Na Liu1, Wei Qing Song1, Jie Liang Lin2, Nan Wang1, Lian ai Liu1, and Wei Yan Miao1
1the First Affiliated Hospital of Dalian Medical University, Da Lian, China, 2Philips Healthcare, BeiJing, China
Synopsis
Multivane (MV) imaging, based on
the blade rotating acquisition in k space, is a novel sequence for motion
reduction imaging. The MV acquisition percentage is critical parameter in MV
imaging for balanced image quality and scan time. This study aims to assess the
performance of compressed sensing for accelerated T2-weighted head MV imaging
under altered MV acquisition percentages. Results indicated that scan time of
head Multivane (MV) T2WI was significantly reduced by compressed sensing (CS)
without reduced image quality.
Synopsis
Multivane (MV) imaging, based on
the blade rotating acquisition in k space, is a novel sequence for motion
reduction imaging. The MV acquisition percentage is critical parameter in MV
imaging for balanced image quality and scan time. This study aims to assess the
performance of compressed sensing for accelerated T2-weighted head MV imaging
under altered MV acquisition percentages. Results indicated that scan time of
head Multivane (MV) T2WI was significantly reduced by compressed sensing (CS)
without reduced image quality.Summary of Main Findings
Scan time for head MV T2WI was
reduced significantly by CS without obvious reduction in image quality. Images
acquired with combination of the CS factor 3 and the MV percentage of 260 showed
adequate performance in both subjective and objective evaluations, thus
recommended for routine MR scans.Introduction
Multivane (MV) is a recently introduced
technique for motion reduction in MRI using rotating blades in the k-space,
which is capable of detection and correction of affects introduced by low
frequency in-plane motion in the center k-space for each blade, thus reduced
motion artifacts in images [1]. MV percentage refers to the total area
of rotating acquisition of blades in this acquisition mode of windmill compared
to the area collected in the Cartesian manner when fully filling the k space[2].
The higher MV percentage tends to be associated with the better image quality,
but also the longer scan time. The compressed sensing (CS) technique has now
been widely used for scan time reduction in MRI through the sparse data sampling,
while ensuring the image quality to meet the diagnostic requirements. The
purpose of this study is to explore the best combination of MV percentage
and CS acceleration factor (AF) for fast and motion-robustness head T2-weighted
imaging (T2WI).Materials and methods
Data
from 12 healthy subjects were included (6 females, 31.25±12.47 years). All
subjects underwent the conventional and MV T2WI on a 3.0 T MR scanner (Ingenia
CX, Philips Healthcare, Best, the Netherlands). MV T2WI adopted a routine
clinical setup using a SENSE acceleration factor 2, and the CS AF of 2, 3 and 4
were used with combination of different MV percentages (160,210,260 or 310). Detailed
parameters were shown in Table 1. Regions of interest were placed manually at
both sides of the Centrum semiovale, tail-shaped nucleus at the maximum level, press part of corpus
callosum, cerebrospinal fluid by an experienced radiologist for the measurement of signal to noise ratio (SNR) and contrast to noise ratio (CNR). The SNR and CNR were repeated
measured for twice with the averaged values for further analyses. Two experienced observers used a
five-point scoring method to subjectively evaluate the quality of images (the
scoring system was listed in Table 2). The Kappa test was adopted to
evaluate the consistency of the scores from the two radiologists. If the
consistency was in good agreement, the corresponding images would be adopted
for further analysis by senior physicians. The Frideman test was used to
assess the difference of SNR, CNR and score among all sequences. This study has
been approved by the local IRB.Results
The
scores by the two observers were in good agreement (Kappa = 0.762). There
was no statistical differences in the subjective scores of each sequence (P>0.05,
Table 3). For sequences with the same CS factor, the SNR and CNR tended to be
increased with the increased of MV percentage but without significantly
difference (P>0.05). For sequence with the same MV percentage, the
SNR and CNR tended to be reduced with the increase of CS factors without
significantly difference (P>0.05). For the sequence with similar scan
times, the combination of CS-AF3 and MV percentage of 260 showed the highest
SNR and CNR (P>0.05, Table 4).Conclusions
Scan time for head MV T2WI could be
reduced significantly by CS without obvious reduction in image quality. Images
acquired with the combination of the CS factor 3 and the MV percentage of 260 showed
adequate performance in both subjective and objective evaluations, thus
recommended for routine MR scans.Acknowledgements
No acknowledgement found.References
1.
Kang,
KA,Kim, YK ,Kim, E, et al. T2-Weighted Liver MRI Using the
MultiVane Technique at 3T: Comparison with Conventional T2-Weighted MRI. Korean
J Radiol.2015 ;16(5) :1038-46
2.
Sodhi,
KS ,Bhatia, A ,Lee, EY. Prospective Evaluation of
Free-Breathing Fast T2-Weighted MultiVane XD Sequence at 3-T MRI for Large
Airway Assessment in Pediatric Patients. AJR Am J Roentgenol.2021
04 ;216(4) :1074-1080