Ruiying Liu1, Jee Hun Kim2, Chaoyi Zhang1, Hongyu Li1, Peizhou Huang1, Xiaojuan Li2, and Leslie Ying1
1Department of Biomedical Engineering, Department of Electrical Engineering, University at Buffalo, Buffalo, NY, United States, 2Program of Advanced Musculoskeletal Imaging (PAMI), Cleveland Clinic, Cleveland, OH, United States
Synopsis
Most
parallel imaging methods require calibration data for reconstruction. Low-rank-based
methods allow calibration-free reconstruction from randomly undersampled,
multi-channel data. This abstract presents a novel reconstruction method to combine
multichannel blind deconvolution (MALBEC), a calibration-free method, and GRAPPA,
a calibration-based method for highly accelerated imaging. The method
sequentially performs MALBEC and GRAPPA with specially designed sampling masks
such that the benefits of low-rank structure and the availability of
calibration data can be utilized jointly. Our results demonstrate that the
proposed method can achieve an acceleration factor that is the product of the
factors achieved by MALBEC and GRAPPA alone.
Introduction
In parallel MRI1, calibration-based methods(e.g., GRAPPA2-3) have been widely used,
which requires fully sampled center k-space for calibration. However, the
acceleration factor is usually limited to two in clinical scans. Some
calibration-free reconstruction methods (e.g., MALBEC4-8) have been proposed to
achieve higher acceleration. These methods require random undersampling in 2D to formulate the reconstruction problem as a low-rank matrix recovery problem
without knowledge of calibration data or coil sensitivities. Despite their potential, the calibration-free methods
alone cannot achieve very high acceleration factors either. In this study, we take advantage of the
benefits of both calibration-free and calibration-based methods and propose a
novel k-space reconstruction method, which integrates MALBEC and GRAPPA to
achieve an
acceleration factor that is the product of the factors achieved individually. Among many calibration-free methods, we choose
MALBEC due to its superior reconstruction quality, ease of parameter selection,
and fast computation. The sampling pattern is specially designed such that MALBEC
and GRAPPA are performed in a sequential manner. The performance of the method
is evaluated using 3D Dual-Echo Steady-State (DESS) knee images. Method
Our proposed method performs MALBEC and GRAPPA
sequentially, as shown in Figure 1. The undersampled k-space data $$$Y_{\Psi,\Omega }$$$ are acquired with a specially designed pattern,
which can be represented as an integration of two undersampling operators $$$\Psi$$$ and $$$\Omega$$$. The first operator $$$\Psi$$$ uniformly undersamples k-space by a factor of $$$\psi$$$ in the phase-encoding direction (GRAPPA-like
undersampling), and the second operator $$$\Omega$$$ randomly undersamples the remaining data by a
factor of $$$\omega$$$ after $$$\Psi$$$ in both phase- and slice-encoding directions. As
a result, the total acceleration reduction is $$$\psi\times\omega$$$. In the first part of reconstruction, the undersampled k-space data $$$Y_{\Omega }$$$ (generated by eliminating the zero-valued phase-encoding
lines in $$$Y_{\Psi,\Omega }$$$ from all
channels) are reconstructed using MALBEC. Specifically, denoting $$$y_{\Omega,c}[p,q]$$$ as the k-space data at location $$$(p,q)$$$ for channel $$$c$$$, each data point can be represented as the convolution of the desired unknown intermediate-reconstructed
k-space data $$$s[m,n]$$$ and the k-space of
the coil sensitivities for the corresponding channels $$$H_{c}[m,n]$$$, $$y_{\Omega,c}[p,q]=\sum_{m=0}^M\sum_{n=0}^NH_{c}[m,n]s[p-m,q-n](1)$$
where $$$c=1,2 \cdots C,(p,q)\in\Omega$$$. MALBEC formulates the reconstruction of $$$s$$$ from the undersampled data as a low-rank
matrix recovery problem4-6 without knowledge of $$$Y_{ m}$$$ or $$$H_{c}$$$. First, we initialize the estimated and $$$H_{c,0}$$$ as $$$s_{0}=F \big(sos\big( F^{-1}\big(T \bullet y_{ \Omega,c}\big)\big)\big), H_{c,0}=F^{-1}\big(T \bullet y_{\Omega,c}\big)/s_{0}(2)$$$. where $$$T$$$ represents a smooth turkey window, $$$sos$$$ represents the sum-of-square operator. The
method then solves for $$$s$$$ and channel responses $$$H_{c}$$$ alternatively based on Eq.(1). Specifically, in
the $$$s$$$-step and $$$H$$$-step, $$$s$$$ and $$$H_{c}$$$ are solved by using: $$s=argmin_{s}\|\sum_{k=1}^K\Omega(s \ast H_{c})\ast\overline{H_{c}}-\sum_{k=1}^K y_{\Omega,c}\ast\overline{H_{c}}\|^{2}(3)$$
$$H_{c}=argmin_{H_{c}}\|\Omega(Hankel(s) \bullet H_{c})-y_{\Omega ,c}\|^{2}(4)$$
We generate the Hankel structured data matrix to
represent the convolution problem. The two
steps are performed iteratively and the iteration usually converges fast (~ 2
iterations). After both $$$s$$$ and $$$H_{c}$$$ are obtained, the multi-coil
intermediate-reconstructed k-space data $$$Y_{m}$$$ can be obtained from their convolution. Adding
previously eliminated zero-valued phase-encoding lines transforms $$$Y_{m} $$$ to $$$Y_{\Psi }$$$, the uniformly
undersampled data. Finally, GRAPPA is
performed on $$$Y_{\Psi }$$$ to obtain the final desired image.
Total 10 in-vivo knees
including 5 knees with metal artifact from ACL reconstruction were scanned on a
3T Siemens scanner with a 15-channel knee coil. A 3D DESS sequence was scanned
with scan parameters as follows; TE =
6.02ms, TR =
17.55ms, and flip angle=20°. FOV= 140 mm, slice
thickness 0.7mm, matrix size of 384x307x160 with
GRAPPA factor of 2 $$$ (\psi =2) $$$. We further manually undersampled the
GRAPPA-undersampled data with 2D random undersampling patterns $$$\Omega$$$ $$$ (\omega =2,3,4)$$$ to simulate combined
acceleration factors of $$$2\times2$$$, $$$2\times3$$$, $$$2\times4$$$.
PSNRs and NMSE were calculated
for reconstructions, using the 2× undersampled data using GRAPPA as the
reference. For knees without metal artifacts, cartilage was automatically
segmented into six
compartments: MFC/LFC, MT/LT, TRO, and PAT cartilages9. Two metrics were used
to compare the reference and accelerated images: DICE coefficients of segmented
cartilage, and the absolute difference of average cartilage thickness of each
compartment10.Results and Discussion
Fig2 shows representative reconstructed
images in sagittal and axial views. For comparison, we replaced MALBEC with
SparseBLIP in our combined method. The
MALBEC+GRAPPA showed superior results compared to SparseBLIP+GRAPPA in terms of
reconstruction quality as well as PSNRs and NMSE shown in each image, especially when metal artifact was present. The performance of MALBEC+GRAPPA
reconstruction does not degrade with metal artifact, as shown in Table1(a). In
knees without metal artifacts, all DICE were higher than 0.9 except for
patellar cartilage with acceleration factor(AF=8), all absolute differences
of cartilage thickness were smaller than 6%, except for patellar cartilage with
AF=8, Table1(b). Not unexpected, DICE
decreases with an increase in AF, and the difference in cartilage thickness
increases with an increase in AF.Conclusion
In this abstract, we developed a novel
acceleration method combining MALBEC and GRAPPA. Experimental
results in 3D knee DESS images show that the proposed method can further
accelerate the acquisition time by a factor of 4 on top of the GRAPPA factor of
2, which reduces the scan time from 6 minutes to 1.5 minutes. More data sets
will be used for evaluating tissue quantification accuracy (cartilage thickness
and composition, with/without metal artifacts) in future studies. Acknowledgements
This work is supported in part by NIH/NIAMS R01 AR077452.References
[1] Ying L, Liang ZP. Parallel
MRI using phased array coils (2010) IEEE
Signal Processing Magazine, 27 (4),
pp. 90-98.
[2] Griswold
MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A.
Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn
Reson Med. 2002 Jun;47(6):1202-10.
[3] Lustig
M, Pauly JM. SPIRiT: Iterative self-consistent parallel imaging reconstruction
from arbitrary k-space. Magn Reson Med. 2010 Aug;64(2):457-71.
[4] Lyu
J, Nakarmi U, Zhou Y, Zhang C, and Ying L. Calibration-free Parallel Imaging
Using Randomly Undersampled Multichannel Blind Deconvolution (MALBEC). Proc. Intl.
Soc. Mag. Reson. Med. 24 (2016)
[5] Shin
PJ, Larson PE, Ohliger MA, Elad M, Pauly JM, Vigneron DB, Lustig M.
Calibrationless parallel imaging reconstruction based on structured low-rank
matrix completion. Magn Reson Med. 2014 Oct;72(4):959-70.
[6] Haldar
JP, Zhuo J. P-LORAKS: Low-rank modeling of local k-space neighborhoods with
parallel imaging data. Magn Reson Med. 2016 Apr;75(4):1499-514.
[7] She H, Chen RR, Liang
D, DiBella EV, Ying L. Sparse BLIP: BLind Iterative Parallel imaging
reconstruction using compressed sensing. Magn Reson Med. 2014 Feb;71(2):645-60.
[8]
Trzasko, JD, Manduca A. Calibrationless parallel MRI using CLEAR. Signals,
Systems and Computers (ASILOMAR), 2011 Conference Record of the Forty Fifth
Asilomar Conference on. IEEE, 2011.
[9] Gaj
S, et al. Automated cartilage and meniscus segmentation of knee MRI with
conditional generative adversarial networks. Magnetic resonance in
medicine 84.1 (2020): 437-449.
[10]
Maier J, Black M, Bonaretti S, Bier B, Eskofier B, Choi JH, Levenston M, Gold
G, Fahrig R, Maier A. Comparison of Different Approaches for Measuring Tibial
Cartilage Thickness. J Integr Bioinform. 2017 Jul 28;14(2):20170015.