Manjunathan Nanjappa1, Brian Raterman1, Bradley D. Bolster, Jr2, Kannengiesser Stephan3, Ning Jin2, and Arunark Kolipaka1
1Department of Radiology, The Ohio State University, Columbus, OH, United States, 2MR Collaborations, Siemens Medical Solutions USA, Malvern, PA, United States, 3Siemens Healthineers, Erlangen, Germany
Synopsis
Magnetic resonance elastography
(MRE) has become a standard clinical tool to stage liver fibrosis and it is
important to optimize the imaging protocols to achieve robust stiffness
measurement. Imaging was performed using a 3T MRI scanner on 20 healthy adult volunteers
with varied driver powers and acquisition parameters. The stiffness value at
60Hz across the settings was found to be stable for all the subjects. 40-50%
driver power with 20-35mT motion encoding gradient amplitude would deliver
adequate shear waves to achieve diagnostic quality of liver stiffness maps.
Introduction
Magnetic Resonance Elastography
(MRE) is a non-invasive technique to estimate the mechanical properties of soft
tissues. An alternative to palpation, MRE is particularly useful to diagnose
liver disease, because it eliminates the need for needle biopsy or surgical
procedure [1]. Given its diagnostic
accuracy, in recent years MRE has become a standard clinical tool to stage
liver fibrosis. MRE quality, such as the size of confidence
region and the reliability of stiffness value, can be sensitive to where the
passive driver is positioned on the patient, the power of the active driver
applied, and imaging protocols, i.e., the amplitude of motion encoding gradient
(MEG) etc. This study was aimed to demonstrate the optimized passive driver
positioning on patient and find the shear wave driver power and MEG strength needed
to achieve robust image quality.Methods
All imaging was performed using a
3T MRI scanner (MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany) on 10
healthy volunteers (ages 21-65 years, 5
M) with no known health problems. Body mass index (BMI) and belly circumference
(BC) was measured after obtaining written informed consent. The passive driver
(Resoundant Inc, Rochester, MN USA) was placed on the upper abdomen at the
fifth intercostal ribcage but more to the lateral side. Scout images of
sagittal and axial views were acquired to verify the placement of the driver
(figure 1).
A prototype single shot spin echo-
echo planar imaging (SE-EPI) MRE sequence was used to acquire axial slices of
the liver with the following parameters: FOV:320x320mm, TE:47ms, TR:1000ms,
slice thickenss:6mm, acquisition matrix:100x100, bandwidth: 2380Hz/Px, motion
encoding gradient (MEG) frequency:60Hz, MEG amplitude:20, 25, 30 and 35mT/m;
Only through plane motion was encoded, and GRAPPA with acceleration factor of 2
with 24 GRE reference lines was used. The Spectrally Selective Adiabatic Inversion
Recovery (SPAIR) technique was used to suppress the fat. Excitation frequency was 60.1Hz and driver
amplitude was varied to 30%, 50%, and 70%. The magnitude images were post
processed to compute the first harmonic amplitude (FHA) of the displacement
field and the stiffness maps were calculated using a multi-model direct inversion algorithm (MMDI) [2]. Stiffness
values and FHA were reported using a semi-automatic region of interest (ROI)
selection technique in the liver. ROI areas were reported as a percentage of
the segmented liver area.Results
The representative MRE magnitude
image, wave image and stiffness maps from one volunteer acquired with 30%
driver amplitude and 30mT/m MEG are shown in figure 2. The BMI of the volunteers
ranged from 17.98 to 30.18 with a mean of 24.10 ± 3.44 and the BC from 28.5 to
42.8 inches with mean of 34.40 ± 4.33 inches. FHA ranged from 4.23 to 12.84
µm/rad and increased with MEG amplitude and driver power. In the stiffness
maps, globally the number of pixels inside the liver region with 95% confidence
increased gradually (though marginally) with increasing MEG amplitude between
20mT to 30mT at varying driver power, and then plateaued at 35mT (Figure 3).
The mean liver coverage with 95%
confidence was found to be maximum between 40-50% power levels and the coverage
ratio ranged from 64% to 85% as shown in figure 3. The stiffness value across
the settings was found to be stable for all the subjects though one subject had
significantly higher stiffness as compared to the others (figure 4).Discussion
Passive driver placement more to
the lateral side of the abdomen (as shown in the scout images figure 1) helps to
achieve better wave penetration and anatomical coverage of the liver. About 40%
driver power was determined to be adequate to generate reliable stiffness
estimates for all body habitus irrespective of belly circumference or BMI.
As expected, the stiffness was not altered
with increase in driver power. However, at increased power levels (i.e.
>50%) a deeper portions of the liver were covered in the stiffness map (with
95% confidence interval). But with higher power, the periphery experienced
inter-voxel phase dispersion due to overdriving leading to unreliable
measurements in the superficial liver and therefore the net area of the
confidence region in stiffness map didn’t increase.
The ratio of the 95% confidence
region of stiffness map to the segmented liver area ranged from 62% to 78% with
increasing driver power. Major portions of the liver could be used when
reporting stiffness but the %area appears lower than 90%. This measure could be
biased lower because the calculated liver area includes vessels which are
excluded from the measured ROI. Additionally, when the liver is soft the
measurements at deeper portions of the liver becomes unreliable due to wave
attenuation. Even in those settings the coverage of reportable stiffness values
is still very good.Conclusion
In this study we demonstrated that
with a lateral driver positioning and good driver contact with the body, the
ability to achieve large measurable stiffness ROIs is relatively insensitive to
driver power and MEG amplitude over a large range of values while remaining
independent of body habitus as characterized by BMI and BC. 40-50% power at
60Hz with 20-35mT/m motion encoding amplitude was shown to result in measurable
shear waves in adult subjects sufficient to achieve diagnostic quality liver
stiffness maps. Acknowledgements
N/AReferences
1. Muthupillai
R, Lomas DJ, Rossman PJ, et al. Magnetic resonance elastography by direct
visualization of propagating acoustic strain waves. Science 1995;
269(5232):1854–7.
2. Silva AM, Grimm
RC, Glaser KJ, et al. Magnetic resonance elastography: evaluation of new
inversion algorithm and quantitative analysis method. Abdom Imaging 2015;40(4):810–817.