Kazuo Kodaira1, Michinobu Nagao2, Masami Yoneyama3, Mana Kato1, Takumi Ogawa1, Yutaka Hamatani1, Isao Shiina1, Yasuhiro Goto1, and Shuji Sakai2
1Department of Radiological Services, Tokyo Woman's Medical University Hospital, Tokyo, Japan, 2Department of Diagnostic imaging & Nuclear Medicine, Tokyo Woman's Medical University Hospital, Tokyo, Japan, 3Philips Japan, Tokyo, Japan
Synopsis
Conventional whole-heart-coronary-magnetic-resonance-angiography (WHC-MRA) using NAV has a limitation of long
scan time. Single-breath-hold-multi-shot-gradient-EPI can
significantly reduce scan time of WHC-MRA, but it may be difficult depending
on the gradient-spec of MRI-system. 3D-turbo-field-echo-planar-imaging (TFEPI) with Compressed-SENSE
(C-SENSE) has the possibility to solve these problems. However, it is yet not applied for WHC-MRA. C-SENSE is suitable for subjects with a high
sparse such as vessels, and can accelerate scan time of WHC-MRA while ensuring
image quality. We propose a new combination of TFEPI with C-SENSE for single-breath-hold WHC-MRA, and examine image quality and scan time in comparison to the
conventional methods.
Introduction
Various sequences have been proposed for whole heart coronary magnetic resonance angiography (WHC-MRA)1-4. In conventional WHC-MRA, it is typically performed during
free-breathing, and respiratory motion is prospectively compensated by using a
1D-right-diaphragmatic-navigator (NAV)5. However, an important limitation of the
NAV method is its relatively
long acquisition time6.
Single-breath-hold
technique using Turbo-field-echo-planar-imaging
(TFEPI) could significantly reduce the acquisition time of
WHC-MRA while providing image quality similar to that of conventional
free-breathing gradient echo sequence7. TFEPI
sequence is a hybrid technique that combines TFE and EPI scan7 (Fig.1). It enables
high-speed imaging with EPI readout, resulting in shorter scan time by
allowing more echoes to be acquired during one heartbeat. However, there is a problem that scan time is extended
depending on the gradient-spec of MRI-system. On the other hand, even if SENSE-factor
is increased to shorten scan time, image
quality is expected to degrade due to increased noise.
3D TFEPI with Compressed SENSE (C-SENSE) has the possibility to solve these problems. C-SENSE has recently been introduced as acceleration technique. By combining TFEPI with
C-SENSE (CS-TFEPI), we hypothesized that scan time of single-breath-hold
WHC-MRA can be further accelerated without degrading the image quality by denoising.
In this study, we investigated the feasibility of single-breath-hold CS-TFEPI for WHC-MRA by comparing with single-breath-hold TFEPI with SENSE and conventional
free-breathing TFE.Methods
A total of six volunteers (5 males and 1 female;
age range: 23~44) were examined on a 3.0T-MRI (Ingenia, Philips Healthcare).
The study was approved by the local IRB, and written informed consent was
obtained from all subjects.
We compared image quality and scan time among
WHC-MRA using CS-TFEPI or SENSE, and TFE.
Four sequences were acquired with cardiac triggering.
For respiratory compensation,
TFEPI was acquired with single-breath-hold and TFE was acquired
with NAV.
Imaging-parameters; TFEPI: FOV=300×282mm2, voxel-size=1.56×1.93×3.6mm3, TR/TE/FA=9.3/4.3/20,
TFE-factor=17, EPI-factor=7, shot-duration=158.7ms, SENSE/C-SENSE factor=SENSE:2.5×1.5
(TFEPI-S2.5×1.5) /SENSE:3.6×2.0 (TFEPI-S3.6×2.0) /C-SENSE:3.6×2.0 (TFEPI-CS3.6×2.0);
TFE: FOV=300×288mm2, voxel-size=1.6×1.6×1.8mm3, TR/TE/FA=3.2/1.39/12,
TFE-factor=31, shot-duration=100ms, SENSE-factor=2.0×1.2 (T1TFE-S2.0×1.2).
Each actual scan time was recorded. Curved
Planer Reconstructions (CPRs) were performed using Ziostation2 (Ziosoft Co,
Tokyo) and image quality was evaluated by visual score at the 10-points
(RCA: #1/2/3/4 LAD: #5/6/7/8 CX: #11/13) based on American-Heart-Association classification.
For overall image quality, sharpness and noise and artifacts, we evaluated them
as 4-point grades (grade “4” was excellent, “1” was severe) by two blinded
readers. Visual evaluation was assessed by Steel-Dwass test.
For quantitative comparison, signal-to-noise ratio (SNR) and contrast-to-noise ratio
(CNR) were measured. The SNR was assessed in the blood and myocardium.
To allow quantitative SNR measurements, we used a noise-measurement-method proposed by Zwanenburg et al8. The standard-deviation
of a region of interest of the corresponding area in the noise image was used
as metric for the noise. SNRblood and SNRmyocardium were then calculated as follows:
SNRA = SI(A) / SDnoise(A)
Where SI are the mean average signal intensity of the blood and
myocardium respectively, and the corresponding SDnoise is the standard-deviation at the same location on the noise images.
Subsequently, we measured the
CNR for comparing image contrast quantitatively. The
CNR was estimated for blood and myocardium (CNRblood-myocardium). The CNRblood-myocardium was calculated by the following
equations:
CNRA-B = [SI(A) - SI(B)] / 0.5 [SDnoise(A) + SDnoise(B)]
The SNR and CNR were assessed by one-way repeated measures
analysis of variance (ANOVA) and the post-hoc Tukey-Kramer test.Results
Figure
2, 3 shows the results of visual evaluation. Regarding proximal coronary
arteries, for overall image quality and noise and artifacts, TFEPI-CS3.6×2.0 and T1TFE-S2.0×1.2 were showed significantly
higher than TFEPI-S3.6×2.0 and TFEPI-S2.5×1.5 (p<0.0001). In addition, there was no significant
difference between TFEPI-CS3.6×2.0 and T1TFE-S2.0×1.2. For the sharpness, TFEPI-CS3.6×2.0 was showed significantly lower than T1TFE-S2.0×1.2 (p<0.0001), but significantly higher than TFEPI-S3.6×2.0 and TFEPI-S2.5×1.5 (p<0.0001).
Regarding
distal coronary arteries, for overall image quality, sharpness and noise and
artifacts, TFEPI-CS3.6×2.0 was showed significantly lower
than T1TFE-S2.0×1.2 (p<0.0001),
but significantly higher than TFEPI-S3.6×2.0 and TFEPI-S2.5×1.5 (p<0.001).
Figure 4
shows SNR and CNR comparison among four sequences. For the SNRblood, SNRmyocardium and CNRblood-myocardium, TFEPI-CS3.6×2.0 was showed significantly higher
than T1TFE-S2.0×1.2 and TFEPI-S3.6×2.0 (p<0.0001), except
for the comparison with TFEPI-S2.5×1.5.
Figure 5 shows the
representative images and average scan time using TFEPI-S2.5×1.5, TFEPI-S3.6×2.0, TFEPI-CS3.6×2.0 and T1TFE-S2.0×1.2.Discussion & Conclusion
TFEPI-S2.5×1.5
had a low visual score due to artifacts caused by relatively long breath holding
time, and TFEPI-S3.6×2.0 had a low visual score due to increased
noise with high SENSE factor. On the other hand, TFEPI-CS3.6×2.0 had high
visual score regarding proximal coronary arteries as a result of high SNR and
CNR due to the effect of denoising.
Regarding distal coronary arteries, TFEPI-CS3.6×2.0
had low visual score due to its low spatial resolution and relatively long shot-duration
time. Nevertheless, coronary MRA
is often acquired after contrast-enhancement. By using contrast agent, the diagnostic
ability of the distal coronary arteries can be improved by increasing the blood
SNR. Thus, CS-TFEPI has clinically usefulness because it provides useful information with short breath holding time of about 20 seconds and also reduces the burden on the patient. Furthermore, 4D-coronary-MRA using 3D TFEPI has been introduced9, but it takes a long scan time for whole heart acquisition. This study suggests that CS-TFEPI can be extended to whole-heart 4D-coronary-MRA.Acknowledgements
No acknowledgements
found.References
(1) Kazuo
Kodaira, et al. Whole
heart coronary MRA using image-based 2D
navigator (iNav) and conventional Nav system: comparison of
image quality and scan time. Proceedings of the Annual Meeting
& Exhibition of ISMRM (2021).
(2) Oliver
M. Weber, PhD et al. Free-Breathing, Three-Dimensional Coronary Artery Magnetic
Resonance Angiography: Comparison of Sequences. Journal of Magenetic Resonance
Imaging. 2004; 20: 395-402.
(3) Teresa
Correia et al. Technical note: Accelerated nonrigid motion-compensated isotropic 3D coronary MR angiography. Med Phys. 2018 Jan; 45(1): 214-222.
(4) Yuji
Iyama et al. Single-breath-hold whole-heart coronary MRA in healthy volunteers
at 3.0-T MRI. Springerplus. 2014; 3: 667.
(5) Markus
Henningsson et al. Prospaective Respiratory Motion Correction for Coronary MR
Angiography Using a 2D Image Navigator. Magn
Reson Med. 2013; 69: 486-494.
(6) Mehdi H. Monghari et al. Three-dimensional
Heart Locator and Compressed Sensing for Whole-heart Magnetic Resonance
Angiography. Magn
Reson Med. 2016 May; 75(5): 2086–2093.
(7) Yuji
Iyama et al. Single-Breath-Hold Whole-heart Unenhanced Coronary MRA Using
Multi-shot Gradient Echo EPI at 3T: Comparison with Free-breathing
Turbo-field-echo Coronary MRA on Healthy Volunteers. Magn
Reson Med Sci. 2018; 17: 161-167.
(8) Zwanenburg et al. MR Angiography of the
cerebral perforating arteries with magnetization prepared anatomical reference
at 7T: comparison with time-of-flight. J Magn Reson Imaging. 2008; 28:
1519–1526.
(9) Yasuhiro Goto, et al. Feasibility of
Non-contrast dynamic TFEPI 4D-coronary MRA with PPU. Proceedings of the Annual Meeting & Exhibition
of ISMRM (2021).