEST effects (0.03% vs. -0.54%),
indicating the proposed fast multi-slice CEST imaging is robust and
accurate.
Introduction
Chemical exchange saturation transfer (CEST) MRI is
sensitive to detecting tissue changes following acute stroke, tumors, and
epilepsy1-3. CEST MRI is versatile, but the CEST measurement depends not
only on labile proton concentration and exchange rate but also on experimental
parameters, such as the duration of radiofrequency (RF) saturation (Ts)
and the relaxation recovery delay (Td). Although the use of long RF
irradiation enhances the CEST effect, it unavoidably prolongs the total scan
time, making CEST MRI challenging for routine clinical use. For typical human
CEST scans, the RF saturation time can range from 0.2 to 3.5 s at 3 T3-5. Therefore,
CEST protocols vary substantially among centers, and it is urgent to unify
results obtained under different Ts and Td. A
quasi-steady-state (QUASS) CEST algorithm has been recently proposed to reduce impacts
of experimental Ts and Td parameters on the CEST
measurement6. We hypothesized that the QUASS algorithm can be generalized for
the multi-slice acquisition in addition to Ts and Td
correction and tested it for fast multi-slice imaging in tumor patients.Theory
A schematic diagram of the fast
multi-slice spin echo (SE) echo planar imaging (EPI) CEST MRI sequence is shown
in Figure 1. The apparent CEST Z-spectrum is
calculated by normalizing the saturated scan signal ($$$I'_{sat}$$$) with the unsaturated control
scan signal ($$$I'_{0}$$$)7:
$$\frac{I'_{sat}\left(\Delta\omega\right)}{I'_{0}}=\frac{\left(1-e^{-R_{1w}T_{d}}\right)e^{-R_{1\rho}\cdot T_{s}}+\frac{R_{1w}}{R_{1\rho}}\cos^{2}\theta\left(1-e^{-R_{1\rho}\cdot T_{s}}\right)}{1-e^{-R_{1w}\cdot \left(T_{s}+T_{d}+pld\right)}}e^{-R_{1w}\cdot pld}+\frac{1-e^{-R_{1w}\cdot pld}}{1-e^{-R_{1w}\cdot \left(T_{s}+T_{d}+pld\right)}}\qquad\qquad\left[1\right]$$
where $$$R_{1w}$$$ is the bulk water longitudinal relaxation
rate, $$$I_{0}$$$ is the equilibrium magnetization, $$$R_{1\rho}$$$ is the spin-lock relaxation rate
and $$$\theta=\arctan\left(\frac{\gamma B_{1}}{\Delta \omega}\right)$$$, in which $$$\gamma$$$ is the gyromagnetic ratio and B1 and $$$\Delta \omega$$$ are the amplitude and offset frequency
of the RF saturation, respectively. Under the assumption that Ts and
Td are much longer than the post-label delay (PLD) time, the
PLD-corrected apparent CEST Z-spectrum can be derived as,
$$\frac{I'^{pldcor}_{sat}\left(\Delta \omega\right)}{I'^{pldcor}_{0}}\cdot \left\{\frac{1-e^{-R_{1w}\left(T_{s}+T_{d}\right)}}{1-e^{-R_{1w}\cdot T_{d}}}\right\}=e^{-R_{1\rho}\cdot T_{s}}+\frac{R_{1w}\cdot \cos^2 \theta}{R_{1\rho}\cdot \left(1-e^{-R_{1w}\cdot T_{d}}\right)}\cdot \left(1-e^{-R_{1\rho}\cdot T_{s}}\right)\qquad\qquad\left[2\right]$$
in
which the superscript pldcor denotes
the PLD-corrected apparent signals. $$$R_{1\rho}$$$ can be numerically solved, and the
QUASS-corrected CEST Z-spectrum can be calculated as
$$\left(\frac{I_{sat}\left(\Delta \omega\right)}{I_{0}}\right)^{QUASS}=\frac{R_{1w}}{R_{1\rho}}\cos^{2}\theta\qquad\qquad\left[3\right]$$
Methods
Informed written consent was obtained following IRB-approved protocol. Three healthy volunteers (1 male and 2 females, age 25-39) and a brain tumor patient underwent CEST brain MRI using a 3T MR scanner (Magnetom Prisma, Siemens Medical Solutions, Erlangen, Germany) with a 64-channel head and neck coil. Two sets of multi-slice brain images with different pairs of saturation duration and relaxation delays (Ts/Td =1 s/1 s and 2 s/2 s) were acquired in the axial orientation. Multiple rectangular RF pulse (duration = 499.8 ms, duty cycle = 99.9 %, B1 = 0.7 µT) were exploited to emulate continuous wave saturation. The alternating offset frequencies of RF saturation were varied from -5 to 5 ppm, with increments of 0.125 ppm. The imaging parameters were: imaging readout time = 547 ms, TE = 33 ms, FOV = 220 × 220 mm2, in-plane matrix = 110 × 110, the number of slices = 8, interleaving slice ordering, slice thickness = 5 mm with 25% slice gap, with fat suppression, 1 average, and readout bandwidth = 1976 Hz/pix. Total imaging time was 3 min 31 s and 6 min 17 s for Ts/Td of 1 s/1 s and Ts/Td of 2 s/2 s, respectively. T1w-weighted images were acquired using eight inversion recovery delay (0.25, 0.5, 0.75, 1.0, 1.25, 1.5, 2.0, and 3.0 ms) with relaxation delay of 5.0 s, and 2 average.Results/Discussion
Figure 2 shows CESTR′,
PLD-corrected CESTR′,
and CESTRQUASS of healthy human brain images at 3.5 ppm with Ts/Td of 1/1 s and 2/2 s. The
apparent CEST effects noticeably differed in the direction of the multi-slice
acquisition and under different Ts and Td (Figs. 2A, B).
Although PLD-corrected apparent CEST effects showed little PLD dependency,
there was a noticeable CEST contrast discrepancy between Ts/Td
of 1/1 s and 2/2 s (Figs. 2C, D). In contrast, the QUASS CEST effects were more
consistent between Ts/Td of 1/1 s and 2/2 s (Figs. 2E, F). Specifically, for WM, the
PLD-corrected CESTR′ was 3.32±1.09% and
3.83±1.11% for Ts/Td of 1/1 s and 2/2 s, respectively,
while CESTRQUASS being 4.91±1.23% and 4.87±1.20%. In GM, the PLD-corrected CESTR′ was 2.00±1.09% and 2.54±1.09%
for Ts/Td of 1/1 s and 2/2 s, respectively, while their
corresponding CESTRQUASS being 3.36±1.56% and 3.28±1.34%. In Figure 3, the Bland-Altman analysis
bias was -0.54% for PLD-corrected CESTR′ and 0.03% for CESTRQUASS,
indicating that the proposed QUASS algorithm minimizes the bias in the CEST
measurement, advantageous over simple PLD correction. Compared with the
contralateral normal tissue, the tumor region shows greater CESTR in Figure 4.
In particular, the apparent CEST effect increased with the increase
of Ts and Td, while the QUASS CEST effects showed little
Ts and Td dependency.Conclusion
Our study demonstrated
fast multi-slice CEST brain MRI with healthy volunteer and brain tumor patient scans. In addition, the QUASS algorithm minimizes the effect of the
choice of saturation duration, relaxation delay, and post-label delay time on
the CEST MRI quantification. In summary, the QUASS algorithm enables an
expedited CEST MRI acquisition strategy, thereby increasing its clinical
translatability, particularly at high magnetic fields, while providing accurate
steady-state CEST measurement.Acknowledgements
No acknowledgement found.References
[1]
Sun PZ et al., Detection of the ischemic penumbra using pH-weighted MRI.
J Cereb Blood Flow Metab 2007;27(6):1129-1136.
[2] Daivis KA et al., Glutamate imaging (GluCEST)
lateralizes epileptic foci in nonlesional temporal lobe epilepsy. Sci Transl Med. 2015;7(309):309ra161.
[3] Togao O et al., Amide proton transfer imaging
of adult diffuse gliomas: correlation with histopathological grades. Neuro
Oncol. 2014;16(3):441-448.
[4] Harston GW et al., Identifying the ischaemic
penumbra using pH-weighted magnetic resonance imaging. Brain 2015;138(1):36-42
[5] Goerke S et al., Clinical routine acquisition
protocol for 3D relaxation‐compensated APT and
rNOE CEST‐MRI of the human brain at 3T. Magn Reson Med
2021.
[6] Sun PZ., Quasi‐steady
state chemical exchange saturation transfer (QUASS CEST) analysis—correction of
the finite relaxation delay and saturation time for robust CEST measurement.
Magn Reson Med 2021;85(6):3281-3289.
[7] Jin T, et al., Spin‐locking
versus chemical exchange saturation transfer MRI for investigating chemical
exchange process between water and labile metabolite protons. Magn Reson Med
2011;65(5):1448-1460.