Nicholas Rubert1, Quin Lu2, Gaurav Jategaonkar1, and Luis Goncalves1,3
1Radiology, Phoenix Children's Hospital, Phoenix, AZ, United States, 2Philips, Best, Netherlands, 3University of Arizona, Phoenix, AZ, United States
Synopsis
We examined undersampled real-time bSSFP reconstructions for fetal cardiac imaging. Two compressed sensing reconstructions, ICTGV and L plus S, were compared to k-t SENSE with respect to spatial and temporal blurring with and without fetal motion. Results with a digital fetal cardiac MRI phantom demonstrated ICTGV was able to capture expansion and contraction of the fetal heart even during periods of fetal motion while k-t SENSE and L plus S were not. Digital phantom results were confirmed on a 1.5T MRI scanner with implementation of ICTGV and acquisitions demonstrating pulsation of the fetal heart during periods of fetal motion.
INTRODUCTION
Random fetal motion, large maternal
anatomy, and small fetal cardiac anatomy place high demands on spatial and
temporal resolution of fetal imaging. Highly undersampled 2D Cartesian real-time imaging
with balanced steady-state free precession (bSSFP) sequences has been used to
capture the dynamics of the fetal heart1,2. Further, slice-to-volume (SVR) super-resolution reconstruction has been shown to effectively correct for motion between
frames of a real-time sequence or between acquisitions2,3. However, SVR may not recover
individual image frames if they are corrupted by artifact. In this work we examined robustness to image
artifacts from fetal motion for three highly undersampled real-time Cartesian reconstruction schemes: k-t SENSE4, L plus S5, and ICTGV6. k-t SENSE relies on a regular
undersampling pattern4, while L plus S and ICTGV utilized a quasi-random
undersampling pattern7. Simulation of fetal motion and image acquisition was
performed with a digital fetal cardiac MR (CMR) phantom for image quality comparison8. Simulation results were confirmed by implementing
compressed sensing reconstructions on a 1.5T Philips scanner and imaging a
normal fetal heart for a volunteer patient.
METHODS
Two-dimensional bSSFP realt-time image formation in transverse, sagittal, and coronal orientation with a
24-element set of coils was simulated under two different motion conditions: 1)
A stationary fetus. 2) Fetal motion due to random movements plus displacement
from maternal breathing. A fetal heart rate of 140 beats per minute was
simulated for both motion conditions. Spatial resolution ranged from 1.88 to
2.17 mm in-plane with a 6.0 mm slice thickness. Two-dimensional k-space was
sampled according to an eight-fold regular undersampling pattern for k-t SENSE
and separate training and high-resolution acquisitions were simulated. Twenty
lines of k-space per image frame were simulated for the acquisition stage and an
additional 10 lines of center k-space were acquired for the training stage.
Simulated k-space was also sampled according to a VISTA pseudo-random
eight-fold undersampling pattern and images were reconstructed with two
compressed sensing algorithms: L + S or ICTGV. Twenty lines of k-space per
image frame were acquired for compressed sensing algorithms. No training data
was necessary for compressed sensing reconstruction . A TR of 3.5 ms was modeled for all acquisitions.
Simulation parameters resulted in a temporal resolution of 70 ms per frame and
a total acquisition time of 4.9 seconds for compressed sensing acquisitions and
7.4 seconds for k-t SENSE. Simulation reconstruction quality was assessed
by measuring the average structural similarity index measure (SSIM) between
ground truth images and reconstructions, considering a 3.2 cm square ROI
containing the fetal heart.
Additionally, VISTA
dynamic scans were implemented on a Philips 1.5T scanner and reconstructed with
ICTGV. 2D bSSFP real-time imaging was performed for a 32-week gestational age
fetus with a normal heart. Acquistion parameters were as follows: TR=3.6 ms,
8-fold undersampling, 65 ms temporal resolution, 6 mm
slice thickness,
4 mm slice spacing, and a scan duration
of 4.2 seconds per slice. Image quality was assessed for its ability to
capture the expansion and contraction of the chambers of the fetal heart. RESULTS
Without motion SSIM ranged
from 0.93 to 0.98 depending on the reconstruction. In the presence of motion,
image quality rapidly degraded for k-t SENSE and L plus S reconstructions but
not ICTGV reconstructions. SSIM ranged from 0.52 to 0.70 for k-t SENSE, 0.72 to
0.81 for L plus S, and 0.90 to 0.94 for ICTGV. In figures 3 and 4 spatial and temporal blurring may be seen in both the k-t SENSE images and the L plus S
images. Meanwhile in the M-mode trace of the ICTGV image in figure 4, the
pulsation of the fetal heart is still visible.
A single frame of an ICTGV
reconstruction and an accompanying M-mode trace is demonstrated for a Philips
1.5T scanner (figure 5). We remark that the pulsation of the heart is evident
with the actual reconstruction. Furthermore, this acquisition happened to occur
during a period of fetal motion. We remark that in the real scan
ICTGV evidently captures the expansion and contraction of the fetal heart even
in the presence of motion.DISCUSSION
Since random fetal motion is always a
possibility during acquisition of fetal images, robustness to motion is a key
feature of any reconstruction algorithm designed for fetal imaging. Little variation in SSIM was observed under ideal motion-free
simulation conditions. However, ICTGV was found to perform significantly better in the presence of
fetal motion. Actual ICTGV images demonstrate that real-time images acquired at
a frame rate of 65 ms can capture the dynamics of the fetal heart, and minor
imperfections in the pulse sequence due to eddy currents do not corrupt the
acquisition. Comparing ICTGV and k-t SENSE acquisitions the ICTGV acquisition had the further advantage that no additional training or
low-resolution data was required. At high undersampling factors where each
frame of k-space is only reconstructed from 20 lines or fewer, avoiding
acquisition of a separate low-resolution scan can significantly shorten imaging
time.CONCLUSION
Due to its robustness to random
motion, ICTGV has great potential for fetal imaging applications. Further work
will combine this acquisition with slice-to-volume registration and
super-resolution reconstruction for anatomic and flow imaging.Acknowledgements
No acknowledgement found.References
1. van Amerom JF, Lloyd DF, Price AN, Kuklisova
Murgasova M, Aljabar P, Malik SJ, Lohezic M, Rutherford MA, Pushparajah K,
Razavi R, Hajnal JV. Fetal cardiac cine imaging using highly accelerated
dynamic MRI with retrospective motion correction and outlier rejection.
Magn Reson Med. 2018 ;79(1):327-38.
2. van Amerom JF, Lloyd DF, Deprez M, Price AN,
Malik SJ, Pushparajah K, van Poppel MP, Rutherford MA, Razavi R, Hajnal JV.
Fetal whole‐heart 4D imaging
using motion‐corrected multi‐planar real‐time
MRI. Magn Reson Med. 2019;82(3):1055-72.
3. Roberts TA, van Amerom
JF, Uus A, Lloyd DF, van Poppel MP, Price AN, Tournier JD, Mohanadass CA,
Jackson LH, Malik SJ, Pushparajah K. Fetal whole heart blood flow imaging using
4D cine MRI. Nat Commun. 2020;11(1):1-3.
4. Tsao J, Boesiger P,
Pruessmann KP. k‐t
BLAST and k‐t SENSE: dynamic
MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med. 2003;50(5):1031-42.
5. Otazo R, Candes E,
Sodickson DK. Low‐rank
plus sparse matrix decomposition for accelerated dynamic MRI with separation of
background and dynamic components. Magn Reson Med. 2015; 73(3):1125-36.
6. Schloegl M, Holler M,
Schwarzl A, Bredies K, Stollberger R. Infimal convolution of total generalized
variation functionals for dynamic MRI. Magn Reson Med. 2017;78(1):142-55.
7. Ahmad R, Xue H, Giri S,
Ding Y, Craft J, Simonetti OP. Variable density incoherent spatiotemporal
acquisition (VISTA) for highly accelerated cardiac MRI. Magn Reson Med 2015;74(5):1266-78.
8. Roy CW, Marini D, Segars
WP, Seed M, Macgowan CK. Fetal XCMR: a numerical phantom for fetal
cardiovascular magnetic resonance imaging. J Cardiovasc Magn Reson. 2019 ;21(1):1-2.