Qike Song1 and Yu Shi1
1radiology, Shengjing Hospital of China Medical University, Shenyang,, China
Synopsis
To determine the feasibility of multi-frequency MRE for assessing
pancreatic stiffness in healthy volunteers and patients with pancreatic
adenocarcinoma (PDAC).A total of 40 healthy volunteers and 10 patients with
PDAC were prospectively recruited between March 2020 and October 2021.Each volunteer and patient underwent 3.0-T
pancreatic MRE at frequencies of 30Hz, 40Hz, 60Hz, 80Hz and 100Hz.Shear stiffness of healthy pancreas and PDAC,
pancreatic width and volume, waist circumference, and wave distance were
measured. Image quality assessment was performed according to image quality
score (IQS: 1~4,≥3 considered qualified).
Synopsis
To determine
the feasibility of multi-frequency MRE for assessing pancreatic stiffness in
healthy volunteers and patients with pancreatic adenocarcinoma (PDAC). We
propose a multi-frequency pancreatic MRE scanning scheme to explore the optimal
MRE imaging frequency for healthy volunteers PDAC patients. Our research shows that
MRE at 30Hz and 40 Hz provides qualified wave images and reliable measurements
of pancreatic stiffness, and MRE at 30Hz, 40 Hz and 60 Hz acceptable for PDACs.Methods
A total of 20 healthy volunteers and 5 patients with PDAC and main
pancreatic duct obstruction (MPD ≥3 mm) were prospectively recruited between
March 2020 and April 2021.
All examinations were performed on an MR scanner (Signa HDX 3.0T system;
GE Healthcare, Milwaukee, WI, USA) equipped with an eight-channel phased-array
body coil. All subjects were instructed to fast for 6–8 hours before the
examination. A rectangular semi-rigid driver (13× 13 cm) was placed against the
upper abdomen, centred on the xiphisternum, and secured with a 20-cm-wide
elastic band that wrapped around the body to ensure good coupling between the
driver and the body. All subjects were
placed in supine position. An active pneumatic driver system situated outside
the scan room generated mechanical vibrations at the frequency of 30 Hz, 40 Hz,
60Hz, 80Hz and 100Hz for delivery to the upper abdomen via a plastic tube.
Thirty-two consecutive slices were acquired within 5 breath holds (4 of 24 s
and 1 of 12 s) at 30 Hz; 5 breath holds (4 of 22 s and 1 of 11 s) at 40 Hz; 3
breath holds (3 of 21 s) at 60 Hz; 3 breath holds (3 of 22 s) at 80 Hz and 3
breath holds (3 of 23 s) at 100 Hz. The imaging parameters for 3D SE-EPI MRE
were as follows: The repetition time at each frequency is the same, repetition
time = 1375ms; at each frequency the echo times for any particular patient were
the same, and ranged from about 37.7–40.5 ms; depending on the patient weight;
field of view = 36–44 cm; matrix size = 96 × 96; slice thickness = 3.5 mm;
phase offsets = 3; parallel imaging acceleration factor = 3. In addition to MRE
sequences, we also obtained routine plain pancreatic MRI studies, including
respiratory-triggered T2-weighted imaging with/without fat suppression and
T1-weightedimaging (T1WI).
The post-processing software was supplied by the Mayo
Clinic, Rochester, MN, USA, compatible with the MRE pulse sequence. A direct
inversion of the Helmholtz wave equation was performed to estimate the tissue
stiffness(1). X, Y and Z curled wave images, magnitude images, and
local frequency estimation (LFE) confidence maps were also automatically
generated. Regions of interest (ROIs) were drawn on the level that
can show the largest cross-section of 5 subregions, including 3 subregions (head
[uncinated], body and tail) for healthy pancreas, and 2 subregions (PDAC and
its pancreatic distal remnant [1-2cm away from PDAC]) for patients at the
magnitude images, avoiding the pancreatic border, large vessels, and the
surrounding tissues(2). We refer to the 4-point wave
image quality score (IQS) method(3) and combine the confidence map to comprehensively
evaluate the image quality. The ratings for image quality and
measurements of pancreatic stiffness were assessed separately for head, body,
and tail in volunteers and for PDAC and pancreatic remnant for patients).The wave
distance was described as minimum perpendicular distance between anterior
abdominal wall and each ROI, which was also obtained from axial T1-weighted
images. Pancreatic volume estimates were calculated as the product of voxel
count and voxel size.Results
In logistic
regression analysis, qualified IQS in healthy volunteers was independently
associated with smaller BMI (Odds Ratio [OR] = 0.89, P = 0.003), shorter wave
distance (OR = 0.70, P = 0.028), lower frequency (30 Hz, OR = 71.13, P
<0.001; 40 Hz, OR = 90.85, P <0.001, respectively). Qualified IQS
in PDAC was independently associated with frequency (30Hz, OR = 21.0, P
=0.016; 40Hz, OR = 21.0, P =0.016, 60Hz, OR = 9.33, P =0.033,
respectively), when using 100Hz as reference. The mean shear stiffness in healthy
pancreas at 30Hz and 40 Hz was 0.74±0.07kPa and 1.18±0.10kPa, respectively. The
mean shear stiffness of PDAC at 30Hz, 40Hz and 60Hz was 1.41±0.14kPa,
3.50±0.34kPa and 4.76±0.52kPa, respectively.Discussion
Theoretically, the data at 60-100Hz should give rise to
improved resolution because the shorter shear wave length at the higher
frequency can improve the calculation of stiffness. However, MRE at 60-100Hz appeared to
deliver waves superficially, showing narrow penetration and less parallel
waves, with illumination only immediately under the skin or confined to the
pancreatic body.
In PDAC, MRE at 60Hz also produced diagnostic quality images. This mainly due to PDAC lesions had a
much higher mean stiffness value than that of the normal pancreas. The higher mean stiffness value indicates longer
wavelengths and less energy loss from the propagated waves.Conclusion
MRE at 30Hz
and 40 Hz provides qualified wave images and reliable measurements of
pancreatic stiffness, and MRE at 30Hz, 40Hz and 60Hz acceptable for PDACs.Acknowledgements
No acknowledgement found.References
1. Manduca, A., D.S.
Lake, S.A. Kruse, et al. Spatio-temporal directional filtering for improved
inversion of MR elastography images.Med Image Anal, 2003. 7(4):465-73.
2. Venkatesh, S.K., M.
Yin,R.L. Ehman. Magnetic resonance elastography of liver: technique, analysis,
and clinical applications.J Magn Reson Imaging, 2013. 37(3):544-55.
3. Ji, R., J. Li, Z.
Yin, et al. Pancreatic stiffness response to an oral glucose load in obese
adults measured by magnetic resonance elastography.Magn Reson Imaging, 2018.
51:113-119.