Zhensen Chen1,2, Chuyu Liu3, Yishi Wang4, Weibo Chen5, Rui Guo6, He Wang1,7, and Xiaolei Song3
1Institute of Science and Technology for Brain-Inspired Intelligence, Fudan University, Shanghai, China, 2MOE Frontiers Center for Brain Science, Fudan University, Shanghai, China, 3Department of Biomedical Engineering, School of Medicine Tsinghua University, Beijing, China, 4Philips Healthcare, Beijing, China, 5Philips Healthcare, Shanghai, China, 6Department of Medicine (Cardiovascular Division), Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA, United States, 7Human Phenome Institute, Fudan University, Shanghai, China
Synopsis
To date, chemical exchange saturation
transfer (CEST) is rarely used in abdominal imaging of human body due to its
high sensitivity to motion. In this study, we developed a free-breathing
abdominal CEST sequence that uses a precise respiratory synchronization strategy
and includes a water presaturation module. Preliminary in-vivo experiments
showed that the proposed sequence outperformed non-triggered and breath-holding
abdominal CEST in reducing the slice position inconsistency between the
frequency offsets and motion-induced noise.
Introduction
Chemical exchange saturation transfer (CEST) MRI is a promising molecular
imaging technique, and has been demonstrated useful in diagnosis of various
diseases. To date, CEST is rarely used in abdominal imaging of human body due
to its high sensitivity to motion. A few previous studies have used
breath-holding to perform abdominal CEST imaging, in which one breath-hold was
required for the acquisition of each frequency offset [1-3].
However, there may remain large inconsistency in slice position over the
frequency offsets due to the difficulty in achieving the same breathing level,
and it would cause uncomfortableness to patients and long total acquisition
time when acquiring many frequency offsets. Therefore, this study aimed to
develop a free-breathing sequence for abdominal CEST imaging.Methods
Sequence design: The
diagram of the proposed sequence is shown in Figure 1. Compared to the
conventional CEST sequence, the proposed sequence is unique in two aspects: (1)
Precise respiratory synchronization is implemented to ensure that the readout
is located inside a pre-designated window of expiratory phase [αlb,
αub], where the
expiratory phase α is defined
as the time interval between readout and start of expiration phase (tacq)
divided by the total duration of expiration phase (texp). To achieve
this, respiratory triggering is used and the trigger delay (td_tri)
is adaptively adjusted according to the predicted total durations of
inspiration phase and expiration phase from previous respiratory cycles, such
that the readout is placed at a targeted expiratory phase αtar. In addition, immediately after the trigger
signal of the subsequent respiratory cycle, the real expiratory phase αreal
of the readout is calculated,
and data will be re-acquired if αreal falls outside [αlb, αub]. (2) To remove the influence of
repetition time variation among respiratory cycles, a water presaturation
module that is followed by a delay time (td_wsat)
is performed before the CEST saturation module.
Simulation: To investigate
influence of the water presaturation module on CEST contrast, simulation of the
proposed sequence under different delay times after the water presaturation
module (i.e. td_wsat) was
performed, by using the Bloch-McConnell equation-based simulation code
downloaded from http://www.cestsources.org[4].
In-vivo experiments: For
proof of concept, preliminary in-vivo experiments of the proposed sequence were
performed on two healthy volunteers (males, 23 years and 54 years). Both
volunteers provided written informed consent before participation. The
experiments were performed on a Philips Ingenia CX 3T MR scanner (Philips
Healthcare, Best, The Netherlands), using a 16-channel body coil and a 16-channel
posterior coil as the receivers. For all CEST scans, a CEST saturation duration
(tsat) of 1800 ms and a single-slice single-shot TSE readout (~300
ms, in coronal direction) were used. For the proposed respiratory
synchronization strategy, the targeted expiratory phase αtar and the
window [αlb, αub] were set to 0.5 and [0.3, 0.7],
respectively.Results and Discussion
Figure 2 shows the simulated Z-spectra with water presaturation, as
well as the conventional one without water presaturation. As can be seen, the Z-spectra
with water presaturation are higher than the conventional one at all frequency offsets,
and the CEST peak in MTR asymmetry is slightly diminished. This influence of
water presaturation probably can be corrected by using the recently proposed
quasi-steady CEST analysis approach [5],
which will be investigated in the near future.
A demonstration of the proposed respiratory synchronization strategy
for free-breathing abdominal CEST imaging on volunteer #1 is presented with
animation in Figure 3. Compared to the non-triggered one, respiratory
triggering did largely reduce the motion-induced slice location inconsistency
between the frames, and the proposed precise respiratory synchronization
strategy (i.e. the one with strict data check) seems to outperform the
conventional one (i.e. the one with simple data check).
A comparison of CEST scans using breath-holding and the proposed
respiratory synchronization strategy (i.e. triggered, with strict data check) on
volunteer #1 is presented with animation in Figure 4. The proposed strategy is
apparently much better than breath-holding in reducing the total acquisition
time as well as slice location inconsistency, although there remains one frame
with large position shift.
Figure 5 shows the comparison of B0 map, Z-spectrum and MTR asymmetry
map between the hepatic CEST scans with non-triggered and the proposed
respiratory synchronization strategy on volunteer #2. The proposed method resulted
in a more homogeneous MTR asymmetry map with less artifacts on the boundary
(white arrows) and fewer noisy oscillations (red arrows) on Z-spectrum.Conclusion
The proposed free-breathing abdominal CEST sequence with water
presaturation and precise respiratory synchronization is feasible and can
result in shorter acquisition time and less motion artifact than breath-holding
CEST. More solid validations and comparisons for the proposed sequence are on-going. Acknowledgements
This work was
supported by the Research Start-up Funds from Fudan University to Dr. Zhensen
Chen, National Natural Science Foundation of China (82071914), and the startup
package from Tsinghua University to Dr. Xiaolei Song.References
- Chen,
S.Z., et al., Chemical exchange
saturation transfer (CEST) MR technique for in-vivo liver imaging at 3.0 tesla.
Eur Radiol, 2016. 26(6): p.
1792-800.
- Deng, M., et al., Chemical Exchange Saturation Transfer (CEST) MR Technique for Liver
Imaging at 3.0 Tesla: an Evaluation of Different Offset Number and an
After-Meal and Over-Night-Fast Comparison. Mol Imaging Biol, 2016. 18(2): p. 274-82.
- Tang, Y., et al., Noninvasive Detection of Extracellular pH in Human Benign and Malignant
Liver Tumors Using CEST MRI. Front Oncol, 2020. 10: p. 578985.
- Zaiss, M. CEST sources. 2014 [cited
2021 March 1]; Available from: http://www.cest-sources.org.
- Sun, P.Z., Quasi-steady state chemical exchange saturation transfer (QUASS CEST) analysis-correction
of the finite relaxation delay and saturation time for robust CEST measurement.
Magn Reson Med, 2021. 85(6): p.
3281-3289.