Jihun Kwon1, Takashige Yoshida2, Masami Yoneyama1, and Marc Van Cauteren3
1Philips Japan, Tokyo, Japan, 2Tokyo Metropolitan Police Hospital, Nakano, Japan, 3Philips Healthcare, Best, Netherlands
Synopsis
T2-weighted (T2W) imaging is used to depict
prostate anatomy. However, T2W imaging with high spatial resolution, which is
required by PI-RADSv2, is not always performed because of the long scan time.
In this study, we investigated the use of Compressed SENSE AI (CS-AI)
reconstruction to accelerate high-resolution T2W imaging of the prostate. The
image quality of CS-AI reconstruction, conventional Compressed-SENSE (C-SENSE),
and SENSE were compared qualitatively and quantitatively. Clinical cases
demonstrated a significant noise reduction in accelerated CS-AI. CS-AI also showed
a significantly higher contrast-to-noise ratio in CS-AI compared to SENSE and
C-SENSE.
Introduction
MRI plays an important role in the
diagnosis and staging of prostate cancer patients. In particular,
high-resolution T2-weighted (T2W) imaging is used to depict prostate anatomy
and assess extraprostatic extension (EPE)1. To standardize imaging protocols and reduce variability in
multiparametric prostate MRI (mpMRI), the minimum technical standard for
prostate MR imaging was established by Prostate Imaging Reporting and Data
Systems (PI-RADS)2. However, recent reports indicated that there are variations among
facilities in their adherence to the minimum technical standards for PI-RADSv23–6. Among acquisition parameters, adherence to frequency resolution (≤
0.4 mm) and phase resolution (≤ 0.7 mm) in T2W turbo spin echo (TSE) imaging
were particularly low3,6. This is because comprehensive mpMRI protocol with high-resolution
imaging prolongs examination time to such an extent that it is not feasible in
clinical practice. Although many scan acceleration techniques, such as parallel
imaging7 and compressed sensing8 have been developed, the use of high acceleration factors may result
in too much noise breakthrough.
In this study, Compressed SENSE AI (CS-AI)
reconstruction9,10 was used to acquire high-resolution T2W images while accelerating
the scan compared to parallel imaging (SENSE). It is hypothesized that the
image quality of high-resolution T2W imaging can be significantly improved
while accelerating the scan by using the CS-AI reconstruction algorithm. The
purpose of this study was to acquire accelerated high-resolution T2W images for
the prostate using the CS-AI reconstruction and quantitatively compare the
image quality with SENSE and conventional Compressed-SENSE (C-SENSE).Methods
The study was approved by the local IRB,
and written informed consent was obtained from all subjects. A total of 14 patients
were examined on a 3.0T whole-body clinical system (Ingenia Elition X, Philips
Healthcare) using a 32-channel phased-array coil. Multi-slice 2D T2W TSE
sequence was acquired in the axial plane.
The image quality was compared between
three acceleration methods: SENSE, C-SENSE, and CS-AI. SENSE images were
acquired with acceleration factor 2 and the scan time was 3:04 min. C-SENSE and
CS-AI images were acquired with acceleration factor 3 and the scan time was
2:15 min. Following parameters were common to all examinations: FOV=200x200 mm2,
slice thickness=3 mm, no intersection gap, 20 slices, in-plane acquisition
resolution 0.4×0.6 mm2, in-plane reconstruction resolution 0.2×0.2
mm2, TR/TE=7386/98ms, number of signals averaged (NSA)=1.
The image quality was quantitatively evaluated
by calculating the contrast-to-noise ratio (CNR). First, oval regions of
interest (ROI) were placed on SENSE image in three areas; transitional zone
(TZ), outer peripheral zone (PZ), and obturator internus muscle as a proxy for background
(bg). The ROIs were then copied to C-SENSE and CS-AI images. The CNR was
defined as:
$$CNR={ROI_{organ} - ROI_{bg} \above{1pt}\sqrt{\frac 1 2\times(SD_{organ}^2+SD_{bg}^2)}}$$
where ROIorgan and ROIbg are
the averages of signal intensity measured in the prostate (i.e. TZ or
PZ) and background ROIs, respectively, and SDorgan and SDbg are the
standard deviations in the prostate and background ROIs, respectively. Paired
t-tests were performed to determine statistical significance between scan
acceleration techniques. P ≤ 0.05 was considered statistically significant.
The CS-AI model used in this study is the
extension of the previously introduced AI-based reconstruction algorithm,
Adaptive-CS-Net9,10. In CS-AI, the iterative optimization procedure in the C-SENSE
reconstruction chain is unrolled for a fixed number of reconstruction blocks.
Each block consists mainly of Unet-like architecture, which performs as a
denoiser. The model was trained on more than 700,000 images, including 2D and
3D data, and multiple contrasts and anatomical areas.Results and Discussions
Figures 1 to 4 show comparison of SENSE
(acceleration factor 2.0), C-SENSE (acceleration factor 3.0), and CS-AI
(acceleration factor 3.0) for axial high-resolution T2W images. Overall, SENSE
and C-SENSE images became very noisy and were not suitable for diagnosis.
However, the CS-AI significantly reduced the noise and produced visually
sharper images, which led to the better depiction of small structures. The
contrast inside the TZ and the PZ improved visually on CS-AI.
Figure 5 (a) and (b) show the CNR comparison
between SENSE, C-SENSE, and CS-AI, measured on TZ and PZ, respectively. The
average CNR were 3.89, 4.20, and 5.26 in the TZ, and were 6.93, 7.37, and 8.47
in the PZ for SENSE, C-SENSE, and CS-AI, respectively. The differences of the
average CNR between three sequences were all statistically significant for both
TZ (SENSE vs C-SENSE: p=0.003, SENSE vs C-SENSE: p<0.001, and C-SENSE vs
CS-AI: p<0.001) and PZ (SENSE vs C-SENSE: p=0.004, SENSE vs C-SENSE: p<0.001,
and C-SENSE vs CS-AI: p<0.001). It has to be noted that the CS-AI
demonstrated superior image quality compared to SENSE despite having a higher
acceleration factor, i.e. k-space undersampling rate. Scan time
reduction in CS-AI compared to SENSE may have suppressed possible motion
artifacts and improved the image quality. However, quantitative evaluation also
showed CS-AI to have better image quality than C-SENSE, which was acquired at
the same scan duration as CS-AI. These results suggest that the CS-AI can
acquire high-resolution T2W images while significantly accelerating the scan
and improving the image quality.Conclusion
High-resolution T2W images for the prostate
were acquired with SENSE and accelerated C-SENSE and CS-AI reconstruction.
Clinical examples demonstrated significant noise reduction in accelerated
CS-AI. The quantitative evaluation showed substantially higher CNR in CS-AI compared to SENSE
and C-SENSE.Acknowledgements
No acknowledgement found.References
1. Israël B, Leest M van der, Sedelaar M, Padhani AR, Zámecnik P, Barentsz JO. Multiparametric Magnetic Resonance Imaging for the Detection of Clinically Significant Prostate Cancer: What Urologists Need to Know. Part2: Interpretation. Eur Urol. 2020;77(4):469-480. doi:10.1016/j.eururo.2019.10.024
2. Weinreb JC, Barentsz JO, Choyke PL, et
al. PI-RADS Prostate Imaging – Reporting and Data System: 2015, Version 2. Eur
Urol. 2016;69(1):16-40. doi:10.1016/j.eururo.2015.08.052
3. Esses SJ, Taneja SS, Rosenkrantz AB.
Imaging Facilities’ Adherence to PI-RADS v2 Minimum Technical Standards for the
Performance of Prostate MRI. Acad Radiol. 2018;25(2):188-195.
doi:10.1016/j.acra.2017.08.013
4. Cuocolo R, Stanzione A, Ponsiglione A,
et al. Prostate MRI technical parameters standardization: A systematic review
on adherence to PI-RADSv2 acquisition protocol. Eur J Radiol.
2019;120(August):108662. doi:10.1016/j.ejrad.2019.108662
5. Burn PR, Freeman SJ, Andreou A,
Burns-Cox N, Persad R, Barrett T. A multicentre assessment of prostate MRI
quality and compliance with UK and international standards. Clin Radiol.
2019;74(11):894.e19-894.e25. doi:10.1016/j.crad.2019.03.026
6. Sackett J, Shih JH, Reese SE, et al.
Quality of Prostate MRI: Is the PI-RADS Standard Sufficient? Acad Radiol.
2021;28(2):199-207. doi:10.1016/j.acra.2020.01.031
7. Pruessmann KP, Weiger M, Scheidegger MB,
Boesiger P. SENSE: Sensitivity encoding for fast MRI. Magn Reson Med.
1999;42(5):952-962.
doi:10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S
8. Lustig M, Donoho D, Pauly JM. Sparse
MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58(6):1182-1195.
doi:10.1002/mrm.21391
9. Pezzotti N, de Weerdt E, Yousefi S, et
al. Adaptive-CS-Net: FastMRI with Adaptive Intelligence. arxiv.
2019;(NeurIPS). http://arxiv.org/abs/1912.12259
10. Pezzotti N, Yousefi S, Elmahdy MS, et
al. An Adaptive Intelligence Algorithm for Undersampled Knee MRI
Reconstruction. IEEE Access. 2020;8:204825-204838.
doi:10.1109/ACCESS.2020.3034287