Manjunathan Nanjappa1 and Arunark Kolipaka1
1Department of Radiology, The Ohio State University, Columbus, OH, United States
Synopsis
Magnetic Resonance Elastography is
an emerging alternative diagnostic tool for palpation and it is known that
stiffness is altered in chronic pancreatitis and as well as in pancreatic
tumors. In this study we compared performances of a drum-shaped rigid driver
with a flat soft-pad passive driver in 10 volunteers at 40Hz and 60Hz diver
frequencies. The variations of mean stiffness values measured between the drivers
were found to be insignificant at a given frequency. Additionally, stiffness measured
at 40Hz is lower compared to 60Hz. Therefore, users can make their choice of
the driver based on patient comfort, longevity and cost.
Introduction
Pancreatic cancer is the third
leading cause of death in United States. There is a need for noninvasive
imaging methods to diagnose various pancreatic diseases such as pancreatic
cancer, chronic pancreatitis (CP) etc.
There are many limitations with current noninvasive imaging methods. It
is known that pancreatic stiffness is altered in diseases such as pancreatic
tumors or CP [1]. Magnetic resonance elastography (MRE) is a noninvasive method
to estimate stiffness of soft tissues without the need for invasive biopsies. Anatomically
pancreas is a complex structure and situated deep in the abdomen. Therefore,
the major challenge of MRE is to be able to generate adequate vibrations for
better wave penetration in the pancreas. Therefore, the aim of the study was to
determine the performance of two passive driver designs at two different
frequencies to determine variation in stiffness measurements. Methods
All imaging was performed in 10
healthy volunteers using a 3T MRI scanner (MAGNETOM Prisma, Siemens Healthcare,
Erlangen, Germany) after obtaining written informed consent. A spin echo- echo
planar imaging (SE-EPI) MRE sequence was performed to obtain 9 slices of
pancreas. The image acquisition was
performed on each subject using 1. Drum shaped (rigid) plastic diaphragm
passive driver (commercial driver used for liver MRE) from Resoundant Inc,
Rochester, MN USA and 2. Flat soft pad passive driver developed as described in
[2]. In both the acquisitions, only passive drivers were replaced without
moving the subject while using same active driver (Resoundant Inc, Rochester,
MN USA), and post processing algorithm.
The passive drivers were placed at
the epigastrium and partially on the ribcage centered around the pancreas and the following imaging parameters were used:
FOV: 380x380 mm; Matrix: 128x256
(interpolated to 256x256); Slice
thickness: 2.5 mm; EPI factor:
32; GRAPPA: 2. This protocol is repeated twice at different
driver frequencies 60Hz/80Hz MEG (TR/TE:
1050/36ms and 40Hz/120Hz MEG (TR/TE: 1125/27ms ). Both in plane and through plane
wave images were acquired to obtain 3D displacement field. 3D local
frequency estimation inversion was performed to report the stiffness values.
Center 5 slices were used for reporting the stiffness values to avoid any edge
effects. Students paired t-test was
performed to determine the statistical difference between stiffness values
obtained using two driver setups. P<0.05 is considered insignificant.Results
Figure 1 shows magnitude image with
red contour delineating the pancreas along with good discernible waves (all the
components of wave field) using rigid driver and soft driver with corresponding
stiffness maps at 40Hz. Similarly, Figure 2 shows magnitude image with red
contour delineating the pancreas along with good discernible waves (all the
components of wave field) using rigid driver and soft driver with corresponding
stiffness maps at 60Hz. Figure 3 shows box plot of stiffness values obtained at
40Hz and 60Hz using both the driver setups with no significant difference. Figure 4 shows line plots of stiffness values
at 40Hz and 60Hz using both driver setups. Discussion
This
study demonstrated that commercial rigid drum driver and soft pad driver
generated adequate waves in the pancreas. There was no significant difference
in stiffness values obtained using both drivers. Additionally, 60Hz waves
penetrated deep into the pancreas without much attenuation similar to 40Hz
vibrations. 60Hz vibrations will generate shorter wavelengths compared to 40Hz
and will potentially enable to generate relatively robust stiffness
measurements using the current inversion (which assumes waves are propagating
in the infinite medium). Additionally,
as expected 40Hz generated lower stiffness values compared to 60Hz frequency.
Therefore, this study recommends 60Hz frequency of vibrations with commercial
rigid driver (used for liver MRE) without the need of developing new driver for
imaging pancreas. Additionally, we also
suggest that based on the patient’s comfort the both drivers can be used
accordingly. Conclusion
This
study demonstrated that both the drivers (rigid drum and soft pad) are equally
efficient and delivers stiffness values that are not significantly different to
each other when the measurements are obtained in similar settings. As expected
the overall ranges of stiffness values measured at different driver frequencies
varied for both the drivers used.Acknowledgements
N/AReferences
1. Janssen J,
Schlorer E, Greiner L. EUS elastography of the pancreas: feasibility and
pattern description of the normal pancreas, chronic pancreatitis, and focal
pancreatic lesions. Gastrointest Endosc. 2007;65(7):971–978.
2. Shi
Y, Glaser KJ, Venkatesh SK, Ben-Abraham EI, Ehman RL. Feasibility of using 3D
MR elastography to determine pancreatic stiffness in healthy volunteers. J Magn
Reson Imaging. 2015 Feb;41(2):369-75. doi: 10.1002/jmri.24572.