Wei Liu1, Michael Koehler2, Flavio Carinci2, Adam Kettinger2, Thorsten Feiweier2, Kun Zhou1, and Mario Zeller2
1Siemens Shenzhen Magnetic Resonance Ltd, Shenzhzen, China, 2Siemens Healthcare GmbH, Erlangen, Germany
Synopsis
In this work, we propose to compensate the residual eddy current field caused by diffusion gradients at the time when chemically selective fat suppression gets applied with an additional gradient applied after the EPI readout. Using an analytic solution, the amplitude of the eddy currents can be cancelled under certain assumptions. The experimental results based on a volunteer scan demonstrate improved fat suppression in diffusion MRI with the proposed method.
INTRODUCTION
Diffusion weighted imaging (DWI) is a
valuable complement to traditional techniques and improves the sensitivity for
both detection and characterization of diseases. However, it can be affected by
eddy current artifacts due to the rapid switching of diffusion gradients, which
introduces eddy current in the nearby conductors, inducing local magnetic
fields that interfere with the spatial encoding gradients1. The
typical eddy current artifacts in DWI include shearing, stretching and
rotation, which will lead to misalignment among different volumes acquired by
different diffusion gradients. Such effect varies across different diffusion
gradient directions and will increase with higher b-values, which usually
require diffusion gradients with strong amplitude and long duration. Different
techniques have been proposed to improve such eddy current induced distortion2-5.
In addition to the common eddy current induced distortion artifacts, the
long-term eddy current components can generate a residual field which persists after
the image acquisition. This may impact the succeeding chemically selective RF-pulses
and lead to insufficient fat suppression.
In this study, we propose
to compensate such residual eddy current field by introducing an additional
gradient, in which the amplitude of the eddy currents can be cancelled by
considering an analytical solution under certain assumptions. The approach was validated
in a volunteer scan. The results show that the proposed method can
improve fat suppression in DWI.METHODS
Eddy current time dependence can be
broadly described by exponential decays with long- and short-time constants. The
time constant indicates the rate of exponential build up and decay of field
perturbations after a gradient change6. To simplify the analytical
calculation, we consider the eddy currents to be mono-exponential with only a
single long-time constant. The field generated by the eddy currents after an
ideal, rectangular shaped gradient pulse G (amplitude G0,
duration TG) can be described by a temporal decay function $$$g(t)$$$:
$$g_{0}\approx G_{0}(exp(\frac{-T_{G}}{\tau})-1)$$
$$g(t) = g_{0} exp(\frac{-t}{\tau}) \tag1$$
where τ is the long-time constant, and t = 0 occurs immediately after the end of G with the eddy current field $$$g_{0}$$$.
The
diffusion pulse sequence with eddy current compensation is shown in Fig. 1. Two
monopolar gradients DG1 and DG2 with equal area are used for
diffusion encoding, and one chemically selective fat saturation RF-pulse
is applied before each excitation. To eliminate the diffusion gradients induced
eddy current effect on the succeeding fat saturation pulse, an additional gradient
will be applied after the EPI readout. The required compensation parameters
can be given by:
$$G_{d}(exp(\frac{-T_{d}}{\tau})-1)exp(\frac{-TD_{DG1}}{\tau}) +G_{d}(exp(\frac{-T_{d}}{\tau})-1)exp(\frac{-TD_{DG2}}{\tau})=G_{ec}(exp(\frac{-T_{ec}}{\tau})-1)exp(\frac{-TD_{EC}}{\tau}) \tag2$$
where Gd is the amplitude of the paired diffusion gradients DG1 and
DG2; Td is the duration of DG1 and DG2; Gec is the amplitude of the eddy current
compensated gradient (ECCG) applied after the EPI readout; Tec is the duration of ECCG; τ is the eddy current long-time constant; TDDG1, TDDG2 are the durations
from the end of DG1 and DG2 to the beginning of the chemically selective fat saturation pulse, respectively; TDEC is the duration from
the end of ECCG to the beginning of the chemically selective fat saturation pulse.
To demonstrate improved fat suppression with the proposed
method, we calculated the ECCG parameters based on the applied diffusion gradients
to generate a prototypic single shot diffusion sequence and assessed the fat
suppression performance with a volunteer scan.EXPERIMENTS AND RESULTS
All measurements were performed on a commercial 1.5T
scanner equipped with a 6-channel body array and an 18-channel spine array. One
volunteer was scanned with informed consent. Conventional and prototypic single
shot diffusion sequences were acquired with the same slice thickness and
coverage using the following parameters: FOV = 400x275 mm2, matrix
size = 134x92, in-plane GRAPPA factor = 2, b = 50 s/mm2 with 1
average, b = 800 s/mm2 with 4 averages, diffusion mode = 4-Scan Trace,
30 slices with 6 mm thickness, 20% slice gap, FatSat mode = SPAIR; TE/TR = 60ms/6300ms with total scan time of 2:23 min. Fig. 2 compares exemplary DW images of the liver acquired
using conventional and prototypic diffusion sequences. Inhomogeneous fat suppression was observed in the high-b value image acquired
with the conventional sequence. After appropriate eddy current compensation,
Fig. 2D shows improved fat suppression in the edge of images, where the eddy
current effects are usually more severe.DISCUSSION
We have demonstrated that it is possible
to mitigate eddy current related imperfections of fat suppression by adding
an extra eddy current compensation gradient into a diffusion sequence. To
achieve a more accurate compensation, the eddy current measurement experiment
should be done in advance to estimate the long-time constant correctly.
Please note that the residual fat signal
depends on the diffusion gradient strength, duration between the diffusion
gradients and chemically selective fat saturation pulse, and on the eddy current
decay. Thus, it exhibits different appearance in low-b and high-b value images,
which will also lead to incorrect ADC calculation. The proposed method can also
mitigate such incorrection by improving fat suppression.CONCLUSION
The proposed method allows for a practical, accurate compensation
of the residual eddy current cancellation after the EPI readout and thus enjoys a more accurate spectral selection of
fat signal, which
is more attractive in diffusion applications at low field.Acknowledgements
No acknowledgement found.References
1. Jezzard,
P., Barnett, A. S., and Pierpaoli, C. Characterization of and
correction for eddy current artifacts in echo planar diffusion
imaging. Magn. Reson. Med. 1998;39:801-812.
2. Reese
TG, Heid O, Weisskoff RM, Wedeen VJ. Reduction of eddy-current-induced
distortion in diffusion MRI using a twice-refocused spin echo. Magn. Reson. Med. 2003;49:177-182.
3. Bodammer
N, Kaufmann J, Kanowski M, Tempelmann C. Eddy current correction in
diffusion-weighted imaging using pairs of images acquired with opposite diffusion
gradient polarity. Magn. Reson. Med. 2004;51:188-193.
4. Andersson,
J. L. R., and Sotiropoulos, S. N. An integrated approach to correction
for off-resonance effects and subject movement in diffusion MR
imaging. Neuroimage. 2016;125:1063-1078.
5. Finsterbusch,
J. Eddy-Current Compensated Diffusion Weighting with a Single Refocusing RF
Pulse. Magn. Reson. Med. 2009;61:748-754.
6.
Bernstein, M.A., King, K.F and Zhou, X.J. Handbook of MRI Pulse
Sequences. Elsevier, Amsterdam; 2004.