Hirohito Kan1,2, Kyosuke Mizuno3, Masahiro Takizawa4, Masashi Shimohira5, Tatsuya Kawai5, Tositaka Aoki6, Satoshi Tsubokura6, and Harumasa Kasai6
1Department of Integrated Health Scieneces, Nagoya University Graduate School of Medicine, Nagoya, Japan, 2Department of Radiology, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan, 3Department of Radiology, Nagoya City University Hospital, Nagoya, Japan, 4FUJIFILM Healthcare Corp., Tokyo, Japan, 5Department of Radiology, Nagoya City University Graduate School of Medical Scieneces, Nagoya, Japan, 6Nagoya City University Hospital, Nagoya, Japan
Synopsis
To develop and validate the
non-contrast time-resolved MR angiography (NC TR-MRA) with consecutive beam
pulse with variable saturation flip angle method, we performed a phantom study
using steady flow and the volunteer study targeted to the pulmonary artery. The
length of visualized flow signals was exhibited proportional to the number of
beam saturation pulses in the phantom study. In human study, the visualization
range of the pulmonary artery was also extended with increasing the number of
beam pulses. The novel consecutive beam pulse with variable saturation flip
angle method can depict blood vessels in each flow phase.
Introduction
Time-resolved MR angiography
(TR-MRA) using contrast agents helps diagnose recanalization of pulmonary
arteriovenous malformation (PAVM) in patients undergone coil embolization1.
However, the use of contrast agents increases the risk of a paradoxical
embolism if recanalization is present. The TR-MRA without contrast agents may
be a more appropriate method for a survey of recanalization and follow-up of the
lesions that underwent coil embolization. To develop and validate the
non-contrast TR-MRA (NC TR-MRA) with consecutive beam pulse with variable saturation
flip angle method, we performed a phantom study using steady flow and the volunteer
study targeted to the pulmonary artery.Methods
All studies were conducted on 3 T
MRI (FUJIFILM
Healthcare Corp.).
Non-contrast time-resolved MR
angiography
This sequence is based on the
three-dimensional fast spin-echo sequence (FSE) with respiratory trigger (Figure
1a). The beam saturation pulses2 consecutively irradiate an upstream
of blood flow at 20 ms interval. The saturated spins are recovered by the
longitudinal relaxation during the time gap between the main scan and each
pulse. It is needed to vary the flip angle of each beam saturation pulse
according to the T1 value of venous blood and the time gap. The flip angle θFA
of beam saturation pulse is
determined by the time gap t and T1 value according to Equation 1.
$$$\theta_{FA}=\arccos(1-\frac{1}{\exp(\frac{-t}{T1})}) $$$ (1)
Increasing the number of beam
saturation pulses extends the distance of the vessel that the blood signal
vanishes. Since the scan without beam saturation pulse depicts all ranges of
blood vessels, the NC TR-MRA at each flow phase is calculated by subtracting each
scan with a varying number of beam saturation pulses from the scan without the
saturation pulse (Fig. 1b).
Phantom study
The steady flow phantom consisted
of the flow pump (Micro pump MC320, Newa, Italy) and polyvinyl chloride tube of
10-mm inside diameter shown in Figure 2. The T1 value of the flowing liquid was
adjusted to 1615 ms to simulate venous blood3,4. Three different
flow velocities were employed to simulate physiological pulmonary arterial
blood flow based on previous literature5; 14.2, 26.2, and 34.6 cm/s.
The number of beam saturation pulses was varied as 0, 5, 10, 15, 20, 25, and
30. The parameters of FSE sequence were the following: Field of view (FOV); 224
× 224 mm2, matrix size; 128 × 128, slice thickness; 3 mm, number of
reconstruction slice; 140, effective TE; 4.2 ms, and TR; 4000 ms assuming an
interval of respirations. We evaluated the line profiles along the flow
direction and visualization range of flow signal of maximum intensity
projection (MIP) in the NC TR-MRA at each flow phase.
Human study
A healthy volunteer (29 years old,
male) was recruited to test a novel NC TR-MRA. The number of beam saturation
pulses was varied as 0, 3, 6, 9, 15, and 24. The FSE sequence parameters were
set in a similar fashion to the phantom study, excepting the FOV; 350 × 350 mm2,
matrix size; 256 × 176, slice thickness; 4 mm, and the number of reconstruction
slice; 110. To evaluate the change in signal intensity of MIP of original
images at each flow phase, the ROIs were set on the right interlobar pulmonary artery
and the proximal and distal portions of the right basal artery, as shown in
Figure 5b.Results
The representative MIPs of NC
TR-MRA in the phantom study are shown in Figure 3. The length of visualized
flow signals was exhibited proportional to the number of beam saturation
pulses. The signal profiles along the flow direction were consistent with the
visual findings (Figure 4a). The difference in flow velocity was also observed as
the change in signal profiles under the fixed number of saturation pulses (Figure
4b). The MIPs of NC TR-MRA in the right pulmonary artery were shown in Figure 5a.
The MIP of original images varying the number of saturation pulses demonstrated
signal fading proximally as the number of pulses decreased. Many beam
saturation pulses were needed to achieve sufficient signal suppression in the
distal branch (Figure 5c).Discussion
The visualized range of steady flow
in the phantom study was proportional to the number of beam saturation pulses.
The depicted distances of the flow signals were excellently dependent on the flow
velocities. The multiple beam saturation pulses with variable saturation flip
angle allow effectively visualizing the chronological change in flow signal almost
without the signal loss due to the longitudinal relaxation during the time gap.
In the human study, the visualization range of the pulmonary artery was also extended
with increasing the number of beam pulses but not in a proportional manner to
the number of beam saturation pulses, because of a pulsatile flow in the
pulmonary artery and the difference in flow velocity. The advantages of NC
TR-MRA are that the repeatable examination can be performed because the
contrast agents are not needed and that the temporal resolution of 20 ms is
higher than that of the contrast-enhanced TR-MRA, which is around 1 s.Conclusion
The NC TR-MRA with beam saturation
pulse may be applicable to the PAVM and help diagnose recanalization and
follow-up in patients who underwent the coil embolization.Acknowledgements
No acknowledgement found.References
[1]
Kawai T, Shimohira M, Kan H, et al. Feasibility of Time-Resolved MR Angiography
for Detecting Recanalization of Pulmonary Arteriovenous Malformations Treated
with embolization with Platinum Coils. J Vasc Interv Radiol 2014; 25(9):
1339-47.
[2]
Nishihara T, Itagaki H, Moriwake C, et al. Selective TOF MRA Using Beam
Saturation Pulse. Proc Intl Soc Mag Reason Med 2012; 20: 3698.
[3]
Lu H, Clingman C, Golay X, et al. Determining the Longitudinal Relaxation Time
(T1) of Blood at 3.0 T Tesla. Magn Reason Med 2004; 25: 679-682.
[4]
Wu WC, Jain V, Li C, et al. In vivo Venous Blood T1 Measurement Using Inversion
Recovery TrueFISP in Children and Adults. Magn Reason Med 2010; 64(4):
1140-1147.
[5]
Wehrum T, Hagenlocher P, Lodemann T, et al. Age Dependence of Pulmonary Artery
Blood Flow Measured By 4D Flow Cardiovascular Magnetic Resonance: Results of
Population-based Study. J Cardiovasc Magn Reason 2016; 18(1): 3.