Masahiro Takizawa1, Takashi Nishihara1, and Chikako Moriwake1
1Radiation Diagnostic System Division, FUJIFILM Healthcare Corporation, Kashiwa-shi, Japan
Synopsis
Cylinder-shaped
pre-saturation pulse train is developed for non-contrast enhanced spatially-selective
and time-resolved vessel imaging. The target vessel is selected by
cylinder-shaped pre-saturation, and the dynamics of blood flow in the target
vessel is observed by changing the number of applied pre-saturation pulses. The
developed pulse train was demonstrated to visualize dynamics of a target pulmonary
vessel in the lung.
Introduction
Pulmonary
artery coil embolization (PACE) is often used to treat pulmonary arteriovenous
malformations (PAVM). Before PACE, CT or contract enhanced (CE) time-resolved
MRA is used to specify the target vessel. To label the blood flow, a few non-CE
MRA method using IR or saturation pulse are developed. However, it is difficult
to visualize the target vessel using conventional labelling pulses since the
blood vessels have complex structure. We developed a cylinder-shaped
pre-saturation (Cyl.-PSat) pulse for non-CE selective MRA[1]. In this study, a
new pulse train based on the Cyl.-PSat pulses is developed to observe flow
dynamics in the target vessel and tuned for lung imaging.Methods
<Pulse sequence diagram>
Figure 1 shows the sequence diagram of the Cyl.-PSat pulse and the
new pulse train. The Cyl.-PSat pulse is designed to work on 3D FSE sequence. To
suppress residual vessel signals in the target area, a conventional saturation
(Conv.-Sat) pulse, interval time between the Conv.-Sat pulse and main scan were
introduced to the 3D FSE sequence. To observe flow dynamics in the target
vessel, Cyl.-PSat pulse train was inserted between the Conv.-Sat pulse and the main
scan, and the number of cylinders can be increased up to 48.
<Phantom experiment>
Experiment was conducted on a 3T whole body MRI system. A phantom
with small constant flow was scanned to observe relationship between the flow
distance and the Cyl.-PSat count. Before the experiment, actual flow velocity
[cm/sec] was calculated by dividing flow rate (calculated from the time needs 1
liter of water to flow through the phantom) by the phantom area. The resultant actual
flow velocity was 53cm/sec. Anticipated flow length was calculated by dividing
the actual flow velocity by Cyl.-PSat period; obtained by multiplying the Cyl.-PSat
count and the Cyl.-PSat duration. Drawing length of water on the image was
measured and compared to the anticipated flow length.
<Volunteer study>
The
study was approved by the ethics
committee of FUJIFILM Healthcare Corporation. The
Cyl.-PSat was set for the pulmonary artery for the volunteer as shown in fig.2.
To visualize only inflow blood signal, the Conv.-Sat pulse was set on entire
region of right lung. To set blood signal null in the target area at the point
of excitation for main scan, FA of the Conv.-Sat pulse was set to 180 degrees
and the interval time between the Conv.-Sat pulse and main scan was set to 800msec.
And to achieve continuous suppression of inflow blood signal, the duration of Cyl.-PSat
pulse was minimized (20msec). The sequence parameters were as follows; FOV 350mm,
TR/TE 4000/41.4msec, thickness 3mm, matrix 224×176, and respiratory gating. The Cyl.-PSat count
were set to 0 (no Cyl.-PSat), 10, 20, 30, and 40. Figure 3 shows the scheme of the
post processing. The inflow blood signal was suppressed by Cyl.-PSat and the saturated
distance of the target vessel becomes long according to increasing the
Cyl.-PSat count. The blood signal of the target vessel is extracted by
subtracted images acquired with and without Cyl.-PSat.Results
<Phantom experiment>
Figure 4 shows the result of phantom experiment. By increasing the Cyl.-PSat
count, the water saturation length was increased, and the water gradually
disappeared in the image. In the Cyl.-PSat count 10 case, the measured drawing length
on the image was 11.6cm. The anticipated flow length was 10.6cm, and they were almost
same and the difference is less than 10%.
<Volunteer study>
Blood
flow signals in the target vessel were visualized using Cyl.-PSat pulses (Fig. 5).
The length of the target vessel becomes longer as the Cyl.-PSat count becomes
larger.Discussion
The phantom study and a volunteer study clearly demonstrated that
the proposed pulse train can visualize the flow dynamics of the spatially
selected vessel. Furthermore, quantitative evaluation of the ratio between
drawing length and the pulse count may indicate the velocity in the target
vessel.
In this study, we used respiratory gating to suppress motion
artifact caused by breathing. The subtraction processing between w/ and w/o
Cyl.-PSat was needed to visualize target vessel, thus motion artifact cases
mis-registration of target vessel. Non-subtract vessel visualization will be
realized by changing Cyl.-PSat pulse train position from the main scan side to the
Conv.-Sat side.
As for
the vessel visualization, even the Cyl.-PSat count 40 case (continuous suppression
during 800 msec) does not show vein. In the actual PAVM case, the blood flow in
the target vein might be observed in the earlier phase than that in the normal
vein. So it is expected that interval time less than 1000 msec might be useful.Conclusion
Cylinder
shaped pre-saturation pulse train was developed to investigate flow dynamics of
only a target blood vessel. It enabled changing the drawing length of the
target vessel by changing the Cyl.-PSat count, which reflects the flow dynamics
in the target vessel.Acknowledgements
No acknowledgement found.References
[1] T. Nishihara, et. al. Selective MRA for Portal
Venography Using Beam Saturation Pulse, 22nd ISMRM (2014), p. 1501.