Qiaoling Wu1 and Geli Hu2
1Department of Radiology, Peking Union Medical College Hospital,Chinese Academy of Medical Sciences, Beijing, China, 2Philips Healthcare, Beijing, China
Synopsis
Banding artifacts are frequently observed in brachial plexus imaging, particularly within the spinal canal, with the balanced FFE
sequence (b-FFE) at 3.0T due to increased influence of higher magnetic field
inhomogeneity compared with clinical 1.5T. In this work we propose using a RF
phase cycling technique with b-FFE (namely b-FFE-XD) to reduce the banding
artifacts in clinical. This design RF phase limits the intravoxel dephasing of
scan object voxels, therefore the artifacts caused by field inhomogeneity
decrease. Meanwhile, the compressed SENSE technique was
adopted to reduce the scan time. Results demonstrated successful removal of
banding artifacts with the proposed technique.
Synopsis
Banding artifacts are frequently observed
in brachial plexus (BP) imaging, particularly within the spinal canal, with the balanced
FFE sequence (b-FFE) at 3.0T due to increased influence of higher magnetic field inhomogeneity compared with clinical 1.5T. In this work
we propose using a RF phase cycling technique with b-FFE (namely b-FFE-XD) to reduce the banding
artifacts in clinical. This design RF phase limits the intravoxel dephasing of scan object voxels, therefore the artifacts caused by field inhomogeneity decrease than that of b-FFE. Meanwhile, the compressed SENSE technique was adopted to reduce the
scan time. Results demonstrated
successful removal of banding artifacts with the proposed technique.Summary of main findings
The combination of b-FFE-XD with
compressed SENSE techniques can effectively reduce banding artifacts in
BP imaging at clinical 3T MR scanners without extending scan time.Introduction
It remains clinically challenging to
diagnose disfunctions in brachial plexus (BP), particularly within spinal canal,
due to its complex anatomy, which provides sensory and motor innervation to the
ipsilateral shoulder, chest, and hand [1-2]. Despite the use of balanced fast
field echo (b-FFE) technique, this sequence is prone to magnetic field
inhomogeneity and is frequently disturbed by banding artifacts with increased
severity at 3.0T. This study proposed a RF phase-cycling technique as an
addition to the conventional b-FFE sequence (b-FFE-XD hereafter) to suppress the potential
banding artifacts in brachial plexus imaging. In addition, compressed SENSE was
implemented in our study to reduce imaging time. Method
Ten healthy volunteers (4 women, and 6 men,
mean age, 53.30±8.73
years; age range, 31-64 years) underwent B-FFE-XD and B-FFE scans (Coronal
orientation) at 3.0T (Ingenia Elition, Philips healthcare, Best, the
Netherlands) on the brachial plexus with a body coil for RF transmitting and an
anterior coil for signal receiving. The detailed acquisition parameters for
both sequences were kept the same, listed in Table1 as below, except that a RF
phase cycling scheme ([0,180+1*X, 2*X, 180+3*X,4*X,…], where X=n*360/NSA, and n
is the nth average) was applied to the excitation pulse for b-FFE-XD sequence.
Images were evaluated independently by
two radiologists (>3 years radiology experience) in terms of overall image
quality, banding artifact suppression, and diagnostic certainty using a 5-point
scale criteria (5 for the best result, the scoring system was listed in Table2).
Reduction factor of compressed SENSE was set as 1.5 to accelerate the scan
speed.
Consistency of the evaluation results
between the two observers were assessed using intra-class correlation
coefficients (ICC) in SPSS (IBM). If the consistency was good, the scores would be averaged, then performed the subsequent analysis. The
scores of the two sequences were compared using a Wilcoxon signed-rank test
with a significance threshold of P < 0.05.Results
Effective banding artifact suppression without prolonged scan time was
achieved by B-FFE-XD combined with compressed SENSE , representative images
shown in Figure 1. Good agreement was achieved between the scoring of the two
observers, and the ICCs of banding reduction and overall imaging quality scores
were 0.786 ad 0738 (both P<0.001), respectively. The scores on ‘black band artifact’ and on
‘overall image quality’ were both significantly higher of b-FFE-XD than that of b-FFE (4.75±0.55 v.s. 3.10±0.64,
4.30±0.66 v.s. 4.10±0.79,
respectively, both P<0.05), as shown in Table3.
Discussion and Conclusion
Our results demonstrated the promising clinical application of the proposed RF phase cycling scheme for brachial
plexus imaging with compressed SENSE, which yielded improved image quality and reduced banding
artifacts compared to the conventional b-FFE technique without extension of scan time.Acknowledgements
NoReferences
[1]
Arvind Vijayasarathi, algun H. Chokshi. MRI of the brachial plexus: A practical
review. Applied Radiology. 2016; 9-18.
[2] Alberto
Tagliafico, Giulia
Succio, etc. MR imaging of the
brachial plexus: comparison between 1.5-T and 3-T MR imaging: preliminary
experience. Skeletal Radiology. 2011;40:717-724.