Dongyue Si1, Rui Guo2, Bowei Liu1, Daniel A. Herzka3, and Haiyan Ding1
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, Tsinghua University, Beijing, China, 2Department of Medicine, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA, United States, 3National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD, United States
Synopsis
Myocardial
T1 and T2 mapping enable quantitative detection of
various cardiac diseases. Here we propose a fast 3D free-breathing simultaneous
T1 and T2 mapping sequence, which acquires four volumes for
joint estimation of T1 and T2. A very small flip angle is
used for efficient sampling of the equilibrium longitudinal magnetization. Whole-heart
T1 and T2 maps were acquired with high resolution of
1.5×1.5×4mm3 within 5 min in normal human subjects with image
quality comparable to conventional 2D methods.
INTRODUCTION
Cardiovascular
magnetic resonance parametric mapping can provide quantitative tissue
characterization in various cardiac disease such as fibrosis, myocarditis,
iron, and amyloid (1)
among others. Many approaches are commercially available for both T1
(e.g. Modified look-locker inversion recovery (MOLLI) (2),
Saturation recovery single-shot acquisition (SASHA) (3))
and T2 (e.g. Gradient Spin Echo (GraSE) (4) and
T2-Prep based (5))
and are now widely used in clinical practice. However, 2D sequences are limited
in the spatial coverage. Separate mapping acquisitions increase the total scan
time and can result in spatial mismatch which hampers voxel-based analysis. Simultaneous
whole-heart T1 and T2 mapping sequences have been
proposed (6–8).
While these 3D sequences have higher resolution and larger coverage, long scan
times remain a major limitation (6,7).
We propose a 3D free-breathing simultaneous whole heart T1 and T2
mapping sequence, SAturation recovery and Variable flip Angle (SAVA), which
enables highly efficient and simultaneous acquisition of inherently co-registered
T1 and T2 maps. METHODS
Sequence Design:
The proposed free-breathing
sequence, referred to as SAVA T1-T2, sequentially
acquires four 3D volumes with hybrid T1 and T2 weighting
(Figure 1). The
first volume (IMG1) is sampled without preparation pulses to measure
the equilibrium longitudinal magnetization. A very small flip angle (FA) of 2°
is used to minimize the impact of imaging on the magnetization and shorten the
recovery time before the subsequent readout (9). The following three
volumes (IMGi, i=2,3,4) are prepared by both saturation (SAT) and T2
preparation (T2 PREP). The delay times after SAT (TSAT i)
and echo times (TEi) of T2 PREP are variable. TSAT i
linearly distributed from 0.5×TMAX to TMAX, where TMAX
was the maximal available TSAT in one cardiac cycle. FA = 15° is
used for readout to obtain a high signal to noise ratio.
The signal
intensity of the ith volume (Si) is formulated in the
following equation:
$$$S_i=S_0 (1-e^{\frac{-T_{SATi}}{T_1} } ) e^{\frac{-TE_i}{T_2}} \sin(FA_i)$$$ [1]
where S0
is the steady-state signal.
Joint
T1 and T2 calculation:
T1 and
T2 are jointly calculated voxel by voxel from the four volumes using
a dictionary matching method combined
with an orthogonal matching pursuit algorithm (10).
The dictionary is generated using Equation
1, with T1 ranging from 60ms to 2500ms in 5ms steps while T2 ranges from 4ms to 100ms in
1ms steps, from 102ms to 200ms in 2ms steps, and from 210ms to 450ms in 10ms
steps (11).
Experiments:
The study was approved
by the local institutional review board. Four normal human subjects (2 males,
23 ± 4 years) were recruited and imaged with SAVA T1-T2,
MOLLI and GraSE on a 3.0T MR scanner (Ingenia CX, Philips Healthcare, Best,
Netherlands). Written informed consent was obtained from all subjects. SAVA T1-T2
imaging parameters were: FOV 300×300×92mm3, voxel size1.5×1.5×8mm3
reconstructed to 1.5×1.5×4 mm3, TR/TE 4.3ms/1.35ms, 35 readouts per
shot, TSATi/TEi 0.5×TMAX/0ms, 0.75×TMAX/25ms,
TMAX/45ms for i=2,3,4, and CS-SENSE factor 2. MOLLI and GraSE
acquired three slices in three breath-holds. Imaging parameters were: FOV 300×300mm2,
voxel size 2×2×10mm3. MOLLI was performed using 5s(3s)3s scheme, TR/TE/flip
angle 2.1ms/0.97ms/35°, SENSE factor 2. GraSE used nine echo times ranging from
8.8ms to 79.6ms, TR 1 heartbeat, EPI factor 7.
The left ventricle
(LV) was manually segmented on T1 and T2 maps and divided
into the 16 AHA segments (excluding apex). The mean, standard deviation (SD),
of T1 and T2 within each segment were calculated. RESULTS
SAVA T1-T2
was successful for all subjects within 4.9 ± 0.5 min (mean heart rate 61 ± 7
bpm). Representative whole heart T1 and T2 maps from one
normal human subject are shown in Figures 2 and 3. Homogeneous T1
and T2 values were observed throughout the whole heart as shown by the
histograms in Figures 2B and 3B. High spatial resolution and
coverage allowed excellent delineation of the ventricular structure (Figures
2C and 3C). The image quality of SAVA T1-T2 was
comparable with MOLLI and better than GraSE as shown in Figure 4. SAVA T1-T2
estimated a higher T1 than MOLLI, and T2 comparable to GraSE
(Figure 5).DISCUSSION
In this study we
proposed a highly efficient sequence for simultaneous whole heart T1
and T2 mapping. A very small flip angle was used for the acquisition
of equilibrium signal to reduce the disturbance to the magnetization, thus
removing the next for extended recovery times and increasing efficiency. Additionally,
the use of a dictionary approach required only four volumes for successful
parameter recovery via joint estimation. As MOLLI had a T1
underestimation (12),
SAVA T1-T2 tend to have a better accuracy with higher T1
than MOLLI. Further investigation is warranted on both phantom and in-vivo
experiments on patients. CONCLUSION
The
proposed SAVA T1-T2 sequence can provide efficient whole
heart T1 and T2 mapping with a high resolution of 1.5×1.5×4mm3
in 5 min. Acknowledgements
No acknowledgement found.References
1. Messroghli DR, Moon JC,
Ferreira VM, et al. Clinical recommendations for cardiovascular magnetic
resonance mapping of T1, T2, T2 and extracellular volume: A consensus statement
by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the
European Association for Cardiovascular Imagin. J. Cardiovasc. Magn. Reson.
2017;19:1–24 doi: 10.1186/s12968-017-0389-8.
2. Messroghli DR, Radjenovic A, Kozerke
S, Higgins DM, Sivananthan MU, Ridgway JP. Modified look-locker inversion
recovery (MOLLI) for high-resolution T 1 mapping of the heart. Magn. Reson.
Med. 2004;52:141–146 doi: 10.1002/mrm.20110.
3. Chow K, Flewitt JA, Green JD, Pagano
JJ, Friedrich MG, Thompson RB. Saturation recovery single-shot acquisition
(SASHA) for myocardial T 1 mapping. Magn. Reson. Med. 2014;71:2082–2095 doi:
10.1002/mrm.24878.
4. Sprinkart AM, Luetkens JA, Träber F,
et al. Gradient Spin Echo (GraSE) imaging for fast myocardial T2 mapping. J.
Cardiovasc. Magn. Reson. 2015;17:1–9 doi: 10.1186/s12968-015-0127-z.
5. Giri S, Chung Y-C, Merchant A, et al.
T2 quantification for improved detection of myocardial edema. J. Cardiovasc.
Magn. Reson. 2009;11:56 doi: 10.1186/1532-429X-11-56.
6. Guo R, Chen Z, Herzka DA, Luo J, Ding
H. A three‐dimensional free‐breathing sequence for simultaneous myocardial T 1
and T 2 mapping. Magn. Reson. Med. 2019;81:1031–1043 doi: 10.1002/mrm.27466.
7. Milotta G, Bustin A, Jaubert O, Neji
R, Prieto C, Botnar RM. 3D whole‐heart isotropic‐resolution motion‐compensated
joint T 1 /T 2 mapping and water/fat imaging. Magn. Reson. Med.
2020;84:3009–3026 doi: 10.1002/mrm.28330.
8. Guo R, Cai X, Kucukseymen S, et al.
Free‐breathing simultaneous myocardial T 1 and T 2 mapping with whole left
ventricle coverage. Magn. Reson. Med. 2021;85:1308–1321 doi: 10.1002/mrm.28506.
9. Guo R, Chen Z, Si D, Herzka DA, Luo
J, Ding H. Combined SAturation recovery and Variable flip Angle ( SAVA ) for
free-breathing high- resolution three-dimensional cardiovascular magnetic
resonance T1 mapping at 3T. Proc. 27th Annu. Meet. ISMRM 2019.
10. Pati YC, Rezaiifar R, Krishnaprasad
PS. Orthogonal matching pursuit: recursive function approximation with
applications to wavelet decomposition. In: Proceedings of 27th Asilomar
Conference on Signals, Systems and Computers. IEEE Comput. Soc. Press; 1993.
pp. 40–44. doi: 10.1109/ACSSC.1993.342465.
11. Hamilton JI, Jiang Y, Chen Y, et al.
MR fingerprinting for rapid quantification of myocardial T 1 , T 2 , and proton
spin density. Magn. Reson. Med. 2017;77:1446–1458 doi: 10.1002/mrm.26216.
12. Kellman P, Hansen MS. T1-mapping in
the heart: accuracy and precision. J. Cardiovasc. Magn. Reson. 2014;16:2 doi:
10.1186/1532-429X-16-2.