Tao Zu1, Yi-Cheng Hsu2, Yi Sun2, Dan Wu1,3,4, and Yi Zhang1,3,4
1Key Laboratory for Biomedical Engineering of Ministry of Education, Department of Biomedical Engineering, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China, 2MR Collaboration, Siemens Healthcare Ltd., Shanghai, China, 3Department of Neurology, The First Affiliated Hospital, Zhejiang University, Hangzhou, China, 4Cancer Center, Zhejiang University, Hangzhou, China
Synopsis
The relaxation time mapping has proven to be an important
diagnostic tool, but it is limited by the prolonged scan time due to the measurements
of multiple frames at the same location. In this study, the recently proposed auto-calibrated
reconstruction method by joint k-space and image-space parallel imaging (KIPI) is
utilized for the acceleration of relaxation time mapping. Combined with the ESPIRiT
method, KIPI generates improved coil sensitivity maps and allows an
acceleration factor of up to 4-fold for acquiring source images, yielding the
accurate parameter map without obvious errors or artifacts.
Introduction
The relaxation time mapping,
which can characterize intrinsic tissue-dependent information, has proven to be
a valuable diagnostic tool(1-3). However, the
long scan times imposed by conventional methods limit its routine clinical application
since multiple images with modulated sequence parameters are required(4). Recently, an
auto-calibrated reconstruction method by joint k-space and image-space parallel
imaging (KIPI) was proposed to accelerate the CEST acquisition(5). Here, we
modify the KIPI method by combining the ESPIRiT(6) approach to
accelerate the relaxation time mapping.Theory
In relaxation time mapping, the dataset comprises a
series of images acquired with different pulse sequence parameters from the
same volume(4), which
implies that the coil sensitivity maps of different
frames are essentially the same. The conventional SENSE(7) method uses
sensitivity maps explicitly from a separate reference scan, while the GRAPPA(8) method utilizes
sensitivity encoding implicitly via correlations in k-space estimated from
autocalibration signals (ACS). However, it is difficult to obtain accurate
sensitivity maps for SENSE reconstruction, limiting its maximum achievable
acceleration factor in practice. As for GRAPAA, the ACS data needs to be repetitively
acquired for all relaxation time mapping frames, causing a substantial overhead
scan time.
The recently-proposed KIPI(5) method
utilizes variable acceleration factors for different CEST frames and
incorporates the advantages of both SENSE and GRAPPA. KIPI requires a CEST
frame to be sampled at a low acceleration factor (e.g. AF=2), uses GRAPPA to
fill the unsampled k-space, obtains sensitivity maps from GRAPPA-generated full
k-space, and then reconstructs the other highly-undersampled (AF>2) frames
with SENSE and the artifact suppression algorithm(9). However,
when applying KIPI to inversion recovery (IR) relaxation time mapping
sequences, the sensitivity maps obtained by dividing the sum-of-squares (SOS)
image into the image from each channel are not always accurate, especially at
the scalp area. In this work, the optimized sensitivity maps are obtained by
combining the raw and ESPIRiT(6) sensitivity
maps, with the reconstruction flowchart shown in Fig. 1. Methods
Human and phantom experiments were conducted on a 3T
Siemens Prisma MRI system with a multi-channel receiver. An
IR TSE sequence was running for T1 mapping with full k-space
sampling, and time of inversion (TI) = 50, 150, 300, 500, 800, 1300, and 2000ms.
For retrospective undersampling, the TI=2000ms frame was selected as the
calibration frame with AF=2 and 24 ACS lines for KIPI to generate the
sensitivity maps and correction maps. And the remaining six IR frames were
undersampled with AF=4 and without ACS data.
For KIPI reconstruction, the k-space
of the AF=2 calibration frame was firstly refilled by GRAPPA, from which the
raw sensitivity maps were calculated by dividing the SOS image into each
channel image. The ESPIRiT sensitivity maps were derived from the ACS data of
the calibration frame. The values in the noise region of the raw sensitivity
map were replaced by those of the ESPIRiT sensitivity, which was defined as
less than 10% of the maximum signal intensity. Then a locally-weighted
polynomial regression(7) was
implemented to fit the combined sensitivities to get the optimized sensitivity
maps (Fig. 2). Besides, the phase of the sensitivity map was generated individually
due to the gap between raw and ESPIRiT maps. With the optimized sensitivity
maps, KIPI reconstructed the other AF=4 frames by SENSE with artifact suppression(5,9).
For comparison, the regular
GRAPPA and ESPIRiT reconstruction were implemented which used the same data as
the KIPI method. Moreover, KIPI was executed with three sensitivity maps as
shown in Fig. 2 while other conditions were kept the same.Results
Figure 2 shows three kinds of sensitivity maps from a healthy volunteer in one coil channel. The raw sensitivity map has a noisy edge and unsmooth jump(red arrow), while the ESPIRiT one is very smooth but a little blurry. The optimized sensitivity map retains the details and avoids the interference of noise, and its phase map has a high agreement with the raw one.
Figure 3 displays the
reconstructed phantom images and the corresponding T1 maps from undersampled IR TSE data. For
all three schemes of sensitivity maps, KIPI obtains results with high quality,
while GRAPPA and ESPiRiT yield obvious artifacts in both source images (red
arrows) and T1 maps (yellow
arrows).
Figure 4 shows the TI=50ms frame reconstructed by GRAPPA, ESPIRiT, and KIPI. The results of GRAPPA and ESPIRiT have evident unfolding artifacts, indicated by the red arrows, which are absent in KIPI results. Figure 5 displays the T1 maps calculated from the source images of Figure 4. Noticeable artifacts can be seen in the results reconstructed by GRAPPA and ESPIRiT while the KIPI method generates T1 maps that are more consistent with the ground truth. The KIPI reconstruction with optimized sensitivity maps shows the best results among all three kinds of sensitivity maps.Conclusion
KIPI is a novel and
non-iterative auto-calibrated reconstruction method that integrates the
strength of both k-space and image-space parallel imaging. Combining the
ESPIRiT method, KIPI generates more accurate and robust sensitivity maps. KIPI
allowed an acceleration factor of up to 4-fold for 2D relaxation time mapping, essentially
without compromising the image quality. The advantages of KIPI may be exploited to
accelerate other MR parameter mapping that acquires multiple image frames at
the same location.Acknowledgements
NSFC
grant numbers: 61801421 and 81971605. Leading Innovation and Entrepreneurship
Team of Zhejiang Province: 2020R01003. This work was supported by the MOE
Frontier Science Center for Brain Science & Brain-Machine Integration,
Zhejiang University.References
1. Bottomley
PA, Hardy CJ, Argersinger RE, Allenmoore G. A REVIEW OF H-1
NUCLEAR-MAGNETIC-RESONANCE RELAXATION IN PATHOLOGY - ARE T1 AND T2 DIAGNOSTIC.
Medical Physics 1987;14(1):1-37.
2. Mackay
A, Whittall K, Adler J, Li D, Paty D, Graeb D. IN-VIVO VISUALIZATION OF MYELIN
WATER IN BRAIN BY MAGNETIC-RESONANCE. Magnetic Resonance in Medicine
1994;31(6):673-677.
3. Sibley
CT, Noureldin RA, Gai N, Nacif MS, Liu S, Turkbey EB, Mudd JO, van der Geest
RJ, Lima JAC, Halushka MK, Bluemke DA. T1 Mapping in Cardiomyopathy at Cardiac
MR: Comparison with Endomyocardial Biopsy. Radiology 2012;265(3):724-732.
4. Crawley
AP, Henkelman RM. A COMPARISON OF ONE-SHOT AND RECOVERY METHODS IN T1-IMAGING.
Magnetic Resonance in Medicine 1988;7(1):23-34.
5. Zu
T, Sun Y, Wu D, Zhang Y. Accelerated Chemical Exchange Saturation Transfer
Acquisition by Joint K-space and Image-space Parallel Imaging (KIPI).
Proceedings of the 29th Annual Metting of ISMRM 2021:2598.
6. Uecker
M, Lai P, Murphy MJ, Virtue P, Elad M, Pauly JM, Vasanawala SS, Lustig M.
ESPIRiT-an eigenvalue approach to autocalibrating parallel MRI: Where SENSE
meets GRAPPA. Magnetic Resonance in Medicine 2014;71(3):990-1001.
7. Pruessmann
KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: Sensitivity encoding for fast
MRI. Magnetic Resonance in Medicine 1999;42(5):952-962.
8. Griswold
MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A.
Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magnetic
Resonance in Medicine 2002;47(6):1202-1210.
9. Zhang
Y, Heo H-Y, Jiang S, Zhou J, Bottomley PA. Fast 3D chemical exchange saturation
transfer imaging with variably-accelerated sensitivity encoding (vSENSE).
Magnetic Resonance in Medicine 2019;82(6):2046-2061.