Elizaveta Motovilova1,2, Frida Maria Galaviz Huerta3, Jana Vincent4, James Shin1, Ek Tsoon Tan2, Fraser Robb4, Victor Taracila4, Darryl Sneag2, and Simone Angela Winkler1
1Department of Radiology, Weill Cornell Medicine, New York, NY, United States, 2Department of Radiology and Imaging, Hospital for Special Surgery, New York, NY, United States, 3Department of Psychology, New York University, New York, NY, United States, 4GE Healthcare, Aurora, OH, United States
Synopsis
It
is commonly undesirable to see an RF coil in MR images. Novel research into
innovative coil designs often involves the use of new materials that can be
visible in MR images. In this abstract, we address this limitation by doping
the coil substrate with Magnevist, drawing from a little-known phenomenon of
signal hypointensity via T2-shortening for large concentrations of
gadolinium-based (Gd) contrast agents that is especially applicable to
hydrogen-rich materials. We show successful reduction of MR signal of >60%
using our previously developed stretchable, liquid-metal based Ecoflex coil (6)
using relaxometry measurements and in vitro experiments.
Introduction
In MR imaging, it is commonly
undesirable to see the outline of the RF coil used because the expectation is
for the focus to be on the anatomy. Novel research into innovative coil designs
often involves the use of new materials that can be visible on MR images. It
would be a clear impediment to limit the selection of coil materials to ensure
their invisibility and could severely limit the potential of these novel
designs. An example of such innovative coil technology is in the arena of
stretchable/flexible designs: liquid metal-based radiofrequency (RF) coils1-6
build on the inherent stretchability of liquid conductors but also require a
suitable stretchable substrate. In our previous work (6),
we used pure Ecoflex and observed strong MR signal from the polymer, which
motivated the quest for a universal solution to reduce signal intensity in arbitrary
coil materials. In this abstract, we address this limitation by doping the coil
substrate with Magnevist, drawing from a little-known phenomenon of signal
hypointensity stemming from T2 shortening for large concentrations of gadolinium-based
(Gd) contrast agents. We show successful reduction of MR signal using the
Ecoflex coil example6.Methods
Theory. For a gradient-recalled-echo (GRE) sequence with given flip
angle $$$\alpha$$$, repetition time TR,
and echo time TE, signal intensity depends on T1 and T2
relaxation times7
$$SI_{GRE}\sim sin(\alpha)\cdot \frac{1-e^{-TR/T_1}}{1-cos\alpha\cdot e^{-TR/T_1}}$$ [Eq.1].
Gd contrast agents facilitate
relaxation of nearby hydrogen protons and shorten both T1 and T2.
For T1-shortening, increased signal intensity occurs, but for T2-shortening,
and especially at high concentrations of gadolinium, decreased signal intensity
occurs7.
Our goal is to utilize T2-shortening to reduce the appearance of
Ecoflex polymer on MR images.
Sample preparation. Three
samples of commonly used Ecoflex polymer (00-10, 00-30, and 00-50) were prepared
by mixing the corresponding crosslinker and prepolymer at a ratio of 1:1. For the
doped samples, Ecoflex 00-30 was mixed with different volume ratios of a Gd
contrast agent (Magnevist) (Figure 1a). After mixing, all samples were degassed
in a vacuum chamber to remove air bubbles, transferred to graduated plastic test
tubes, and cured at room temperature (Figure 1b).
Relaxometry measurements. Samples
were equidistantly arranged around a standard homogeneous cylindrical phantom
(OD=9.5cm, L=30cm, Figure 2a), positioned centrally inside a 32-channel head
coil, and scanned on a 3T MRI scanner (GE Healthcare, MR750). T2
mapping was performed by acquiring a single axial slice with a spin-echo (SE) sequence
(TR=1500ms, slice thickness 5mm, 32 TEs from 8.3-133ms). Post-processing was performed
using Matlab, with circular, manually drawn, regions of interest to determine
the average signal intensity. The T2 value of each sample was found
by fitting the average signal intensity to the following equation $$SI_{avg.}=S_0\cdot e^{-TE/T_2}$$ [Eq.2].
Coil fabrication and bench
tests. Three
geometrically identical coils were fabricated as outlined previously6. The polymer substrate was fabricated using i)
pure Ecoflex 00-30 polymer and Ecoflex 00-30 mixed with Magnevist at a ratio of
ii) 5:100 and iii) 1:10 (Figure 3a). S11 parameters (Figure 3b) and unloaded quality
factor were measured using a vector network analyzer (Keysight, E5071C).
Phantom imaging. The
coils were placed on a standard rectangular silicone phantom (W=22cm, L=33cm,
H=16cm) for imaging also at 3T using an FSE sequence (TR=315ms, TE=25.5ms, FOV=24cm,
pixel size 0.9x0.9mm, slice thickness=3mm, ETL=2).Results
Relaxometry
measurements. Figure
2b shows one of the SE images acquired with an intermediate echo time (TE) of 33.2ms.
The signal is decreased in samples with high Magnevist content (samples
#10-12). With the addition of Magnevist, the signal increases by up to 24% at 1:1,000
concentration (Figure 2c). As concentration is increased further, the signal
gradually drops to only 30% of its initial value at a concentration of 1:10.
Figure 2d shows T2 with respect to the Magnevist concentration to remain
relatively stable around 180ms (up to 1:10,000) and then to drop to 39%
(70.35ms) of its original value for higher concentrations.
Coil fabrication and bench
tests. The three coil prototypes exhibited similar S11 parameters (-27+/-3dB)
and Q factors (22.2+/-1.5) indicating maintained frequency stability and
unloaded losses.
Phantom
imaging.
Figures
4a-c show FSE images of a homogenous phantom acquired with the 3 coils. The
coils doped with Magnevist were less visible on the MR scans, yielding an
imaging signal reduction of up to 83% and 92% for 5:100 and 1:10 ratios,
respectively. Figures 4d-f show the corresponding SNR maps calculated by
dividing the signal maps (a)-(c) by the standard deviation of the background noise.
The SNR remains stable for all coils (426±5).
Figures 4g-I show the signal intensity measured inside the phantom along the
vertical line going through the center of each coil, confirming that doping of
the polymer with Magnevist does not deteriorate coil sensitivity.Conclusions
This study demonstrates that
doping coil materials with Gd-based contrast agents (Magnevist) can reduce their
appearance in MR images without sacrificing coil sensitivity. Signal intensity
reduction depends on sequence parameters, but even for a SE sequence with a
relatively short TE=33.2ms, signal reduction is greater than 60% in an Ecoflex
substrate. This method could also prove interesting for other applications, such
as the design of phantoms with low MR signal intensity, e.g. for the mimicking
of lung tissue or bone.Acknowledgements
This work was supported by the
National Institutes of Health under NIH R00EB024341, and GE Healthcare. The
authors would like to acknowledge Muc Chu, Jojo Borja, and Jonathan Dyke of the
Citigroup Biomedical Imaging Center for the helpful technical discussions, and Dr.
Hollis Potter of the Hospital for Special Surgery for the guidance and research
support of the project.References
1. Varga
M, Mehmann A, Marjanovic J, Reber J, Vogt C, Pruessmann KP, et al. Adsorbed
eutectic GaIn structures on a neoprene foam for stretchable MRI coils. Advanced
Materials. 2017;29(44):1703744.
2. Mehmann A, Varga M, Vogt C, Port A,
Reber J, Marjanovic J, et al. On the bending and stretching of liquid metal
receive coils for magnetic resonance imaging. IEEE Transactions on Biomedical
Engineering. 2018;66(6):1542-8.
3. Mehmann A, Vogt C, Varga M, Port A,
Reber J, Marjanovic J, et al. Automatic resonance frequency retuning of
stretchable liquid metal receive coil for magnetic resonance imaging. IEEE
transactions on medical imaging. 2018;38(6):1420-6.
4. Port A, Luechinger R, Albisetti L,
Varga M, Marjanovic J, Reber J, et al. Detector clothes for MRi: A wearable
array receiver based on liquid metal in elastic tubes. Scientific Reports.
2020;10(1):1-10.
5. Port A, Luechinger R, Brunner DO,
Pruessmann KP. Elastomer coils for wearable MR detection. Magnetic Resonance in
Medicine. 2021;85(5):2882-91.
6. Motovilova E, Tan ET, Taracila V,
Vincent JM, Grafendorfer T, Shin J, et al. Stretchable self-tuning MRI receive
coils based on liquid metal technology (LiquiTune). Scientific reports.
2021;11(1):1-10.
7. Rohrer
M, Bauer H, Mintorovitch J, Requardt M, Weinmann H-J. Comparison of magnetic
properties of MRI contrast media solutions at different magnetic field
strengths. Investigative radiology. 2005;40(11):715-24.