Mark Gosselink1, Aris van Ieperen1, Wout Schuth2, Martino Borgo2, Dimitri Welting1, Edwin Versteeg1, and Dennis Klomp1
1Department of Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Futura composites, Heerhugowaard, Netherlands
Synopsis
We propose a setup to boost spatiotemporal encoding for MRI in the human brain using a conventional receiver
array (32ch) and fast gradients (6100T/m/s). Moreover, we demonstrate that the operation frequency of a high-end gradient-amplifier
can be increased to ultrasonic frequencies so to avoid unpleasant acoustic
noise. By using this amplifier with 8 RF transmitters, 32 receivers and a 2-channel
gradient coil, a light-weight setup is constructed for operation in a 7 Tesla MRI system.
The setup is demonstrated for operation in the human brain showing good FLAIR
data and no peripheral nerve stimulation at the slew rate of 6100T/m/s.
Introduction
Silent MRI can be obtained when using an extra gradient
insert that can drive an EPI readout above the human hearing frequency and can
be operated in conjunction with the conventional body gradients [1]. When
driving gradients in EPI mode, the selwrate can be increased by an order of
magnitude before causing peripheral nerve stimulation. Here we propose a
light-weight design that includes a second gradient axis, so to facilitate
spiral readouts above 20kHz, and investigate if human brain MRI performance is
maintained and peripheral nerve stimulation remains absent.Methods
The gradient amplifier used is a NG500 1.1 (Prodrive
Technologies, Son, Netherlands). The FPGA firmware of the gradient amplifier
has been modified to allow for a higher frequency range output. The new
firmware modifies the output driver to allow the output of a harmonic of the
base frequency, increasing the output frequency of one amplifier to 23kHz and t
a second amplifier to 27kHz. The internal end-stage filter was removed and
brought to the gradient coil as a resonant circuit.
A dual axis gradient insert was built (Futura, the
Netherlands) that has been made with windings in the Y and Z axes, (Fig 1). The
high driving frequency of the insert allowed for minimal use of epoxy, as
minimal movement is expected due the inertia of the coil above 20 kHz. Therefore,
the entire setup weighs less than 25kg. Capacitors have been added to the
inductive windings to filter unwanted amplifier currents while enhancing
gradient field efficiency.
To minimize eddy currents, no RF shield was used
between the insert gradient and the RF coils. Eight ultrathin (2mm) 7T 1H
dipoles transmitter, matched to ~50 ohm at 298MHz, are integrated into the
design so to provide sufficient space for a standard 32 channel receiver array
(NOVA, USA).
To measure the efficacy of the ultrasonic gradient
setup a field camera (Skope, Swiss) has been placed inside the insert gradient
and RF setup.
In vivo MRI scans were obtained with the entire setup.
First, constructive B1+ phase shimming was applied, followed by a B1 map to
verify flip angle setting. Then a FLAIR sequence was obtained using the 32
receivers and 8ch transmitters in the presence of the insert gradient. Finally,
the 6100mT/m/s (for Z) and 3040mT/m/s (for Y) were applied for 64 times in 10ms
shots to the human brain of a healthy volunteer to verify absence of peripheral
nerve stimulation. Results
With the gradient amplifier driven at less than 30% of
its range, already 6100T/m/s at 42mT/m (Z) and 3040mT/m/s at 18mT/m (Y) was
obtained, resulting in the expected operating frequency of 23 and 27kHz.
In the presence of the gradient insert, the 8 channel
dipole array could obtain 15.6uT in the human brain (Fig 2), while the 32
receivers provided excellent FLAIR images of the human brain (Fig 3).
Using the identical gradient waves as measured by fieldcameras,
repeated 64 times, no peripheral nerve stimulation was observed by the healthy
volunteer. Discussion
While a fixed resonance above 20kHz is pleasant for
sound perception and simplifies cable management, it coincides with limitations
in flexibilities to setup the MRI experiment. Furthermore, no standard
MR-sequences exist for 5 gradient channels. Nonetheless, the setup demonstrates
similar performance as a state-of-the-art 7T 8ch-32ch head coil, even in the
presence of a close fitting two channel gradient insert. Moreover, as the third
axis is not implemented, full visual view is maintained. Conclusion
We successfully increased the operation frequency of a
high-end gradient amplifier by altering the firmware and bringing the end
filter to the coil. Moreover, a low weight setup was constructed that houses 8
transceivers, 32 receivers, 2 channel cooled gradient coil for operation in a 7T
MRI system for MRI experiments in the human brain, even providing access for
visual stimulation. The operation of the setup was confirmed using field
cameras and MRI images from the human brain, while no PNS was observed when
operating spirals at 6100 and 3040T/m/s in the brainAcknowledgements
No acknowledgement found.References
Versteeg, E. et al. MRM2021