Jihun Kwon1, Takashige Yoshida2, Masami Yoneyama1, Johannes M Peeters3, and Marc Van Cauteren3
1Philips Japan, Tokyo, Japan, 2Tokyo Metropolitan Police Hospital, Tokyo, Japan, 3Philips Healthcare, Best, Netherlands
Synopsis
Time-of-flight MR angiography (TOF-MRA) is
a non-contrast-enhanced imaging technique widely used to visualize intracranial
vasculature. In this study, we investigated the use of ultra-thin slice (up to
0.4 mm) to improve the delineation of the cerebral arteries in TOF-MRA. To reduce
the noise while preserving the image quality, Compressed SENSE AI (CS-AI)
reconstruction was used. Our results showed that the improved noise reduction
by CS-AI enabled better visualization of vessels, especially on the thinner
slices compared to conventional Compressed-SENSE. The usefulness of CS-AI was
also demonstrated in clinical cases with moyamoya disease and suspected
aneurysm patients.
Introduction
Time-of-flight MR angiography (TOF-MRA) is
a non-contrast-enhanced imaging technique widely used to visualize intracranial
vasculature1. The cerebral arteries are highly tortuous and sometimes difficult
to visualize clearly by TOF-MRA. Improving the delineation of arteries in the
middle cerebral artery (MCA) is of particular interest and acquiring TOF-MRA
with very thin slices would improve the visualization of tortuous vessels. However,
thinner slices also suffer from more noise in the image, which makes ultra-thin
slice imaging challenging.
In this study, Compressed SENSE AI (CS-AI)
reconstruction2,3 was used to reduce the noise while preserving the image quality
under ultra-thin slice TOF-MRA. It is hypothesized that the slice thickness can
be significantly reduced while maintaining the image quality by using the CS-AI
reconstruction algorithm, thus leading to the better visualization of cerebral
vascular patterns. The purpose of this study was to acquire ultra-thin slice 3D
TOF-MRA using the CS-AI reconstruction and compare the image quality with conventional
Compressed-SENSE (C-SENSE).Methods
Three healthy volunteers and two patients;
one with moyamoya disease and another with suspected aneurysm were examined on
3.0T whole-body clinical systems (Ingenia Elition X, Philips Healthcare) using
a 32ch head coil, following informed consent with an IRB-approved protocol.
Volunteer study: TOF-MRA was acquired by
C-SENSE and CS-AI at slice thicknesses 1.0, 0.8, 0.6, and 0.4 mm, and then
reconstructed to 0.5, 0.4, 0.3, and 0.2 mm, respectively. Total scan times were 3:50,
4:46, 6:22, and 9:41 min, respectively. In-plane resolution was 0.55x0.75 mm2
and reconstructed to 0.32x0.32 mm2. 90 slices.
Patient study: TOF-MRA was acquired with
C-SENSE and CS-AI at voxel size 0.4x0.8x0.8 mm3 and reconstructed to
0.2x0.2x0.4 mm3. 150 slices, total scan time = 4:09 min. Following
parameters were common to all volunteer and patient examinations: FOV=200x200
mm, TR/TE=25/3.5 ms, flip angle=18, acceleration factor=5.4, number of signals
averaged (NSA)=1.
Quantitative evaluation: image quality was
evaluated by measuring the number of vessels in TOF-MRA. On the volunteer
image, a straight line was drawn in the left MCA area on a 90-mm slab maximum
intensity projection image (MIP). The signal profile along that line was then
obtained for the C-SENSE and CS-AI. Region of interest (ROI) was placed on the
background stationary tissues in the brain parenchyma. A signal was defined as
coming from a vessel if the signal was greater than STave+5STSD, where STave and STSD are the mean signal intensity and standard
deviation in the stationary tissue ROI, respectively.
The CS-AI model used in this study is the
extension of the previously introduced AI-based reconstruction algorithm,
Adaptive-CS-Net2,3. In CS-AI, the iterative optimization procedure in the C-SENSE
reconstruction chain is unrolled for a fixed number of reconstruction blocks.
Each block consists mainly of Unet-like architecture, which performs as a
denoiser. The model was trained on more than 700,000 images, including 2D and
3D data, and multiple contrasts and anatomical areas.Results and Discussions:
Figure 1 shows the comparison of TOF-MRA
between C-SENSE and CS-AI for slice thickness 1.0, 0.8, 0.6, and 0.4 mm.
Overall, in C-SENSE, the noise level increased and image quality degraded as
the slice thickness reduced. In CS-AI, the noise level did not increase
significantly even at 0.4 mm slice thickness.
Figure 2 shows the MIP of the TOF-MRA
images shown in Fig.1. Because thinner slices can better represent tortuosity
of vessels, partial signal loss in the internal carotid artery (red arrows)
seen in slices thicker than 0.6 mm was resolved in 0.4 mm. Background noise in
CS-AI was visually well suppressed compared to C-SENSE. Figure 3 (b) and (c)
show the signal profile for 1.0 and 0.4 mm slices, respectively, along the line
drawn in figure 3(a). For both 1.0 and 0.4 mm thicknesses, the baseline signal
intensity in CS-AI was lower than that of C-SENSE, while the signal intensity
of each peak was higher in CS-AI than C-SENSE. This suggests improved
signal-to-noise in CS-AI compared to C-SENSE. The
number of vessels detected was 6 and 8 on the 1.0 mm slice, and was 6 and 9 on
the 0.4 mm slice for C-SENSE and CS-AI, respectively. This trend was also seen in the other two
subjects. This suggests that the improved noise reduction by CS-AI can
visualize more vessels compared to C-SENSE, especially when the slice is
thinner.
Figure 4 shows the TOF-MRA of the patient
with moyamoya disease. CS-AI showed better noise reduction compared to C-SENSE.
On the sagittal image, horizontal banding-like artifact was present in C-SENSE
(red arrows) but was diminished in CS-AI.
Figure 5 shows the TOF-MRA of the patient
with a suspected aneurysm. On C-SENSE it was difficult for clinicians to
determine if there was an aneurysm or not (red arrow). But on CS-AI the better
delineation of the vessel helped make a diagnosis that it is likely not an
aneurysm, but that the vessel is bending towards a confusing direction.Conclusion
We acquired ultra-thin slice 3D TOF-MRA for
the brain using the CS-AI and compared the image quality with the C-SENSE. We
showed that the improved noise reduction by CS-AI enabled better visualization
of vessels, especially on the thinner slice thicknesses. The usefulness of
CS-AI was also demonstrated in clinical cases.Acknowledgements
No acknowledgement found.References
1. Dumoulin CL, Cline HE, Souza SP,
Wagle WA, Walker MF. Three‐dimensional time‐of‐flight magnetic resonance
angiography using spin saturation. Magn Reson Med. 1989;11(1):35-46. doi:10.1002/mrm.1910110104
2. Pezzotti N, de Weerdt E, Yousefi S, et
al. Adaptive-CS-Net: FastMRI with Adaptive Intelligence. arxiv.
2019;(NeurIPS). http://arxiv.org/abs/1912.12259
3. Pezzotti N, Yousefi S, Elmahdy MS, et
al. An Adaptive Intelligence Algorithm for Undersampled Knee MRI
Reconstruction. IEEE Access. 2020;8:204825-204838.
doi:10.1109/ACCESS.2020.3034287