Dan Zhu1,2 and Qin Qin1,2
1F.M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, MD, United States, 2The Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, MD, United States
Synopsis
The traditional RF-prepared B1
mapping technique consists of one scan with an RF preparation module for
FA-encoding and a second scan without this module for normalizing. To
reduce the T1-induced k-space filtering effect, this method is
limited to 2D FLASH acquisition with a two-parameter method. Based on the point spread function
analysis of FLASH readout, a novel 3D RF-prepared three-parameter method for B1-mapping is
proposed to correct the T1-induced
quantification bias. The
proposed technique was compared with existing methods in the brain, breast, and abdomen
and demonstrated with high accuracy for ultrafast B1 mapping.
Introduction
Rapid
measurement of the spatial variation of the flip angles (FA) for the organs of
interest is desired for many MRI applications. The RF-prepared
two-parameter method1,2 starts
with a B1-insensitive saturation module and a fixed
saturation delay, then applies an RF preparation module for
B1-encoding in the first scan and removes this module for normalizing in
the second scan, both followed with FLASH acquisition. To reduce the T1-induced
k-space filtering effect, this method is limited to a short 2D readout. In this
work, a novel 3D RF-prepared method for ultrafast B1-mapping is proposed with a
third scan to record and
correct the bias from the k-space filtering
effect (three-parameter method). We compared the 3D RF-prepared method with existing B1
mapping methods in the brain, abdomen and breast with different RF-shim settings at
3T. Theory
The proposed protocol is illustrated in
Figure 1. The ratio between the longitudinal
magnetization after an RF-prep pulse ($$$M_{FA}$$$) with FA=$$$\alpha$$$ and the normalizing
magnetization without this RF-prep pulse ($$$M_{norm}$$$) is:$$\frac{M_{FA}}{M_{norm}}=\frac{M_{norm}\cdot\cos{(\alpha)}}{M_{norm}}=\cos{(\alpha)}.\tag{1}$$
The point spread function analysis of
FLASH shows that the prepared longitudinal magnetization before the FLASH
acquisition ($$$M_{FA}$$$, or $$$M_{norm}$$$) and the image signal ($$$S_{FA}$$$,
or $$$S_{norm}$$$) obeys a linear (not proportional) relationship (slope=$$$a$$$,
intercept=$$$b$$$). $$S_{norm}=a\cdot M_{norm}+b\tag{2}$$$$S_{FA}=a\cdot M_{FA}+b\tag{3}$$
In this work, we propose to obtain the
intercept b via a third image, by acquiring the signal immediately after the saturation
module: $$S_{sat}=a\cdot 0+b=b.\tag{4}$$
Using a three-parameter method, the bias
in FA estimation can thus be corrected:$$\frac{S_{FA}-S_{sat}}{S_{norm}-S_{sat}}=\frac{a\cdot M_{norm}\cdot\cos{(\alpha)}+b-b}{a\cdot M_{FA}+b}=\cos{(\alpha)}.\tag{5}$$
The relationship of the actual FA ($$$\alpha$$$) and the nominal FA ($$$\alpha^{nom}$$$) is
usually described with a
scale factor $$$\kappa$$$ (with 100% as consistent between the two): $$\kappa=\frac{\alpha}{\alpha^{nom}}\times100\%=\frac{\cos^{-1}{\left(\frac{S_{FA}-S_{sat}}{S_{norm}-S_{sat}}\right)}}{\alpha^{nom}}\times100\%.\tag{6}$$
It is important to note that subtractions
between $$$S_{FA}$$$, $$$S_{norm}$$$ and $$$S_{sat}$$$ should consider
opposite polarities and complex subtraction is preferred in this case.Methods
In vivo datasets of three brains, three
breasts and five abdomens were obtained from six volunteers (4 females, 25-45yrs),
performed on a 3T Philips Ingenia scanner with a dual-source parallel RF
excitation system. A 32-channel head coil, a 16-channel bilateral breast coil
and a 32-channel chest coil were used for signal reception, respectively.
For the three-parameter method, the
RF-prep utilized $$$\alpha^{nom}$$$=60°. The FLASH
acquisition had a low-high ordering. For brain scans: TR/TE=3.4/2.3ms, FOV=220×220×120mm3,
resolution=4×4×4mm3, readout bandwidth(BW)=2975Hz/pixel. The number
of k-lines per shot were varied: 25,50,100,225,450. Corresponding FA=18°,13°,9°,6°,5° for optimal
signal contrast. Different saturation delays=0.75,2.0,5.0,10.0,20.0s. For
breast and abdomen
scans, FOV=350×350×180mm3, resolution=6×6×6mm3,
BW=2854Hz/pixel. Compressed-sensing factor=3 in all scans. Abdomen scans were
acquired during one breath-hold.
Five
other B1 mapping methods for comparison: 1) Single-slice double-angle method (DAM) with 60°/120° excitation
FA with TR=20s; 2) Single-slice Saturated
DAM (SDAM)3,4 with a saturation delay of 0.5s
and TR=0.58s; 3) Multi-slice SDAM with 30 slices; 4) Multi-slice DREAM5-8 with 30 slices, TR/TESTE/
TEFID=4.5/1.7/2.3ms, STEAM/FLASH FA=60°/15°, TR extension=8.5s; 5) 3D AFI9 with TR1/TR2/TE=20/100/4.6ms, FA=60°.
Standard
single-source B1 shimming (“Fixed”) was used for all three organs.
For breast and abdomen scans, a dual-source B1 shimming (“Adaptive”)
was also performed. A vendor-provided B1 shimming method based on geometries of
individual breasts (“SMART”)
was also applied for breast scans.
B1+ maps of the proposed
method were calculated with Eq. (6) with both magnitude and complex subtractions. A mask for each B1 map was manually drawn from
respective raw images. Within the mask area, the B1 maps were
quantitatively compared with the DAM (SDAM for abdomen scans) using four
metrics: 1) root mean square error (RMSE), 2) error mean, 3) error standard
deviation (SD), 4) concordance correlation coefficients (CCC).Results
Figures 2,3,4 shows the results
of an axial
slice of a subject’s brain, breast, and abdomen with the calculated B1+
scale maps from the 2D DAM method, the 3D RF-prepared two-parameter method and
the proposed three-parameter method using both magnitude- and
complex-subtractions, and their respective difference maps, under different
shim settings respectively.
When combining comparison results from
all three organs of all the subjects, including three orientations for each of
the three brains, three shimming conditions for each of three breasts, two
shimming conditions for each of the five abdomens, the RMSE, error mean, error
SD, and CCC values of each B1 mapping method comparing to 2D DAM or
SDAM methods for each organ are plotted in Figure 5.
Compared to the 3D RF-prepared
two-parameter method, the three-parameter method using complex-subtraction
yielded the averaged RMSE 38% lower in the brain (2.4% vs. 3.9%), 45% lower in
the breast (4.2% vs. 7.7%), and 27% lower in the abdomen (8.7% vs. 11.9%); its
CCC values were 5.9% higher in the brain (95.9% vs. 90.6%), 22% higher in the
breast (83.0% vs. 68.1%), and 28% higher in the abdomen (68.2% vs. 53.1%).
Based on these quantitative metrics, the 3D RF-prepared
three-parameter method with complex-subtraction delivered consistently lower RMSE, error
mean, error SD, and higher CCC values than multi-slice DREAM and 3D AFI, and
were close to the results of 2D or multi-slice SDAM (Figure 5).Conclusion
The proposed 3D RF-prepared
three-parameter method with complex-subtraction was demonstrated with high accuracy for
B1 mapping in brain, breast, and abdomen, while only taking a
fraction of time of existing methods.Acknowledgements
No acknowledgement found.References
1. Chung
S, Kim D, Breton E, Axel L. Rapid B1+ Mapping Using a Preconditioning RF Pulse
with TurboFLASH Readout. Magn Reson Med. 2010;64(2):439-446.
2. Breton E, McGorty K, Wiggins GC, Axel
L, Kim D. Image-guided radio-frequency gain calibration for high-field MRI. NMR Biomed. 2010;23(4):368-374.
3. Cunningham CH, Pauly JM, Nayak KS.
Saturated double-angle method for rapid B1+ mapping. Magn Reson Med. 2006;55(6):1326-1333.
4. Sung K, Nayak KS. Measurement and
characterization of RF nonuniformity over the heart at 3T using body coil
transmission. Journal of Magnetic
Resonance Imaging. 2008;27(3):643-648.
5. Nehrke K, Bornert P. DREAM-a novel
approach for robust, ultrafast, multi-slice B1 mapping. Magn Reson Med. 2012;68(5):1517-1526.
6. Nehrke K, Versluis MJ, Webb A, Bornert
P. Volumetric B1+ mapping of the brain at 7T using DREAM. Magn Reson Med. 2014;71(1):246-256.
7. Hooijmans MT, Dzyubachyk O, Nehrke K, et
al. Fast multistation water/fat imaging at 3T using DREAM-based RF shimming. J Magn Reson Imaging. 2015;42(1):217-223.
8. Nehrke K, Sprinkart AM, Bornert P. An
in vivo comparison of the DREAM sequence with current RF shim technology. MAGMA. 2015;28(2):185-194.
9. Yarnykh VL. Actual flip-angle imaging
in the pulsed steady state: a method for rapid three-dimensional mapping of the
transmitted radiofrequency field. Magn
Reson Med. 2007;57(1):192-200.