Abdullah S. Bdaiwi1,2, Mariah L. Costa1,2, Joseph W. Plummer1,2, Matthew M. Willmering1, Laura L. Walkup1,2,3,4, and Zackary L. Cleveland1,2,3,4
1Center for Pulmonary Imaging Research, Cincinnati Children’s Hospital Medical Center, Cincinnati, OH, United States, 2Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH, United States, 3Department of Radiology, Cincinnati Children’s Hospital Medical Center, Cincinnati, OH, United States, 4Department of Pediatrics, University of Cincinnati, Cincinnati, OH, United States
Synopsis
Hyperpolarized 129Xe MRI can non-invasively measure regional
ventilation by mapping the spin-density of inhaled gas, thus providing insights
into regional disease pathophysiology. However, the quantitative accuracy of 129Xe
ventilation imaging is reduced by B1-inhomogeneity causing spatial variations
in both coil sensitivity and nonequilibrium magnetization decay. These can
cause lung function impairment to be either over or underestimated. We demonstrate
these artifacts can be mitigated by generating flip angle maps from paired
2D-spiral images acquired in the same held breath. This yields quantitatively comparable
results to those obtained with a conventional, 2D gradient recalled echo
sequence in substantially reduced acquisition times.
Introduction
The highly non-equilibrium magnetization afforded
by hyperpolarized (HP) 129Xe MRI enables non-invasive strategies to
quantify lung function and structure1-5. In particular, mapping the spin-density of the inhaled gas enabled regional ventilation
to be imaged, thus providing insights into regional disease
pathophysiology. However, each RF pulse
used to acquire the image (n,…,N) consumes longitudinal magnetization.
Therefore, selecting an optimal flip angle (FA) is crucial to obtaining high image
quality. However, even when the global
FA is optimal, B1-inhomogeneity causes the local FA to deviate from proscribed
value across the imaged volume. Moreover, these artifacts vary between patients,
due to differences in coil fit and loading, and can mimic disease, compromising
image interpretation. Fortunately, these systematic deviations can be corrected
if the local FA is known. Previously, FA maps were obtained by acquiring two consecutive images at the same constant FA
(θC) during a breath-hold using a 2D GRE sequence6. This method requires long breath holds (~10-15 s), which can be
challenging for pediatric
and severely ill patients, or limited lung
coverage. However, spiral acquisitions acquire k-space data more efficiently,
reducing overall scan duration (<5s)7-10. In this study, we describe and test a method to acquire sensitivity
maps and ventilation images during a single breath-hold scan and compare the resulting
correction maps to those obtained using 2D-GRE. These maps can be used to
detect and compensate for intensity variation artifacts caused by B1-inhomogeneity.Theory and Methods
An analytical model
for FA uncertainty (relative error, ϵθc) was derived using
error propagation (Figure 1A) and evaluated using MATLAB (MathWorks, Natick, MA) across
a range of parameters for 2D-GRE and 2D-spiral (number of phase encoding, Nph=25−150, number
of spirals, NS=10−35 and θc=1o−40o).
To validate this model, 2D simulations for GRE and
spiral were performed by modeling signal decay according to
(Figure 2). For HP 129Xe imaging, both
2D-GRE and 2D-spiral sequences were designed such that two consecutive images (fully sampled) were acquired with fixed
and slice position during
a breath-hold and T1 could be assumed to be long relative to slice acquisition. To
validate this approach, structured and non-structured phantoms containing 1L of
HP-129Xe were imaged at 3T (Philips
Ingenia,
Best, Netherlands). Finally, HP-129Xe ventilation
imaging was performed in a healthy volunteer (male, 30 years) and a patient
with Cystic Fibrosis (CF, female, 21 years). Xenon was polarized to ~30%
(Polarean 9820, Inc, Durham, NC). MRI parameters included: resolution 3x3x15mm2,
12 slices, TR/TE [GRE=6.16/1.7; spiral=14/0.92ms], FOV [GRE=340×240; spiral=340×340mm2],
scan time [GRE~15; spiral~5s] and acquisitions window [GRE=0.3; spiral=10ms]. Images
were reconstructed using Graphical Programming Interface (GPI)11 with further
analysis performed in MATLAB. The two consecutive images were used to generate FA
maps and
according to the equations in Figure 1A. FA maps
were used to correct first image according to
Image1(GRE)×(1/(θc·cosNph/2(θc)·sin(θc))) and Image1(Spiral)×(NS/(θc·sin(θc)))×(1-cos(θc))/(1-cosNS(θc))Results
The analytical solution for FA uncertainty
showed that there exists a FA that maximizes SNR and minimizes ϵθc
for a given Nph for GRE (Figure 1B-C).
A similar trend was observed for spiral imaging, where SNR is maximized and
ϵθc
is minimized at optimal FA for given NS
(Figure 1D-E). Mean
measured FA agreed well with applied FA in simulations (R2> 0.95;
Figure 3), phantoms (Figure 4) and in vivo imaging (Figure
5) (underestimated
by <15%). Furthermore, regions with increased signal decay (white arrows)
showed an increase in ϵθc, relative to the regions with proscribed FA values. The
reconstructed images were corrected using the measured FA maps, showing more
accurate distribution of xenon signal relative to uncorrected images. Discussion and Conclusion
An analytical model was derived to examine the impact of
acquisition parameters (e.g., N, FA) on the uncertainty in the measured FA. The
analytical model accounts for HP-specific effects but does not account for the
uncertainty resulting from image acquisition and reconstruction. To address
this limitation, we simulated HP 129Xe ventilation imaging for GRE
and spiral. These simulations produced broadly similar results to those
predicted by the model with only modestly elevated errors (Figure 3H and P). HP
gas phantoms further validated the approach for measuring RF-coil sensitivity
maps. The measured FA maps correlated well with the applied FA for both GRE and
spiral sequences.
Ventilation
images and a voxel‐level FA map were successfully obtained during the same breath hold
using spiral (~5s) from both healthy and CF subjects. Mean measured FA maps
agreed well with the applied FAs. Generated FA maps were then used to correct
images for signal decay. The corrected images provide a superior estimate of
regional 129Xe content. Notable improvements were observed in the
appearance of the corrected images from both healthy and CF subjects. More
importantly, spiral FA and correction maps produced nearly identical maps to
those obtained using conventional GRE. However, spiral provided higher SNRs and
three-fold faster acquisition times compared to GRE for the same resolution and
number of slices. This rapid acquisition makes spiral more practical for 129Xe
ventilation imaging, especially for pediatric subjects or patients with highly compromised
respiratory function. Acknowledgements
The authors acknowledge the following sources
for research funding and support: Cincinnati Children’s Research Foundation and
NIH (R01HL131012, R01HL143011, and R00HL138255). References
1. Walkup
LL, Woods JC. Translational Applications of Hyperpolarized 3he and 129Xe. NMR in Biomedicine 2014;27:1429-1438.
2. Goodson
BM. Nuclear Magnetic Resonance of Laser-Polarized Noble Gases in Molecules,
Materials, and Organisms. Journal of
Magnetic Resonance Imaging: An Official Journal of the International Society
for Magnetic Resonance in Medicine 2002;155:157-216.
3. Ruppert
K. Biomedical Imaging with Hyperpolarized Noble Gases. Reports on Progress in Physics 2014;77:116701.
4. Mugler
III JP, Altes TA. Hyperpolarized 129Xe MRI of the Human Lung. Journal of Magnetic Resonance Imaging
2013;37:313-331.
5. Bdaiwi
AS, Niedbalski PJ, Hossain MM, et al. Improving Hyperpolarized 129Xe ADC Mapping in Pediatric and Adult Lungs with Uncertainty Propagation. NMR Biomed 2021:e4639.
6. Miller
G, Altes T, Brookeman J, De Lange E, Mugler Iii J. Hyperpolarized 3He Lung
Ventilation Imaging with B1-Inhomogeneity Correction in a Single Breath-Hold
Scan. Magnetic Resonance Materials in
Physics, Biology Medicine 2004;16:218-226.
7. Salerno
M, Altes TA, Brookeman JR, De Lange EE, Mugler III JP. Rapid Hyperpolarized 3He
Diffusion MRI of Healthy and Emphysematous Human Lungs Using an Optimized
Interleaved‐Spiral Pulse Sequence. Journal of Magnetic Resonance Imaging: An
Official Journal of the International Society for Magnetic Resonance in
Medicine 2003;17:581-588.
8. Willmering
MM, Niedbalski PJ, Wang H, Walkup LL, Robison RK, Pipe JG, Cleveland ZI, Woods
JC. Improved Pulmonary (129) Xe Ventilation Imaging Via 3d-Spiral Ute Mri. Magn Reson Med 2020;84:312-320.
9. Brandon
Zanette YF, Samal Munidasa , and Giles
Santy Comparison of 3d Stack-of-Spirals and 2d Gradient Echo for Ventilation
Mapping Using Hyperpolarized 129xe. ISMRM2020.
10. Doganay
O, Matin TN, Mcintyre A, Burns B, Schulte RF, Gleeson FV, Bulte D. Fast Dynamic
Ventilation MRI of Hyperpolarized 129xe Using Spiral Imaging. Magn Reson Med 2018;79:2597-2606.
11. Zwart NR, Pipe JG. Graphical
Programming Interface: A Development Environment for MRI Methods. Magn Reson Med 2015;74:1449-1460.