Gregory Peng1, Can Wu2, and Qi Peng1
1Department of Radiology, Albert Einstein College of Medicine, Bronx, NY, United States, 2Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
Phase-cycling (PC) has been proposed as an
effective method to minimize the impact of B0 and B1
field inhomogeneities in quantitative T1rho mapping, which needs both positive
and negative longitudinal magnetization T1rho preparations. We propose here a
simple magnetization inversion with an adiabatic inversion RF pulse before an
established T1ρ preparation module to achieve an opposite PC. This approach has
been shown to be much less sensitive to B0/B1 field imperfections compared with
a traditional approach of reversing the phase of the last 90° RF
pulse.
Introduction
Quantitative
T1rho (T1ρ) mapping MRI has been shown to be sensitive to pH, macromolecule
density, and metabolic dynamics in tissues such as the cartilage, muscle, and
brain1. One of the
technical challenges in quantitative T1ρ mapping is its high sensitivity to B0
and B1 field inhomogeneities. Phase-cycling (PC) has been proposed
as an effective method to minimize the impact of B0 and B1
field inhomogeneities, and to eliminate the signal contaminations due to T1
relaxation along a GRE readout train.2-4 In a PC sequence, both positive
(Mz+) and negative (Mz-) magnetization-prepared signals are acquired and complex
data subtracted before T1ρ-decay curve fitting. To achieve the opposite PC
preparation, the phase of the last 90° hard pulse can be inversed in phase,
which readily leads to high sensitivity to B1 field inhomogeneities.
Alternatively, two different T1ρ preparation for Mz+ and Mz- can be applied,4 which complicates
the sequence design and potentially introduces extra quantitative deviations. In
this work, we propose magnetization inversion with an adiabatic inversion RF
pulse before a T1ρ preparation module to achieve both a simple sequence
design and insensitivity to B0/B1 imperfections. Methods
A 3D MAPSS sequence based on segmented GRE
readouts was utilized in this study (Figure 1A). The reference T1ρ preparation, herein referred to as
the composite spin-echo (CSL) preparation,5 had a 90°x-TSL/2-180°y-TSL/2-90°x
RF pulse train (termed “CSL-” preparation
herein, where “-” means magnetization prepared to -Mz), which has
been shown to be robust to both B0- and B1-imperfections (Figure
1A). Two different T1ρ
preparations leading to +Mz based on CSL- were tested. The
first had 90°x-TSL/2-180°y-TSL/2-90°-x
as shown in Figure 1B (termed CSL+), with the last 90° RF pulse having an
opposite phase compared to that of the CSL- preparation. In the second method (InvCSL-),
an adiabatic inversion RF pulse (9.21ms, BW=1.38kHz) followed by a gradient
spoiler was inserted before the CSL- module (Figure 2C). All imaging was
performed on a 3T Philips Ingenia scanner with a 1ch-TX/16ch-Rx knee coil on a phantom consisted of three pairs
of lab tubes with agarose concentrations of 2, 3, and 4 % (Figure 2A.). These three T1ρ
preparations were evaluated in a 3D MAPSS sequence with
spinlock frequency of 300Hz and five times of spinlock (TSL) of 0, 17,
34, 51, and 68ms. Other scan parameters were: FOV=140/140/120mm, voxel size=1x1x4mm3,
TR/TE=5.4/2.6ms, TFE factor=16 with centric profile ordering, and compressed
SENSE factor=3. To evaluate the sensitivity to B1 imperfections in each T1ρ preparation,
each sequence was repeated with 90° and 180° hard RF
pulses manually misadjusted to 0.9 and 0.8 times of the nominal B1 amplitude.
Six 3D ROIs from six tubes were generated after eroding seven boundary voxels in
all three directions. The
same 3D ROIs were applied to all the 3D T1ρ maps from different B1
scales to measure their T1ρ mean and standard deviation (SD). Additionally, a reference T1ρ was measured for each T1ρ
preparation method from the central phantom slice of each tube with negligible
B0 and B1 imperfections to calculate the mean percent
error for each preparation approach. Results
Without
manually adjusting B1 scale to 0.8 and 0.9, all three T1ρ
preparations resulted in uniform T1ρ maps on the center-slice for each tube, as
shown in Figure 2D. On the off-center slices, both the CSL- and InvCSL- had
similar results as in Figure 2D, but CSL+ had increased T1ρ deviations from the
center slices (Figure 2E). When B1 was misadjusted, T1ρ maps from
both CSL- and InvCSL- preparations had minimal changes compared with maps from
B1=1 (Figure 2G), while CSL+ preparation induced much larger deviations
as shown in Figure 2F. Comparison of the impact of B1 misadjustment
on all three preparations are shown in the column graphs in Figure 3. Both CSL- and InvCSL- preparations had
consistent T1ρ results at different B1 scales, while average T1ρ
results from CSL+ preparation changed drastically with increasing B1
misadjustment. All quantitative results including the T1ρ reference values from
the center 2D ROIs and mean±stdev from the 3D ROIs are summarized in Table 1. Discussion
PC
has been shown to be an effective method to reduce T1ρ measurement uncertainty
due to B0/B1 field inhomogeneities. An unpaired PC strategy has also
been proposed recently to avoid doubled scan duration needed for PC.4 Nevertheless, two
T1ρ preparation modules leading to both Mz+ and Mz- are still
both needed. The final quantitative quality relies on the quality of both T1ρ preparations.
In this work, we present a simple adiabatic inversion before the T1ρ
preparation module to achieve opposite PC for a hard-pulse T1ρ preparation. It
is shown in this phantom study that InvCSL- had similar performance to B0/B1
field inhomogeneities as the original CSL- preparation, much improved compared with
the traditional CSL+ preparation. Although not shown here, this approach has also
been tested on adiabatic T1ρ to be much less sensitive to B0/B1 field imperfections
compared with an earlier presented approach.6Conclusion
Adiabatic inversion before a T1ρ preparation achieves
much improved insensitivity to B0/B1 inhomogeneities compared with a traditional
approach of reversing the phase of the last 90° pulse. Acknowledgements
No acknowledgement found.References
1. Borthakur
A, Sochor M, Davatzikos C, et al. T1rho MRI of Alzheimer's disease. Neuroimage.
2008;41(4):1199-1205.
2. Li X, Han ET, Busse RF, et al. In
vivo T(1rho) mapping in cartilage using 3D magnetization-prepared
angle-modulated partitioned k-space spoiled gradient echo snapshots (3D MAPSS).
Magn Reson Med. 2008;59(2):298-307.
3. Chen W, Takahashi A, Han E.
Quantitative T(1)(rho) imaging using phase cycling for B0 and B1 field
inhomogeneity compensation. Magn Reson Imaging. 2011;29(5):608-619.
4. Peng Q, Wu C, Kim J, et al.
Efficient Phase Cycling Strategy for High Resolution 3D GRE Quantitative MRI
Mapping. NMR in Biomedicine. 2021;Under Review.
5. Witschey WR, 2nd, Borthakur A,
Elliott MA, et al. Artifacts in T1 rho-weighted imaging: compensation for B(1)
and B(0) field imperfections. J Magn Reson. 2007;186(1):75-85.
6. Wu
C, Peng Q. 3D MAPSS T1ρ Mapping with Adiabatic Spin-Locking RF Pulses for Knee
and Brain at 3T. Proc Intl Soc Mag Reson Med 29. 2021.