Jie Huang1, Yuancheng Liu2, Xianchun Zeng2, Chunqi Qian1, and Erik M. Shapiro1
1Michigan State University, East Lansing, MI, United States, 2Guizhou Provincial People's Hospital, Guiyang, China
Synopsis
The
hepatic artery (HA) input is required in the dual-input, two-compartment model
of studying liver function with dynamic gadoxetate-enhanced MRI. The dynamic
contrast agent concentration in abdominal aorta (AA) is usually measured and
used as the arterial input function instead of that measured in the HA because
its small size. In this study, we determined the relationship between these two
concentrations. We found that the measured arterial input function in AA can be
reliably used as the HA input function with a suppression scaling factor taken
into account.
Introduction
The standard dual-input, two-compartment model of
studying liver function with dynamic gadoxetate-enhanced MRI requires the
hepatic artery (HA) input for liver function measurements1,2. It is a challenge to measure the dynamic contrast agent
concentration in the HA due to the small size of the HA, and instead the
dynamic contrast agent concentration in abdominal aorta (AA) is measured and
used as the arterial input function, assuming that the two concentrations are
the same except a time delay of the former. In this study, we determined the
relationship between these two concentrations.Theory
In a blood vessel at position x, the total derivative
of contrast agent concentration C(x,t) with respect to time t is governed
by the equation:
dC(x,t)/dt = ∂C(x,t)/∂t + V(x,t)∙∂C(x,t)/∂x = -C(x,t)·∂V(x,t)/∂x (1)
where ∂C(x,t)/∂t and ∂C(x,t)/∂x are the partial
derivate of C(x,t) with respect to t
and x, respectively (Fig. 1). V(x,t)
is the blood flow velocity. For constant blood flow V, we have dC(x,t)/dt = 0 or ∂C(x,t)/∂t = -V∙∂C(x,t)/∂x, i.e., a temporal increase in C(x,t) is accompanied with a spatial
decrease in C(x,t), and vice versa.
The solution is C(x2,t+ẟt) = C(x1,t) with ẟt = (x2 – x1)/V. However, for an increased blood flow,
i.e., ∂V(x,t)/∂x > 0, it reduces C(x,t)
as shown in Eq. (1), and this reduction in C(x,t)
should result in a reduced DCE-MRI signal. The DCE-MRI time signal S(x,t)
is related to C(x,t) by the FLASH
equation3:
S(x,t) = ρ0sin(θ)[1 - e-TR·R1(x,t)]/[1 - cos(θ)·e-TR·R1(x,t)] (2)
where R1(x,t) = R1(x,0) + α∙C(x,t), TR is the repetition time, θ is the flip angle, R1(x,0) is the pre-contrast longitudinal relaxation rate of water, and
α is the relaxivity associated with the contrast agent. For constant blood flow
with assumption of R1(x2,0) = R1(x1,0),
we found
S(x2,t+δt)/S(x2,0) = S(x1,t)/S(x1,0) (3)
where S(x,0) is the pre-contrast signal intensity.Methods and Materials
Six pigs underwent a 42.35
min DCE-MRI scan (Siemens 1.5T scanner, vibe pulse sequence, TE/TR=1.77/3.65 ms,
FA=12°, FOV=320mm, matrix 256x256, 64 slices with slice thickness 3mm, 3.51s
acquisition time for each image volume, and 724 volumes). Two scans were
performed with Gd-EOB-DTPA injection, two were performed with Gd-DTPA injection
and the other two were performed with Gd-BOPTA injection. We performed three
region of interest (ROI) analyses using AFNI4. (i): To test Eq. (3), three ROIs with 27
voxels each were placed in the abdominal aorta. The 1st ROI was
placed right above the celiac artery (CA), the 2nd ROI with a
distance of 12 mm was placed above the 1st ROI, and the 3rd
ROI with the same 12 mm distance was placed below the 1st ROI (Fig.
2). (ii): The diameter of CA is less than half of that of AA, and this
substantially reduced size from AA to CA may increase V(x,t) from AA to CA, resulting in a reduced S(x,t) from
AA to CA as discussed above. To test this prediction, the 4th ROI
with 27 voxels was placed in the CA just outside the
AA (Fig. 3). (iii): The 5th ROI with 27 voxels was placed within the
HA to determine the relationship of S(x,t) between
the HA and AA (Fig. 4).Results and Discussion
For each of the five
ROIs in each scan, the mean signal S(t) averaged over all 27
voxels within that ROI was first computed, followed by computing its baseline
mean signal S(0) averaged over the baseline period (22 time points), and
then measuring their relative signal, i.e., ΔS(t)=S(t)/S(0). Then, we computed the group-mean ΔS(t) averaged over the six subjects. Fig. 2
illustrates this group-mean ΔSAA(t) for the three ROIs in the AA. These three
time signals were almost identical. As the blood flow velocity in arteries is
ranged from 49 – 190 mm/s, the maximum time (ẟt) it takes to pass the 24
mm distance should be 0.49s that is much smaller than the 3.51s temporal
resolution of the DCE-MRI. The blood flow velocity in this part of the AA was
most likely the same under the well-controlled experimental condition.
Accordingly, the almost identical ΔSAA(t) in these three ROIs are consistent with Eq.
(3). Fig. 3 illustrates the group-mean ΔSCA(t) in the CA and Fig. 4 illustrates ΔSHA(t) in the HA, respectively. The magnitude of ΔSCA(t) was clearly reduced relative to ΔSAA(t), consistent with our prediction. To
compute this reduction rate (R), we first computed the ratio of [ΔSCA(t) - 1] to [ΔSAA(t) - 1] for each
time point for the post-contrast period, and then its mean and SD, resulted in
R=0.743±0.030. The magnitude of
ΔSHA(t) was further reduced as shown in Fig.
4, and R=0.705±0.032. Using the mean R value, the curve ΔSFIT(t) = R∙[ΔSAA(t) - 1] +1 is
almost identical to the reduced ΔS(t) for both CA
and HA, showing a quantitative relationship of contrast agent concentration
between AA and both CA and HA,
respectively. In conclusion, the measured arterial input function in AA can be
reliably used as the HA input function with the reduction factor R taken into
account.Acknowledgements
No acknowledgement found.References
1 Sourbron,
S., Sommer, W. H., Reiser, M. F. & Zech, C. J. Combined quantification of
liver perfusion and function with dynamic gadoxetic acid-enhanced MR imaging. Radiology 263, 874-883, doi:10.1148/radiol.12110337 (2012).
2 Simeth, J. et al. Quantification of liver function by linearization of a
two-compartment model of gadoxetic acid uptake using dynamic contrast-enhanced
magnetic resonance imaging. NMR Biomed
31, e3913, doi:10.1002/nbm.3913
(2018).
3 Haacke, E. M. et al. New algorithm for quantifying vascular changes in dynamic
contrast-enhanced MRI independent of absolute T1 values. Magn Reson Med 58,
463-472, doi:10.1002/mrm.21358 (2007).
4 Cox, R. W. AFNI: software for
analysis and visualization of functional magnetic resonance neuroimages. Comput Biomed Res 29, 162-173 (1996).