Ke Zhang1,2, Simon M.F. Triphan1, Felix T. Kurz2, Chrstian H. Ziener2, Hans-Ulrich Kauczor1, Heinz-Peter Schlemmer2, and Oliver Sedlaczek1,2
1Department of Diagnostic and Interventional Radiology, Heidelberg University Hospital, Heidelberg, Germany, 2Department of Radiology, German Cancer Research Center, Heidelberg, Germany
Synopsis
Renal
perfusion is an important physiological parameter in health and disease (1).
To measure kidney perfusion
using arterial spin labelling (ASL) the respiratory motion is a major problem. In this study, respiratory
motion information is acquired from a projection signal and used to adjust the
position of the excited slice in real time. The feasibility of free-breathing
multi-slice kidney perfusion imaging using EPI
based pseudocontinuous ASL (pCASL)
with navigator-based slice tracking method is investigated.
PURPOSE
To
apply the navigator-based slice tracking method to prospectively compensate the
respiratory motion for kidney pseudo-continuous arterial spin
labeling (pCASL) using EPI acquisition.METHODS
Measurements
were performed using an 18-channel body and spine receive RF coil on a 1.5T Aera
Siemens scanner (Siemens Healthineers AG, Erlangen, Germany). A single
gradient-echo slice selection and projection readout at the location of the
diaphragm along the inferior-superior (IS) direction is acquired as a
navigator. Navigator acquisition and fat suppression were interleave inserted
before each transverse imaging slice of the readouts of 2D gradient-echo EPI based
pCASL sequence (Fig. 1), with 45 measurements including 4000 training
navigators at the beginning of measurements obtained in 5 minutes. Sequence
parameters were as follows: TE =12 ms, FOV=400×240 mm3, in-plane iPAT factor=3,
matrix size=120×72×8, resolution=3.3×3.3×8 mm3, slice gap = 4 mm, labeling
duration = 1650 ms, postlabeling delay (PLD) =2000 ms, TR = 6500 ms, TI for
SPAIR (SPectral Attenuated Inversion Recovery) fat suppression=90 ms, FOVnav =
200 mm, resnav = 64, Flip anglenav = 15º, TRnav=4.22ms. The labeling pulse train and
readout were triggered by the pulse signal and played out only during the
systolic period. The renal blood flow was
selectively tagged using a tagging plane placed above to the diaphragm,
labeling the arterial blood in the descending aorta.
Before motion analysis, the interleaved navigator
signals during image acquisition were Fourier transformed and truncated to
exclude RF saturation along IS direction from the EPI readout (Fig. 2). The position
for this truncation was calculated based on peaks fitted from the averaged training
navigator and the peaks from the first 32 interleaved navigators. The diaphragm
position was derived by calculating the phase difference of the interleaved
navigator signals at each acquisition after Fourier transform and truncation. The
unwrapped data from different coils were then combined by using coil clustering
(2) based on the first 32 interleaved
navigators. The motion information was then directly sent back to the sequence
and slice positioning was adjusted in real-time. This motion analysis and
real-time feedback was performed on the scanner, implemented in ICE (Image
calculation environment, Siemens Healthineers AG, Erlangen, Germany).RESULTS
The respiratory motion
from navigator signals could be precisely calculated (Fig. 2) and slice
positioning was changed in real-time based on the motion information. Comparing
to the case without, motion compensation reduced liver movement in a selected
slice (Fig. 2). Fig. 3 shows spatial maps of the coefficient of variation (CoV)
of control images from a
represent volunteer in kidney. The value of CoV was calculated as the standard
deviation of the constituent voxels divided by their mean. Motion greatly
increased the CoV, but to a lesser extent when the motion correction was on. The
calculated perfusion maps show significantly higher perfusion in the renal
cortex from navigator-based motion compensated pCASL than from no motion
compensated pCASL (Fig. 4, 5). More detailed structure in the perfusion maps
could be observed in the motion compensated case (Fig. 4, 5).CONCLUSION
This
study demonstrates the feasibility of navigator-based slice tracking technique
in kidney pCASL using EPI readout. The sequence may improve the evaluation of the
perfusion in kidney diseases.Acknowledgements
No acknowledgement found.References
1. Dalal R, Bruss ZS, Sehdev JS.
Physiology, Renal Blood Flow and Filtration. StatPearls. Treasure Island (FL);
2021.
2. Zhang T, Cheng
JY, Chen YX, Nishimura DG, Pauly JM, Vasanawala SS. Robust Self-Navigated Body
MRI Using Dense Coil Arrays. Magn Reson Med 2016;76(1):197-205.