Hao Li1, Charlotte E Buchanan2, David M Morris3, Alexander J Daniel2, João Sousa4, Steven Sourbron4, David L Thomas5,6,7, Susan T Francis2, and Andrew Nicholas Priest1,8
1Department of Radiology, University of Cambridge, Cambridge, United Kingdom, 2Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, United Kingdom, 3Centre for Cardiovascular Science, University of Edinburgh, Edinburgh, United Kingdom, 4Department of Infection, Immunity and Cardiovascular Disease, University of Sheffield, Sheffield, United Kingdom, 5Neuroradiological Academic Unit, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom, 6Dementia Research Centre, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom, 7Wellcome Centre for Human Neuroimaging, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom, 8Department of Radiology, Addenbrooke’s Hospital, Cambridge, United Kingdom
Synopsis
B1
inhomogeneity and non-ideal slice profiles introduce contributions from
stimulated and indirect echoes into the multi-echo spin-echo (MESE) T2
mapping sequence, leading to variations in quantitative T2 values
across scanners and vendors despite the use of a harmonised scan protocol. This
study used an EPG-based fitting method to include corrections for stimulated-echo
effects, applied to phantom and ‘travelling kidney’ data collected on scanners from
three different MR vendors (GE, Philips, Siemens). Compared with conventional monoexponential
fitting, EPG-based fitting substantially reduced the inter-scanner variations
of T2 measurements. This improves the harmonization of the MESE T2
mapping sequence across MR scanner vendors.
INTRODUCTION
T2
mapping has been proposed as a potential tool for renal disease evaluation1. However, B1
field inhomogeneities and imperfect slice selection pulse profiles can lead to
simulated and indirect echoes for a multi-echo spin-echo (MESE) sequence. This can
impair T2 estimation accuracy and cause inconsistencies in measured
T2 between different scanners with varying hardware and protocol
implementations—which would be problematic for multi-centre clinical trials.
Model-based
methods using the extended phase graph (EPG) algorithm2 have been
developed, which can estimate the transmit field and incorporate stimulated
echo contributions for accurate T2 estimation3. This study
aims to harmonize renal T2 measurements from a ‘travelling kidney’
study including three MR vendor scanners, using an EPG-based fitting method to
reduce cross-vendor variations in both volunteer and phantom measurements.METHODS
Data
was acquired on six healthy volunteers (age 32±6 yrs), with four scanned on
each of the three 3T scanners of different vendors (GE, Philips and Siemens;
randomly labelled Vendor 1, 2 and 3), while two volunteers only underwent scans
on two vendors. Only one scanner (vendor 1) was equipped with a dual-channel
transmit system. The ISMRM/NIST system phantom4 was used to
evaluate the accuracy of T2 measurements against known reference
values.
Following
initial harmonisation work5, a
respiratory-triggered MESE sequence was harmonised across vendors: TR 1 breath
(min 3000 ms), TE
12.9–129 ms in 12.9 ms steps, echo train length 10, refocusing flip angle 180°, FOV
38.4 cm, acquired matrix 128×128, 5 slices with thickness/gap of 4.5/1.0 cm, parallel
imaging factor 3.
To evaluate T2
variations caused by B1 inhomogeneity and validate the flip angle
maps estimated by the EPG model, separate B1 maps were acquired
using the available B1 mapping scheme for each vendor (DREAM/
TurboFLASH B1 mapping /Bloch-Siegert method)
(5). The
acquired B1 maps were converted into the range
[0, 1] to compare with the estimated B1
maps from the StimFit cB1 = (1− abs(FAnominal − FAactual)/ FAnominal).
Masks of whole kidneys, cortex and
medulla were acquired based on harmonised native 5(3)3 MOLLI T1 map
images and applied to the T2 maps for quantitative evaluation.
The
StimFit toolbox based on the EPG algorithm3,6 was used as
a fitting model with stimulated echo compensation and compared with conventional
monoexponential fitting using nonlinear least-squares.
The EPG simulation used vendor-specific information about RF and gradient pulse
shapes and timing.RESULTS
Figure 1 compares the T2
measurement on the NIST phantom across the different
vendors using exponential fitting and StimFit. The mean absolute percentage
error (MAPE) was calculated in the physiologically relevant range (45–286 ms). Compared with exponential
fitting, StimFit reduced the MAPE from 11.0%, 5.5% and 19.3% to 6.1%, 4.2% and
4.6% for the three vendors respectively.
Figure 2 shows example T2 maps and cB1 maps from
the same volunteer collected across the three vendors. In regions
where the flip angle is close to the nominal value, the T2 maps
agree well between fitting methods and between vendors. However, flip angle
variations (non-ideal B1) caused overestimation by exponential fitting
in the right kidney for vendor 1, the upper left kidney for vendor 2 and both
kidneys for vendor 3. For vendor 3, in particular, a widespread low B1
(with 82.7% of the nominal flip angle) caused a global overestimate of T2
in all subjects. All these variations are largely corrected using StimFit which
gives much more consistent T2 values between kidneys and between
vendors. Also, the cB1 maps estimated by StimFit show similar
features to the measured B1 maps.
Similar results were found for the other five volunteers. Figure
3 shows scatterplots of the T2
measurements in the left and right kidneys; the values are much more consistent,
between kidneys and between vendors, for StimFit than for exponential fitting.
The T2 values in the cortex and medulla are summarised in Table 1 for the three
vendors. The coefficients of variation (CV) across vendors were reduced from 8.13%
(cortex) and 8.63% (medulla) with exponential fitting to 2.41% and 2.22% with
StimFit. Corresponding reductions in inter-vendor bias and variance canT2 values also be
observed in Bland-Altman plots showing the agreement between different vendors using
the two fitting methods (Figure 4).Discussion
We demonstrate a large variance in renal T2 mapping using a
harmonised MESE scheme with exponential fitting, with up to a ~30 ms difference in whole kidney T2
in the same volunteer between vendors. This would substantially impair the reliability
of T2 mapping as a potential disease biomarker.
Variations
observed between scans were all corrected by StimFit without additional knowledge or data input. The inter-vendor variations we observed are not necessarily
specific to all scanners from these vendors, and can also relate to
configuration issues such as using single or parallel transmit systems. The
specific Vendor 3 scanner used in this study suffered B1
miscalibrations for all subjects, but this issue is less severe in our other
measurements on different Vendor 3 systems. Future research will further investigate factors related to
cross-site variation, as well as increasing the number of ‘travelling kidney’
scans.CONCLUSION
Fitting
with stimulated echo correction reduces inter-scanner T2 variations
to aid the harmonization of the MESE T2 mapping measures across MR
scanner vendors.Acknowledgements
This work was supported by the UK Medical Research Council
(MR/R02264X/1) and NIHR Cambridge Biomedical Research Centre (BRC-1215-20014). The
authors acknowledge Dr Ali Aghaeifar from Siemens Healthineers for providing RF pulse parameters.References
1. Wolf M, de Boer A, Sharma K, et al. Magnetic resonance
imaging T1- and T2-mapping to assess renal structure and function: a systematic
review and statement paper. Nephrol. Dial. Transplant 2018;33:ii41–ii50 doi:
10.1093/ndt/gfy198.
2.
Hennig J. Multiecho imaging sequences with low refocusing flip angles. J. Magn.
Reson. 1988;78:397–407 doi: 10.1016/0022-2364(88)90128-X.
3.
Lebel RM, Wilman AH. Transverse relaxometry with stimulated echo compensation.
Magn. Reson. Med. 2010;64:1005–1014 doi: 10.1002/mrm.22487.
4.
Stupic KF, Ainslie M, Boss MA, et al. A standard system phantom for magnetic
resonance imaging. Magn. Reson. Med. 2021;86:1194–1211 doi: 10.1002/mrm.28779.
5.
Charlotte Buchanan, Hao Li, Fabio Nery, Alexander Daniel, João De Sousa, Steven
Sourbron, Andrew N Priest, David L Thomas SF. Harmonisation of Multiparametric
Renal MRI for Multi-Centre Studies. In: The 29th Annual Meeting of the
International Society for Magnetic Resonance in Medicine. ; 2021.
6.
Lebel RM. StimFit: A toolbox for robust T2 mapping with stimulated echo
compensation. In: In: Proceedings from the 20th Annual Meeting of ISMRM,
Melbourne, Australia. Vol. 37. ; 2012. p. 2558.