Emanuele Camerucci1, Benjamin Elder1, Jeffrey Gunter1, John Huston1, Clifford Jack1, and Petrice Cogswell1
1Mayo Clinic, Rochester, MN, United States
Synopsis
We compared the artifact
induced by the Certas Plus at 3.0T vs 1.5T in 9 patients with idiopathic normal
pressure hydrocephalus (iNPH) on six sequences (3D MPRAGE, Sagittal T1, axial DWI,
axial GRE, axial T2 FLAIR, axial T2 FSE) at three levels (atria of the lateral
ventricles, cerebral aqueduct, and cerebellum hemispheres). The 1.5T scans had
a significantly greater area of artifact on DWI (all levels) and GRE
(cerebellar level) sequences, and a smaller area of artifact on T2 FLAIR
(cerebellar level). Post-shunt imaging at 3T with the Certas plus valve is
feasible and with lesser artifact than on 1.5T.
Introduction
Idiopathic normal
pressure hydrocephalus (iNPH) may be treated with ventriculoperitoneal (VP)
shunt placement (1). After VP shunt placement, imaging
with CT - and more recently MRI - is performed to evaluate for features of
overdrainage and improvement in radiographic features of iNPH, as an indirect
indicator of shunt function. Unfortunately, the shunt valve may induce
artifacts that shadow the adjacent brain regions and obscure evaluation. In our
clinical practice, post-shunt imaging is typically triaged to 1.5T scanners
rather than 3T in an effort to reduce shunt-related artifact based on theory
and prior work (2). While the extent of shunt-related
artifact has been studied for several types of shunt valves (3), only limited in vivo studies
exist on the extent of shunt-related artifact from the more recently available
Certas plus valve (4, 5). The goal of this
study was to compare areas and volumes of shunt-induced artifact between 3T and
1.5T images in iNPH patients with a Certas Plus valve as an initial step in
assessing feasibility of 3T post-shunt imaging.Methods
Patients: We
retrospectively identified a cohort of 9 iNPH patients using an in-house built
software for patient diagnostic code detection. Inclusion criteria were a
Certas Plus valve placed via the right parietal approach for treatment of iNPH and
at least one 3T and one 1.5T MRI performed after the surgery.
Imaging: Imaging was
performed on 1.5T (7 GE, 2 SIEMENS) and 3T (5 GE, 4 SIEMENS). MPRAGE, sagittal
T1, axial DWI, axial GRE, axial T2 FSE, and axial T2 FLAIR were performed
according to standard clinical imaging protocols. Acquisition parameters are
summarized in Figure 1.
Analysis: We calculated
area on artifact (defined as the region of signal blowout, as shown in Figure 2)
in a transversal plane by measuring the 2 greatest dimensions at 3 levels:
atria of the lateral ventricle, cerebral aqueduct, and mid cerebellar
hemispheres. We assumed an elliptic shape and used the formula: $$A = πab$$We also calculated the volume of artifact at the aqueductal level assuming an
ellipsoid shape with the following formula: $$V = (4/3)πabc$$ with a and b representing the 2
dimensions of aqueductal area artifact and c representing the
craniocaudal dimension of the artifact, calculated on a sagittal plane. A
Wilcoxon Rank Sum Test was used to compare areas and volumes of artifacts
between 3T and 1.5T scans. The null hypothesis represented no differences among
field strengths. P-values smaller than 0.05 were considered statistically significant.Results
When considering
all levels together, DWI showed a smaller extent of artifact on 3T vs 1.5T (p=
0.0001). Breaking down comparisons by individual sequences and levels, 3T scans
had a significantly lesser degree of artifacts on DWI (Figure 2. Lateral ventricle
p= 0.03; aqueduct p= 0.008; cerebellum p= 0.04) and on GRE (Figure 3. Cerebellum
p= 0.03). On the contrary, 1.5T imaging had less artifacts in T2 FLAIR (Figure 4. Cerebellum p= 0.02). In terms of artifact volumes, we found that 3T images
had smaller volumetric artifacts as compared to 1.5T on DWI (p= 0.002). The
other sequences did not show a statistical difference between 3T and 1.5T. On
MPRAGE, there was a trend toward smaller extent of artifact on 3T.
On sagittal T1 and axial T2, there was a trend toward smaller extent of
artifact at the level of lateral ventricles and cerebral aqueduct at 3T, though
larger artifact at the level of the cerebellum. All results are reported in Figure 5.Discussion
Our results
indicate that the relative extent of artifact on 3T vs 1.5T depends on the
imaging sequence and anatomic location. The imaging sequences most prone to
susceptibility artifact, DWI and GRE, showed on average less artifact at 3T vs
1.5T. Findings correspond with prior work showing smaller extent of artifact on
3T vs 1.5T DWI when a shunt valve was fixed to the temporal scalp of healthy
volunteers (5). Artifact was on
average similar in extent between 3T and 1.5T on the MPRAGE and sagittal T1
sequence, while T2 FLAIR showed greater extent of artifact on 3T. The differential
performance by sequence may be related acquisition technique as well as
specific parameters, such as TE and BW, which will be further evaluated in
future studies. Regarding the anatomic location, 3T on average performed better
on more cranial sections particularly on MPRAGE, DWI, and GRE, but had more
artifacts at the lowest imaging plane evaluated, particularly on T2 FLAIR. The
difference in artifact extent at more cranial vs caudal levels on 3T vs 1.5T
may reflect differences in shape of the artifact.Conclusion
Our results
indicate that post-shunt imaging in iNPH patients with the Certas Plus valve
can be performed at 3T with the extent of shunt artifact being less than or
similar to that on a 1.5T scanner on most sequences. The decreased artifact
extent at 3T is likely related to difference in scan parameters and implies
that artifact could be further reduced with future optimization of sequence
parameters. Implications for clinical practice include wider availability of
scanners on which post-shunt imaging may be performed and feasibility to
perform potentially higher quality 3T exams without increased extent of
shunt-related artifact compared to 1.5T.Acknowledgements
No acknowledgement found.References
1. Toma
AK, Papadopoulos MC, Stapleton S, Kitchen ND, Watkins LD. Systematic review of
the outcome of shunt surgery in idiopathic normal-pressure hydrocephalus. Acta
Neurochir (Wien). 2013;155(10):1977-80.
2. Matsuura H, Inoue T, Ogasawara K,
Sasaki M, Konno H, Kuzu Y, et al. Quantitative analysis of magnetic resonance
imaging susceptibility artifacts caused by neurosurgical biomaterials:
comparison of 0.5, 1.5, and 3.0 Tesla magnetic fields. Neurol Med Chir (Tokyo).
2005;45(8):395-8; discussion 8-9.
3. Toma AK, Tarnaris A, Grieve JP,
Watkins LD, Kitchen ND. Adjustable shunt valve-induced magnetic resonance imaging
artifact: a comparative study. J Neurosurg. 2010;113(1):74-8.
4. Uchida D, Amano Y, Nakatogawa H,
Masui T, Ando N, Nakayama T, et al. Setting pressure can change the size and
shape of MRI artifacts caused by adjustable shunt valves: a study of the 4 newest
models. J Neurosurg. 2018:1-8.
5. Amano
Y, Kuroda N, Uchida D, Sakakura Y, Nakatogawa H, Ando N, et al. Unexpectedly
Smaller Artifacts of 3.0-T Magnetic Resonance Imaging than 1.5 T:
Recommendation of 3.0-T Scanners for Patients with Magnet-Resistant Adjustable
Ventriculoperitoneal Shunt Devices. World Neurosurg. 2019;130:e393-e9.