Qi Peng1, Gregory Peng1, and Can Wu2
1Department of Radiology, Albert Einstein College of Medicine, Bronx, NY, United States, 2Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
T1ρ
dispersion imaging is an emerging MRI technique for tissue characterization. Multiple
independent repetitions of T1ρ experiments
at different spin-lock frequencies have to be performed to generate tissue T1ρ dispersion curve. In this work, we
demonstrate the feasibility of an integrated imaging and quantification
approach for T1ρ dispersion imaging, which allows
simultaneous generation of T1ρ
maps at multiple spinlock frequencies in one coherent workflow. This opens door
for further data undersampling to exploit redundancy at higher dimension data
space, further reducing total scan time needed for the time-consuming T1ρ dispersion imaging.
Introduction
The
spin-lattice relaxation time in the rotating frame (T1ρ or T1rho,
and the relaxation rate R1ρ = 1/T1ρ), using spin-locking pulse sequences, have been
shown to be sensitive to both local magnetic field fluctuations and chemical
exchange process in the tissue. T1ρ varies as a function of the spin-lock frequency
(FSL), known as T1ρ dispersion, which correlates with tissue
properties.1, 2 Therefore, multiple repetitions of the T1ρ experiments at different FSLs have
to be performed to generate a T1ρ dispersion curve. The MAPSS (Magnetization-Prepared
Angle-Modulated Partitioned k-Space Spoiled Gradient Echo Snapshots) sequence
has been proposed to achieve fast high-resolution 3D T1ρ mapping on the knee cartilage.3 Additionally, it has
been shown that both T2 mapping and T1ρ mapping can be integrated in the same pulse sequence to reduce total scan
duration and simplify post-processing.4, 5 In this work, we demonstrate the feasibility of an
integrated imaging and post-processing approach for T1ρ dispersion imaging, which allows
simultaneous reconstruction of T1ρ maps at multiple spinlock frequencies in one
coherent workflow. This opens door for further data undersampling to exploit
redundancy at higher dimension data space, potentially further reducing total
scan time needed for the time-consuming T1ρ dispersion imaging.Methods
All imaging was performed on a 3T
Philips Ingenia MR scanner. A modified 3D MAPSS sequence was used to acquire T1ρ-weighted images at multiple spin-lock frequencies
for T1ρ dispersion imaging.3 The sequence was on calf muscles of a
46-yo male volunteer with a 1ch-TX/16ch-Rx knee coil
with maximum B1+ of 27 µT. The MAPSS T1ρ preparation module includes a RF train of 90°x-TSL/2-180°y-TSL/2-90°x
to achieve PC- (-Mz) preparation (Figure 1A). The PC+ (+Mz) preparation was
obtained by applying an adiabatic inversion module before the T1ρ preparation module (Figure 1B).6 The MAPSS T1ρ dispersion imaging sequence had 10 FSLs (0 to 900Hz with 100Hz gap),
and with time of spinlock (TSL) of 30ms in addition to the TSL=0 ms
acquisition. Axial 3D volumetric imaging was performed with the following scan
parameters: FOV=140/140/180mm3, acquisition voxel size=1x1x6mm3,
TR/TE=5.4/2.6ms, TFE factor=96 with centric profile ordering and with increasing
RF flip angle sweeping along the GRE readout train, and compressed SENSE factor=4.
MR signal of a voxel with T1ρ at FSL can be modeled as $$$S^{*}(T1\rho(FSL),TSL\pm)=\pm S_{A}e^{-\frac{TSL}{T1\rho(FSL)}}+S_{B},$$$ where SA and SB are complex variable to be determined, and SB
includes the contaminating signal originated from T1 recovery.7 For
the standard T1ρ fitting (termed “Single-fitting” herein), each T1ρ
at different FSL was obtained using paired subtraction between the positive and
negative phase cycling acquisitions followed by a mono-exponential curve-fitting.3 For our proposed T1ρ dispersion fitting method (termed “All-fitting”
herein),
all T1ρ values at different FSLs were obtained simultaneously
using a nonlinear least-square
fitting algorithm using the following optimization function: $$$S_A,S_B,T_1ρ (FSL)=\arg min_{S_A,S_B,T_1ρ (FSL)}{\sum||S(T_1ρ (FSL),TSL±)-S^*(T_1ρ (FSL) ,TSL±)||^{2}},$$$
where S* and S are the modeled and acquired signals,
respectively, from the TSL acquisitions with PC± at FSL. Seven regions-of-interest
(ROIs) were drawn on calf muscles and T1ρ dispersion results were obtained from
both Single-fitting and All-fitting analyses. Student-t test was
performed to evaluate if they have the same mean T1ρ values from the seven ROIs at 10
FSLs. Pearson’s correlation and intraclass correlation coefficient were also calculated
to evaluate the quantitative agreement between these two methods.Results
Representative
T1ρ dispersion imaging results are shown in Figure 2. Figure 2A shows
the seven ROIs defined on a T1ρ-weighted image acquired at TSL=0ms with PC+. Both
the Single-fitting and the All-fitting methods resulted in good quality T1ρ
maps with consistently increasing values at higher FSLs. From visual
inspection, no difference within muscles between the two sets of T1ρ
maps could be identified. The R1ρ results
from the ROI-analyses are shown in Table 1. The R1ρ dispersion curves from the
average R1ρ values of the 7-ROIs are shown in Figure 3, which shows consistent results
from both methods within the whole spectrum of FSLs (from 0 to 900 Hz). From
statistical analysis, there is a strong correlation (r=0.9984, and P<0.0002)
between the R1ρ values from All-fitting and Single-fitting methods. Intraclass correlation
coefficient was 0.9982. Discussion
T1ρ dispersion
imaging can be obtained from multiple independent repetitions of T1ρ mapping at
different FSLs. Given that fact that T1ρ is typically a monotonical, slowly
increasing function of FSL, there is a high degree of redundancy along the FSL
dimension in T1ρ dispersion
imaging. In this pilot study, we demonstrated the feasibility of an integrated dispersion
mapping approach from both the imaging acquisition and the T1ρ map reconstruction
perspectives. Not only did this approach lead to similar results as traditional
methods, but also it could be combined with the compressed sensing acquisition into
the optimization reconstruction framework at a 5D (3D+TSL+FSL) space. This work
opens the door for further data undersampling to dramatically reduce total scan
time needed for the time-consuming T1ρ dispersion imaging, particularly true
when combining with iterative or AI-based reconstruction algorithms.8, 9 Further work is warranted to validate this promising integrated
imaging and reconstruction approach. Acknowledgements
No acknowledgement found.References
1. Koskinen
SK, Virta AM, Niemi PT, et al. T1rho dispersion of rat tissues in vitro. Magn
Reson Imaging. 1999;17(7):1043-1047.
2. Wang P, Block J, Gore JC. Chemical
exchange in knee cartilage assessed by R1rho (1/T1rho) dispersion at 3T. Magn
Reson Imaging. 2015;33(1):38-42.
3. Li X, Han ET, Busse RF, et al. In
vivo T(1rho) mapping in cartilage using 3D magnetization-prepared
angle-modulated partitioned k-space spoiled gradient echo snapshots (3D MAPSS).
Magn Reson Med. 2008;59(2):298-307.
4. Li X, Wyatt C, Rivoire J, et al.
Simultaneous acquisition of T1rho and T2 quantification in knee cartilage: repeatability
and diurnal variation. J Magn Reson Imaging. 2014;39(5):1287-1293.
5. Peng Q, Wu C. High Spatial
Resolution Simultaneous Quantitative T1rho and T2 Mapping at 3T. AAPM Annual
Meeting. 2021.
6. Zhang H, Jiang B, Jian H, et al.
Robust Quantitative T1rho imaging Proc Intl Soc Mag Reson Med 29. 2021.
7. Peng Q, Wu C, Kim J, et al.
Efficient Phase Cycling Strategy for High Resolution 3D GRE Quantitative MRI
Mapping. NMR in Biomedicine. 2021;Under Review.
8. Cao T, Ma S, Wang N, et al.
Three-dimensional simultaneous brain mapping of T1, T2, T 2 * and magnetic
susceptibility with MR Multitasking. Magn Reson Med. 2021.
9. Wang
N, Gaddam S, Wang L, et al. Six-dimensional quantitative DCE MR Multitasking of
the entire abdomen: Method and application to pancreatic ductal adenocarcinoma.
Magn Reson Med. 2020;84(2):928-948.